Phosphate (Inorganic) ver.2 5171377190Recalls & reported problems

Roche Diagnostics GmbH · Class 1

All models of Phosphate (Inorganic) ver.2 (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04015630925094), received between Jan 25, 2023 and Sep 11, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “High Test Results” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results2
  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Low Test Results1
  • No Device Output1
  • Non Reproducible Results1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 11, 2026High Test Results

    There was an allegation of questionable Phosphate (Inorganic) ver.2 results for multiple patient samples on a cobas 8000 cobas c 701 module. The doctor questioned the initial results with prompted the customer to repeat the samples.On (b)(6) 2026, the customer observed multiple patient sample results initially greater than 4 mmol/L, and upon repeat were 1 mmol/L or lower. On (b)(6) 2026, the customer again observed 3 patient sample results initially between 2 and 5 mmol/L, and upon repeat, around 1 mmol/L. The exact patient results were not provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50734305 ↗
  • MalfunctionDec 3, 2025Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THERE WAS AN ALLEGATION OF QUESTIONABLE PHOSPHATE (INORGANIC) VER.2 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS 8000 C702 MODULE. THE CUSTOMER WAS PROMPTED TO REPEAT THE PATIENT SAMPLE AS THEIR QC SHIFTED. THE CUSTOMER RECALIBRATED AND REPEATED QC PRIOR TO REPEATING THE SAMPLE. THE INITIAL RESULT WAS 1.5 MG/DL. THE REPEAT RESULT WAS 1.0 MG/DL. THE REPEAT RESULT WAS DEEMED CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23706931 ↗
  • MalfunctionDec 19, 2024Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results

    THE INITIAL REPORTER COMPLAINED OF DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED FOR PHOSPHORUS (PHOS2) ON A COBAS 8000 C 702 MODULE. THE INITIAL RESULT WAS 0.00 MG/DL. THE OPERATOR REQUESTED REPEAT TESTING. THE REPEAT RESULT WAS 3.79 MG/DL. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20979472 ↗
  • MalfunctionMar 6, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH PHOS2 (PHOSPHORUS) ON A COBAS 8000 C702 MODULE. THE SAMPLE INITIALLY RESULTED IN A PHOSPHORUS VALUE OF 0.9 MG/DL. DUE TO THE VALUE BEING LOW, THE SAMPLE WAS REPEATED, RESULTING IN A VALUE OF 9.5 MG/DL. THE SAMPLE WAS REPEATED SEVERAL TIMES, BUT NOT ADDITIONAL VALUES WERE PROVIDED. THE 0.9 MG/DL VALUE WAS DEEMED CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18847427 ↗
  • MalfunctionJan 25, 2023No Device Output; High Test Results; No Device Output; High Test Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE PHOSPHORUS GEN.2 RESULTS FOR ONE PATIENT FROM COBAS 8000 C702 MODULE SERIAL NUMBER (B)(4). THE CUSTOMER ALLEGED INTERFERENCE IN THE SAMPLES. ON (B)(6) 2022, THE RESULT WAS 9.80 MG/DL. THE SAMPLE WAS REPEATED AND THE SAME RESULT WAS OBTAINED. ON (B)(6) 2022, THE RESULT FROM A NEW SAMPLE WAS 9.58 MG/DL. THE QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16241679 ↗

Other Phosphomolybdate (colorimetric), inorganic phosphorus devices