Skip to content

N/A 305175Recalls & reported problems

Becton, Dickinson And Company · NEEDLE 20GA 1IN · Class 2

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00382903051755), received between Oct 9, 2020 and Feb 20, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (2 reports).

3 other recalls cover devices cleared under the same FDA submission (K021475); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash2
  • Complete Blockage1
  • Delivered as Unsterile Product1

Patient problems

  • No Consequences Or Impact To Patient2
  • Needle Stick/Puncture1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 20, 2024Complete Blockage; Complete Blockage

    NO ADDITIONAL INFORMATION RECEIVED. MATERIAL #:305175. BATCH#:2056681. IT WAS REPORTED BY CUSTOMER THAT THE CLINICAL ANESTHESIA TEAM RECENTLY FOUND BLOCKED HYPO BD NEEDLES. VERBATIM: "NOUS AVONS REÇU UNE RÉCLAMATION POUR LES ARTICLES BD 305175 ET 305106, L¿ÉQUIPE CLINIQUE D¿ANESTHÉSIE A RÉCEMMENT TROUVÉ DES AIGUILLES HYPO BD BLOQUÉES. " "POUR L'ARTICLE # 305175, ON N'EST PAS CERTAIN CAR L'EMBALLAGE A ÉTÉ JETÉ, MAIS IL SE PEUT QUI SOIT UN DE CES LOTS: 2056681, 21117190 OU 20209321". "WE HAVE RECEIVED A COMPLAINT FOR BD 305175 AND 305106 ITEMS, THE CLINICAL ANESTHESIA TEAM RECENTLY FOUND BLOCKED HYPO BD NEEDLES." "FOR ITEM # 305175, WE ARE NOT SURE BECAUSE THE PACKAGING HAS BEEN DISCARDED, BUT IT MAY BE ONE OF THESE LOTS: 2056681, 21117190 OR 20209321". ADDITIONAL INFORMATION: 1. VEUILLEZ NOUS INDIQUER LA DATE OÙ L¿INCIDENT S¿EST PRODUIT. VEUILLEZ CONFIRMER SI LES NUMÉROS DE LOT SONT 21117190 ET 20209321, ET S¿ILS SONT TOUCHÉS ? EITHER ONE WAS A POSSIBILITY. MOST REPORTED INCIDENT: (B)(6) 2024. THE PACKAGING WAS DISCARDED, BUT UPON VERIFICATION OF THE BD NEEDLES IN THE OR, THE LOT NUMBER COULD BE NARROWED DOWN TO 2 -LOT: 2020931 OR 2056681. 2. ÊTES-VOUS EN MESURE DE NOUS FOURNIR UN É

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18739004 ↗
  • MalfunctionAug 2, 2021Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT THE NEEDLE 20GA 1IN PIERCED THROUGH THE SHIELD BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "A PHARMACY TECHNICIAN EXPERIENCED A NEEDLESTICK INJURY WHEN RECAPPING A NEEDLE IN OUR STERILE COMPOUNDING ROOM. THE NEEDLE CAME THROUGH THE SIDE OF THE NEEDLE TIP COVER."

    Patient problems: Needle Stick/Puncture

    MAUDE report 12263201 ↗
  • MalfunctionOct 15, 2020Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT NEEDLE 20GA 1IN LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "IT WAS REPORTED THAT WHEN THE NURSE PULLED THE PLUNGER BACK ON THE 3ML SYRINGE THE MEDICATION SPILLED OUT SURROUNDING THE NEEDLE/SYRINGE CONNECTION POINT. PER ATTACHED FOLLOW UP RESPONSE: FOR 1835952. INCIDENT DATE? (B)(6) 2020 AND (B)(6) 2020. WHAT MEDICATION WAS BEING DRAWN UP? ON (B)(6)- PEVNAR23 AND ON (B)(6)- PREVNAR13. IS PRODUCT AVAILABLE TO RETURN FOR INVESTIGATION? THIS SPECIFIC PRODUCT WAS THROWN AWAY BUT HAVE UNUSED PRODUCTS THAT CAN BE SENT. EVENT DESCRIPTION STATES: THERE HAD BEEN A FEW INSTANCES WHERE ONE OF MY CLINICS HAVE HAD ISSUES WITH A COUPLE BD PRODUCTS WHEN THEY WERE PERFORMING MEDICATION ADMINISTRATION. IN BOTH INSTANCES MEDICATION SPRAYED OUT OF THE NEEDLE/SYRINGE CONNECTION POINT. THE PRODUCTS THAT WERE USED, PRODUCT # AND LOT # ARE BELOW: BD 3ML SYRINGE WITH LUER-LOK TIP PRODUCT # 9870248 LOT # 0122550, BD PRECISION GLIDE NEEDLE 20G X 1¿ PRODUCT # 9872550 LOT # 8324804. PLEASE SEE THE 3 FORMS FOR MORE INFORMATION. ANY INFORMATION REGARDING WHETHER THIS IS A PRODUCT ISSUE OR USER ERROR WOULD BE APPRECIATED. EVENT DETAILS STATES: IN SEPARATE INSTANCE

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10687999 ↗
  • MalfunctionOct 9, 2020Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE NEEDLE 20GA 1IN EXPERIENCED LEAKAGE AT THE CONNECTION SITE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: COMPLAINT 2 OF 5 THERE HAD BEEN A FEW INSTANCES WHERE ONE OF MY CLINICS HAVE HAD ISSUES WITH A COUPLE BD PRODUCTS WHEN THEY WERE PERFORMING MEDICATION ADMINISTRATION. IN BOTH INSTANCES MEDICATION SPRAYED OUT OF THE NEEDLE/SYRINGE CONNECTION POINT. BD PRECISION GLIDE NEEDLE 20G X 1¿ PRODUCT # 9872550, LOT # 8324804. IN SEPARATE INSTANCES, MY NURSING TEAM HAS HAD ISSUES WITH THE BD 3ML SYRINGE WHEN THEY WERE DRAWING UP OR ADMINISTERING MEDICATIONS. THE FIRST INSTANCE CAME ABOUT WHEN NURSING STAFF TIGHTENED A BD PRECISIONGLIDE 20G X 1" NEEDLE ONTO A BD 3ML SYRINGE TO DRAW UP MEDICATION TO ADMINISTER TO A PATIENT. WHEN THE NURSE PULLED THE PLUNGER BACK ON THE 3ML SYRINGE THE MEDICATION SPILLED OUT SURROUNDING THE NEEDLE/SYRINGE CONNECTION POINT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10659134 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K021475. One submission often covers many models, so check the product and lot details in each record.

  • Z-0588-2021 · Nov 18, 2020 · Terminated
    BD PrecisionGlide Needle 30G x 1/2 (0.3 mm X 13mm) REF 305106 - Product Usage: used for general purpose injection and aspiration of fluids from vials, ampoules and parts of the bo
  • Z-0207-2018 · May 31, 2017 · Terminated
    BD Precision Glide Needle 18G x 1 RB, Catalog 305195 Product Usage: These needles are intended for general purpose fluid injection/aspiration, infusion, venipuncture to obtain bl
  • Z-1009-2012 · Oct 8, 2011 · Terminated
    BD 30G1/2" Precision Glide" Needle Needles are used for aspiration and injection of medications

Other Needle, hypodermic, single lumen devices

All Needle, hypodermic, single lumen devices →