Skip to content

Interlink 303347Recalls & reported problems

Becton, Dickinson And Company · 5ml Syringe Luer Lock Tip with Blunt Plastic Cannula · Class 2

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00382903033478), received between Nov 5, 2020 and Apr 5, 2023 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K974363); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Device Contamination with Chemical or Other Material2
  • Device-Device Incompatibility1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Patient Involvement1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 5, 2023Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT BLACK FOREIGN MATTER WAS FOUND ON THE BD LUER-LOK¿ TIP SYRINGE WITH BLUNT PLASTIC CANNULA. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "FOUND CONTAMINATED WITH A BLACK SUBSTANCE THAT LOOKED LIKE DROPS OF INK OR MOLD. IT WAS RECOGNIZED PRIOR TO USE ON THE PATIENT".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16683140 ↗
  • MalfunctionSep 22, 2022Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE BD SYRINGE 5ML LL W/BLUNT PLASTIC CANNULA BROKE OFF IN VIAL.THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THIS IS A PICTURE OF THE PRODUCT THAT WE HAVE SEEN THE TIP BRAKE OFF OF SEVERAL TIMES IN VIALS OF MEDICATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15463923 ↗
  • MalfunctionJun 2, 2022Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT 1 BD SYRINGE 5ML LL W/BLUNT PLASTIC CANNULA HAD CORING ISSUES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER : THE CONSUMER FACILITY REPORTED ISSUES WITH THE BLUNT PLASTIC CANNULAS THAT WHEN THE NURSING STAFF ARE USING THESE TO DRAW UP MEDICATIONS FROM A VIAL, THEY ARE PUSHING PART OF THE RUBBER STOPPER INTO THE BOTTLE. EVENT DATE : UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14578690 ↗
  • MalfunctionNov 5, 2020Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE BD 10ML SYRINGE WITH LUER-LOK TIP WITH BLUNT PLASTIC CANNULA. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN UNPACKING, THE HCP FOUND THAT THE TIP OF THE SYRINGE (BLUNT CANNULA/VIAL ACCESS COMPONENT) WAS DAMAGED.

    Patient problems: No Patient Involvement

    MAUDE report 10796068 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K974363. One submission often covers many models, so check the product and lot details in each record.

  • Z-1055-2015 · Dec 19, 2014 · Terminated
    BD 10 mL Syringe BD Luer-Lok Tip with Blunt Plastic Cannula, Not for Skin Injection ; Catalog #303348, BD 10 mL Syringe Luer-Lok Tip with Blunt Fill Needle 18G x 1 1/2 (1.2mm x 4
  • Z-1461-06 · Jun 21, 2006 · Terminated
    BD Vacutainer Blood Collection Assembly with BD Blunt Plastic Cannula; Catalog #303380.

Other Needle, hypodermic, single lumen devices

All Needle, hypodermic, single lumen devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.