Skip to content

N/A 305155Recalls & reported problems

Becton, Dickinson And Company · NEEDLE 22X1 RB · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00382903051557), received between Sep 14, 2018 and Jul 17, 2024 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Device Markings/Labelling Problem” (2 reports).

3 other recalls cover devices cleared under the same FDA submission (K021475); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Markings/Labelling Problem2
  • Loose or Intermittent Connection2
  • Break1
  • Contamination /Decontamination Problem1
  • Detachment of Device or Device Component1
  • Device Contamination with Chemical or Other Material1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Known Impact Or Consequence To Patient2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 17, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem

    MATERIALS#: 305155. BATCH#: 1300351. IT WAS REPORTED BY THE CUSTOMER THAT CUSTOMER WHO DOES WEEKLY INJECTION NOTICED THAT THERE WAS 3 OF ITEM 305196 LOT 1120197 AND 2 OF ITEM 305196 LOT -2327243 WITHIN THE BOX OF 100 COUNT ITEM 305155.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19762464 ↗
  • MalfunctionApr 13, 2023Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT THE BD PRECISIONGLIDE¿ NEEDLE PULLED OUT OF THE HUB WHEN TRYING TO REMOVE THE NEEDLE FROM THE SYRINGE. THIS COMPLAINT WAS CREATED TO CAPTURE THE 1ST OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "HAPPENED TWICE- WHEN TRY TO CAP AND REMOVE THE NEEDLE OFF SYRINGE THE METAL NEEDLE COMES BUT GRAY PLASTIC REMAINS ON THE SYRINGE (HUB) BREAKS AWAY FROM THE NEEDLE ( WHEN TRYING TO REMOVE THE NEEDLE FROM SYRINGE )".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16740465 ↗
  • MalfunctionApr 12, 2023Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT THE BD PRECISIONGLIDE¿ NEEDLE PULLED OUT OF THE HUB WHEN TRYING TO REMOVE THE NEEDLE FROM THE SYRINGE. THIS COMPLAINT WAS CREATED TO CAPTURE THE 2ND OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ISSUE : HAPPENED TWICE- WHEN TRY TO CAP AND REMOVE THE NEEDLE OFF SYRINGE THE METAL NEEDLE COMES BUT GRAY PLASTIC REMAINS ON THE SYRINGE (HUB) BREAKS AWAY FROM THE NEEDLE ( WHEN TRYING TO REMOVE THE NEEDLE FROM SYRINGE )".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16729094 ↗
  • MalfunctionMar 22, 2021Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE NEEDLE 22X1 RB PULLED OUT OF THE HUB WHILE INJECTING A DOG AND REMAINED BETWEEN THE SHOULDER BLADES. THE NEEDLE WAS ABLE TO BE REMOVED BY HAND FROM THE PATIENTS BODY. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE MEDICATION WAS DRAWN UP WITH A DIFFERENT NEEDLE. THIS NEEDLE WAS PLACED ONTO THE SYRINGE AND GAVE A SUBCUTANEOUS INJECTION TO A SMALL DOG BETWEEN THE SHOULDER BLADES. THE DOG WAS VERY STILL AND AS THE NEEDLE WAS PULLED OUT IT WAS NOTICED THAT THE NEEDLE WAS NO LONGER ON THE HUB. LOOKED AT THE PATIENT AND THE NEEDLE WAS SLIGHTLY STICKING OUR AND WAS ABLE TO PULL IT OUT OF THE PATIENT."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11548394 ↗
  • MalfunctionMar 4, 2021Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT A NEEDLE 22X1 RB HAD FOREIGN MATTER DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "IT WAS REPORTED THAT THERE IS A SMALL STYROFOAM BALL IN THE BEVEL TIP. VERBATIM: PER CUSTOMER'S RESPONSE: (B)(6) 2021. EVENT DATE (B)(6) 2021. NO ONE HURT. OPENED A NEW BD PRECISIONGLIDE NEEDLE 22G X 1 LOT# 9162963 AND WHEN UNCAPPING THE NEEDLE FOUND WHAT APPEARS TO BE A SMALL STYROFOAM BALL IN THE BEVEL TIP."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11417969 ↗
  • MalfunctionFeb 18, 2021Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT THERE WAS NO EXPIRATION DATE LISTED ON THE NEEDLE 22X1 RB BOX. THIS COMPLAINT WAS CREATED TO CAPTURE THE 3RD OF 3 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: PHARMACIST STATED, BOXES DON'T HAVE AN EXPIRATION DATE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11346359 ↗
  • MalfunctionMar 19, 2019Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING USE OF THE BD PRECISIONGLIDE NEEDLE THAT SYRINGE NEEDLE BROKE OFF OF SYRINGE WHILE DRAWING AND WAS STUCK IN THE VIAL. THE NEEDLE BROKE OFF IN CONSUMERS LEG AND IN THE VIAL. FROM PHONE CALL ON (B)(6) 2019 16:06:07: CONSUMER CALLED, I UPDATED THE CONTACT INFORMATION. PROBLEM OCCURRED ABOUT 3 WEEKS AGO. CONSUMER ALSO STATED THAT ONE NEEDLE BROKE OFF IN HIS LEG DURING INJECTION, HE REMOVED IT WITH HIS FINGERS, NO MEDICAL INTERVENTION, NO INJURY. SAMPLES WERE DISCARDED, SENDING REPLACEMENT BOX TO THE PHARMACY. PHARMACIST REPORTED ON BEHALF OF CONSUMER THAT SYRINGE NEEDLE BROKE OFF OF SYRINGE WHILE DRAWING AND WAS STUCK IN THE VIAL. OCCURRENCE DATE IS UNKNOWN, LOT # 5336578, PRODUCT # 305155, EXP 1-31-2021.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8433301 ↗
  • MalfunctionSep 14, 2018Contamination /Decontamination Problem

    IT WAS REPORTED WITH THE USE OF THE BD PRECISIONGLIDE¿ NEEDLE THERE WAS AN ISSUE WITH FOREIGN MATTER. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7877674 ↗

FDA clearance and classification

Cleared through 510(k) K021475 on Jul 19, 2002 (decision: Substantially Equivalent). Submitted by Bd as “Bd Single Lumen Needle, Syringe And Blood Collection Set”. The FDA review took 72 days. FDA record ↗

The FDA regulates product code FMI (needle, hypodermic, single lumen) under 21 CFR 880.5570 as a Class II device. 21 CFR 880.5570 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K021475. One submission often covers many models, so check the product and lot details in each record.

  • Z-0588-2021 · Nov 18, 2020 · Terminated
    BD PrecisionGlide Needle 30G x 1/2 (0.3 mm X 13mm) REF 305106 - Product Usage: used for general purpose injection and aspiration of fluids from vials, ampoules and parts of the bod
  • Z-0207-2018 · May 31, 2017 · Terminated
    BD Precision Glide Needle 18G x 1 RB, Catalog 305195 Product Usage: These needles are intended for general purpose fluid injection/aspiration, infusion, venipuncture to obtain bl
  • Z-1009-2012 · Oct 8, 2011 · Terminated
    BD 30G1/2" Precision Glide" Needle Needles are used for aspiration and injection of medications

Common questions about the N/A 305155

Has the N/A 305155 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00382903051557), based on the FDA recall file dated Nov 27, 2026. However, 3 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the N/A 305155?

The FDA's MAUDE database lists 8 adverse event reports for this device, received between Sep 14, 2018 and Jul 17, 2024. In those reports, the most common specific device problems are “Device Markings/Labelling Problem” (2), “Loose or Intermittent Connection” (2) and “Break” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the N/A 305155?

None of the reports are classified as deaths, 0 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the N/A 305155 FDA approved?

It was cleared through 510(k) K021475 on Jul 19, 2002. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the N/A 305155, and is it still sold?

The FDA's device database lists Becton, Dickinson And Company as the company responsible for it (brand N/A, model 305155). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Needle, hypodermic, single lumen devices

All Needle, hypodermic, single lumen devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page