Phosphate (Inorganic) ver.2 08058610190Recalls & reported problems
Roche Diagnostics GmbH · Class 1
All models of Phosphate (Inorganic) ver.2 (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613336121481), received between Jul 12, 2022 and Jul 16, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Low Test Results2
- Non Reproducible Results2
- High Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 16, 2026Low Test Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25843288 ↗MalfunctionNov 10, 2025Non Reproducible Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE PHOSPHATE (INORGANIC) VER.2 RESULTS FOR 5 PATIENT SAMPLES ON A COBAS C 703 ANALYTICAL UNIT. THE CUSTOMER PROVIDED AN EXAMPLE OF DISCREPANT RESULTS FOR 2 PATIENT SAMPLES TESTED. FOR SAMPLE 1, THE INITIAL RESULT WAS 9.5 MG/DL. THE REPEAT RESULT WAS 5.4 MG/DL. FOR SAMPLE 2, THE INITIAL RESULT WAS APPROXIMATELY 8 MG/DL. THE REPEAT RESULT WAS 2 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23512424 ↗MalfunctionDec 14, 2022High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results
THE INITIAL REPORTER COMPLAINED OF DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED FOR PHOSPHATE (INORGANIC) VER.2 (PHOS2) ON A COBAS PRO C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS 15.6 MG/DL. THIS RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. THE REPEAT RESULT WAS 8.60 MG/DL. THE PHOS2 REAGENT LOT NUMBER AND EXPIRATION DATE WERE NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15975618 ↗MalfunctionJul 12, 2022Low Test Results; Low Test Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH PHOS2 PHOSPHATE (INORGANIC) VER.2 ON A COBAS C 503 ANALYTICAL UNIT (SERIAL NUMBER (B)(4)). NO INCORRECT RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE SAMPLE INITIALLY RESULTED IN A PHOSPHORUS VALUE OF 2.69 MG/DL ACCOMPANIED BY A DATA FLAG INDICATING ATYPICAL KINETIC BEHAVIOR. THE SAMPLE WAS REPEATED, RESULTING IN A PHOSPHORUS VALUE OF 2.68 MG/DL ACCOMPANIED BY A DATA FLAG INDICATING ATYPICAL KINETIC BEHAVIOR. THE SAMPLE WAS DILUTED 1:3 AND REPEATED, RESULTING IN A PHOSPHORUS VALUE OF 2.33 MG/DL ACCOMPANIED BY A DATA FLAG. THE SAMPLE WAS REPEATED AGAIN AT DECREASED SAMPLE VOLUME, RESULTING IN A PHOSPHORUS VALUE OF 2.55 MG/DL ACCOMPANIED BY A DATA FLAG INDICATING ATYPICAL KINETIC BEHAVIOR. THE SAMPLE WAS THEN REPEATED ON A SECOND ANALYZER AT DECREASED SAMPLE VOLUME, RESULTING IN A PHOSPHORUS VALUE OF 3.54 MG/DL. THIS VALUE WAS DEEMED CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14984854 ↗
Other Phosphomolybdate (colorimetric), inorganic phosphorus devices
- VITROS 1513209Ortho-Clinical Diagnostics, Inc. · Phosphomolybdate (colorimetric), inorganic phosphorus1 recall0 reports
- Synchron 467868Beckman Coulter, Inc. · Phosphomolybdate (Colorimetric), Inorganic Phosphorus1 recall0 reports
- Phosphate (Inorganic) ver.2 03183793122Roche Diagnostics GmbH · Phosphomolybdate (colorimetric), inorganic phosphorus11 reports
- Phosphate (Inorganic) ver.2 5171377190Roche Diagnostics GmbH · Phosphomolybdate (colorimetric), inorganic phosphorus5 reports
- ARCHITECT 7D71-23Abbott GmbH · Phosphomolybdate (colorimetric), inorganic phosphorus2 reports
- Dimension Vista® Flex® reagent cartridge PHOS 10445144Siemens Healthcare Diagnostics Inc. · Phosphomolybdate (colorimetric), inorganic phosphorus1 report
- Dimension® Flex® reagent cartridge PHOS 10720277Siemens Healthcare Diagnostics Inc. · Phosphomolybdate (colorimetric), inorganic phosphorus1 report
- Piccolo Renal Function Panel 400-0027Abaxis, Inc. · Phosphomolybdate (Colorimetric), Inorganic Phosphorus1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page