Skip to content

ARCHITECT 7D71-23Recalls & reported problems

Abbott GmbH · Phosphorus · Class 1

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380740161736), received between Feb 24, 2022 and Feb 13, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “High Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results1
  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 13, 2024Low Test Results; Low Test Results

    THE CUSTOMER OBSERVED FALSELY DECREASED PHOSPHORUS RESULTS GENERATED ON THE ARCHITECT C16000 PROCESSING MODULE FOR MULTIPLE PATIENT SAMPLES. THE CUSTOMER PROVIDED THE FOLLOWING EXAMPLE RESULTS: (CUSTOMER PROVIDED NORMAL RANGE 2.5-4.5 MG/DL) (B)(6) INITIAL RESULT = <0.7 MG/DL, REPEAT RESULT ON (C8000) = 2.1 MG/DL. (B)(6) INITIAL RESULT = 1.3 MG/DL, REPEAT RESULT ON (C8000) = 2.8 MG/DL. (B)(6) INITIAL RESULT = 1.2 MG/DL, REPEAT RESULT ON (C8000) = 2.7 MG/DL. (B)(6) INITIAL RESULT = 0.9 MG/DL, REPEAT RESULT ON (C8000) = 2.3 MG/DL. (B)(6) INITIAL RESULT = 1.7 MG/DL, REPEAT RESULT ON (C8000) = 3.2 MG/DL. (B)(6) INITIAL RESULT = <0.7 MG/DL, REPEAT RESULT ON (C8000) = 1.8 MG/DL. (B)(6) INITIAL RESULT = 1.7 MG/DL, REPEAT RESULT ON (C8000) = 3.3 MG/DL. (B)(6) INITIAL RESULT = 1.2 MG/DL, REPEAT RESULT ON (C8000) = 2.6 MG/DL. (B)(6) INITIAL RESULT = 1.6 MG/DL, REPEAT RESULT ON (C8000) = 3.0 MG/DL. (B)(6) INITIAL RESULT = 0.8 MG/DL, REPEAT RESULT ON (C8000) = 2.1 MG/DL. THE CUSTOMER REPORTED AN UNKNOWN NUMBER OF PATIENTS WERE PROVIDED AN UNSPECIFIED TREATMENT BASED ON RESULTS AND IS NOT AWARE OF ANY ADVERSE EFFECT ON THE PATIENTS. THE RESULTS WERE THEN REPEATED ON THE ARCHITECT C8000 AND THE C

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18694258 ↗
  • MalfunctionFeb 24, 2022High Test Results; High Test Results

    THE CUSTOMER STATED THAT DISCREPANT ARCHITECT PHOSPHORUS RESULTS WERE GENERATED FOR PATIENT SAMPLE ID (B)(6). AN INITIAL RESULT OF 0.94 MMOL/L WAS GENERATED (WITHIN NORMAL RANGE) AND THE CUSTOMER DECIDED TO RETEST THE SAMPLE WHICH RETESTED AT <0.21 MMOL/L (BELOW NORMAL RANGE) USING TWO DIFFERENT REAGENT LOTS. THE CUSTOMER USES SAMPLE COLLECTION TUBES WITH A SEPARATING GEL. IT IS SUSPECTED THAT THE SAMPLE PROBE POSSIBLY WENT INTO THE GEL. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13592656 ↗

Other Phosphomolybdate (colorimetric), inorganic phosphorus devices

All Phosphomolybdate (colorimetric), inorganic phosphorus devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.