Liverty 306466Recalls & reported problems
Suzhou BD Medical Device Co., Ltd · Liverty TIPS Access Set · Class 2
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00382903064663), received between Mar 15, 2023 and Aug 31, 2026 — classified as 3 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Fracture” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture5
- Adverse Event Without Identified Device or Use Problem2
- Break1
- Device Contamination with Chemical or Other Material1
- Difficult to Advance1
- Leak/Splash1
- Material Puncture/Hole1
- Packaging Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Abdominal Pain1
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Hyperbilirubinemia1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 31, 2026Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50517988 ↗MalfunctionFeb 13, 2026Material Puncture/Hole
ON (B)(6) 2026. A PATIENT UNDERWENT A TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT PROCEDURE. IT WAS REPORTED THAT THERE WAS AN ISSUE WITH A LIVERTY SET. THE CUSTOMER STATED THAT THE DEVICE HAD A HOLE OR DISTAL DEFECT, UPON FLUSHING BEFORE USE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24330710 ↗MalfunctionDec 22, 2025Device Contamination with Chemical or Other Material; Packaging Problem; Device Contamination with Chemical or Other Material; Packaging Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT PROCEDURE USING THE LIVERTY TIPS ACCESS SET. PRIOR TO THE PROCEDURE, IT WAS REPORTED THAT THE FOREIGN MATTER WAS FOUND INSIDE THE PUNCTURE KIT PACKAGING UPON OPENING. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23861710 ↗MalfunctionNov 24, 2025Fracture; Fracture
ON (B)(6) 2025, A PATIENT UNDERWENT A PROCEDURE USING TRANS JUGULAR LIVER ACCESS SET. DURING THE PROCEDURE. THE ADJUSTABLE TIP OF THE DEVICE FRACTURED DURING CANNULA ROTATION PRIOR TO TIPS PUNCTURE. THIS REQUIRED DELICATELY REMOVING THE FRACTURED CANNULA AND USING A NEW KIT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23628887 ↗MalfunctionApr 30, 2025Fracture; Fracture
ON (B)(6) 2025, A PATIENT WAS UNDERGOING A TRANS JUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT PROCEDURE USING LIVERTY TIPS ACCESS SET. IT WAS REPORTED THAT DURING THE PROCEDURE, THE NEEDLE ALLEGEDLY DID NOT ASPIRATE. ADDITIONALLY, THE CANNULA ALLEGEDLY SHREDDED THE CATHETER. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21937911 ↗InjuryNov 19, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH RESULTS OF A CLINICAL TRIAL THAT THREE WEEKS AND THREE DAYS POST AN INDEX PROCEDURE USING A LIVERTY TIPS ACCESS SET THROUGH RIGHT JUGULAR VEIN, THE SUBJECT ALLEGEDLY HAD AN ADVERSE EVENT OF HYPOKALEMIA AND HYPERBILIRUBINEMIA. IT WAS FURTHER REPORTED THAT THE ADVERSE EVENT RELATIONSHIP WAS POSSIBLY RELATED TO THE TIPS ACCESS SET AND TIPS STENT GRAFT AND CAUSALLY RELATED TO TIPS INDEX PROCEDURE. REPORTEDLY, THE SUBJECT WAS TREATED WITH POTASSIUM CHLORIDE AND RECOVERING. THE CURRENT STATUS OF THE SUBJECT WAS UNKNOWN.
Patient problems: Hyperbilirubinemia
MAUDE report 20721512 ↗InjuryJan 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL THAT ONE DAY POST A TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT CREATION VIA THE RIGHT INTERNAL JUGULAR VEIN, THE SUBJECT DEVELOPED AN ADVERSE EVENT OF ACUTE INCREASED ABDOMINAL PAIN. IT WAS FURTHER REPORTED THAT THE SUBJECT DEVELOPED AN ADVERSE EVENT OF DIARRHEA, BUT THE SEVERITY OF ADVERSE EVENT WAS MILD. REPORTEDLY, THE OUTCOME OF THE ADVERSE EVENTS WERE RESOLVED. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Abdominal Pain
MAUDE report 18606306 ↗MalfunctionJul 5, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT PROCEDURE, THE CATHETER OF THE KIT ALLEGEDLY HAD BROKEN AND DISLODGED INTO PATIENT. IT WAS FURTHER REPORTED THAT THE PHYSICIAN HAD TO SNARE THE PIECE OF THE CATHETER THAT BROKE LOOSE OUT OF THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17257123 ↗InjuryMar 31, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT PROCEDURE, THE CATHETER ALLEGEDLY SHEARED WHEN TRIED TO PUSH THE NEEDLE. IT WAS FURTHER REPORTED THAT THERE WAS ABOUT AN INCH OF THE CATHETER LEFT IN THE PATIENT. FURTHER IT TOOK THREE HOURS TO REMOVE THE SEGMENT FROM THE PATIENT BODY. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 16654467 ↗MalfunctionMar 15, 2023Break; Fracture; Difficult to Advance; Break; Fracture; Difficult to Advance
IT WAS REPORTED THAT DURING A TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT PROCEDURE, THE DEVICE ALLEGEDLY HAD DIFFICULTY IN TRACKING AND SEEING THE NEEDLE THROUGH INTRACARDIAC ECHOCARDIOGRAPHY. IT WAS FURTHER REPORTED THAT THE FLEX POINT OF THE NEEDLE ALLEGEDLY BROKE AND LOST ALL PUSH AS WELL AS PUNCTURE ABILITY. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16543754 ↗
Other Introducer, catheter devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.