BD Ultra-Fine 320119Recalls & reported problems
Becton Dickinson Pen Limited · BD Ultra-Fine Pen Needles, 5mm, 31G, 5bevel, TW · Class 2
All models of BD Ultra-Fine (4)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This BD Ultra-Fine pen needle is a thin needle used with insulin pens. Many of the 162 reports describe needles that clog or do not deliver insulin, will not attach properly to the pen, or are hard to remove after an injection. Of the reports, 154 are malfunctions and 8 describe injuries, with no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 162
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 162 adverse event reports that list this exact device (DI 00382903201198), received between Oct 26, 2017 and Jul 17, 2025 — classified as 8 injuries, 154 malfunctions.
The most frequently reported specific device problem is “Failure to Deliver” (70 reports).
1 other recall cover devices cleared under the same FDA submission (K213478); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Deliver70
- Misconnection19
- Break13
- Complete Blockage12
- Defective Device12
- Device Markings/Labelling Problem11
- Material Protrusion/Extrusion7
- Leak/Splash6
Patient problems
- No Clinical Signs, Symptoms or Conditions129
- No Consequences Or Impact To Patient17
- No Known Impact Or Consequence To Patient9
- Foreign Body In Patient4
- Hyperglycemia4
- Insufficient Information2
- Needle Stick/Puncture2
- Blister1
Recent adverse event reports
The 25 most recent of 162 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 17, 2025Off-Label Use; Failure to Deliver; Off-Label Use; Failure to Deliver
CONSUMER REPORTED THE NEEDLE FROM LAST NIGHT CLOGGED DURING INJECTION. CALLER DOES NOT COMPLETE A FLOW CHECK. ADVISED CALLER HOW TO PROPERLY ATTACH NON PATIENT END TO PEN. LOT # 4268005 CATALOG# UNKNOWN, PEN NEEDLE, 5MM, MINI DATE OF EVENT 07.06.2025 SAMPLE STATUS DISCARD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22536039 ↗MalfunctionApr 21, 2025Failure to Deliver; Failure to Deliver
CONSUMER REPORTED NEEDLE CLOG DURING INJECTION. HE STATED THAT THERE IS NO INSULIN FLOW. CONSUMER DOES NOT RE-USE. LOT #: 3032929 CATALOG #: 320119 DATE OF EVENT: UNKNOWN SAMPLES: DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21876762 ↗MalfunctionApr 21, 2025Misconnection; Defective Device; Misconnection; Defective Device
CONSUMER REPORTED THAT THE NEEDLES WILL NOT ATTACH AND THE NEEDLES ARE DIFFICULT TO REMOVE AFTER INJECTION. CONSUMER DOES NOT RE-USE. LOT #: 4107078 CATALOG #: 320119 DATE OF EVENT: UNKNOWN SAMPLES: DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21872724 ↗MalfunctionApr 17, 2025Misconnection; Defective Device; Misconnection; Defective Device
CONSUMER REPORTED SOME PEN NEEDLES ARE HARD TO ATTACH OR CAN'T BE ATTACHED. STATED, PEN NEEDLES ARE HARD TO REMOVE AFTER INJECTIONS FOR THE ONES SHE CAN ATTACH LOT: 4093075. CATALOG: 320119. DATE OF EVENT: UNKNOWN. SAMPLES: YES CL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21859366 ↗MalfunctionMar 25, 2025Misconnection; Misconnection
CONSUMER REPORTED FINDING 10-15 PEN NEEDLES FROM 2 BOXES THAT ARE HARD TO THREAD ONTO THE PEN PROPERLY. CAN'T USE THEM. INFORMED CALLER PROPER NON-PATIENT END PLACEMENT. (B)(6). LOT # 4051682 1 BOX. LOT # UNKNOWN = 1 BOX. CATALOG# 320119. DATE OF EVENT UNKNOWN. SAMPLE STATUS DISCARD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21688780 ↗MalfunctionMar 21, 2025Misconnection; Misconnection
CONSUMER REPORTED THAT THE NEEDLES WILL NOT ATTACH. SHE ALSO STATED THAT THE PATIENT END NEEDLE IS BENT. CONSUMER NOTICED IT AFTER INSPECTING THE NEEDLE WHEN IT WOULD NOT ATTACH. LOT #: 4184005. CATALOG #: 320119. DATE OF EVENT: 02-27-25. SAMPLES: AVAILABLE - SENDING MAIL KIT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21664804 ↗MalfunctionFeb 11, 2025Off-Label Use; Failure to Deliver; Off-Label Use; Failure to Deliver
CONSUMER REPORTED DURING THE INJECTION THE NEEDLE CLOGS. CALLER DOES NOT PRIME THE NEEDLE. DC. LOT # 4086048, CATALOG# 32011, DATE OF EVENT UNKNOWN, SAMPLE STATUS AWAITING SAMPLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21356071 ↗MalfunctionJan 2, 2025Failure to Deliver; Failure to Deliver
PHARMACIST REPORTED ISSUE ON BEHALF OF CONSUMER. STATED, NO INSULIN FLOW WHEN PRIMING PEN NEEDLES. STATED, THE CONSUMER IS USING THE ECHO PEN. PHARMACIST ASKED IF THE PEN NEEDLES ARE COMPATIBLE WITH THE ECHO PEN. I WAS ABLE TO ASSIST THE PHARMACIST WITH PRIMING TWO PEN NEEDLES WHILE ON THE CALL. LOT: 4107078, CATALOG: 320119, DATE OF EVENT: UNKNOWN, SAMPLES: YES (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21056233 ↗MalfunctionDec 31, 2024Failure to Deliver; Failure to Deliver
CONSUMER REPORTED NEEDLE CLOG DURING PRIMING. SHE STATED THAT THERE IS NO INSULIN FLOW. CONSUMER DOES NOT RE-USE. LOT #: 4093073, CATALOG #: 320119, DATE OF EVENT: UNKNOWN, SAMPLES: DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21048236 ↗MalfunctionNov 18, 2024Failure to Deliver; Defective Device; Failure to Deliver; Defective Device
CONSUMER REPORTED WHEN USING THE PEN NEEDLES WITH HIS FLEXTOUCH PEN, THE NON PATIENT END NEEDLE WILL NOT ATTACH ONTO THE PEN NOR WILL IT PRESS OUT THE INSULIN. HAS TRIED SEVERAL PENS AND PEN NEEDLES FROM THIS BOX. DC CONSUMER REPORTED HAS NOT BEEN ABLE TO TAKE HIS INSULIN DUE TO THIS ISSUE. INFORMED CALLER OF NON PATIENT END PLACEMENT. HE FEELS THE NON PATIENT END IS STRAIGHT. BUT HE HAS VERY POOR VISION. LOT # 3297042. CATALOG# 320119. DATE OF EVENT UNKNOWN. SAMPLE STATUS AWAITING SAMPLES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20707547 ↗MalfunctionNov 12, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
CONSUMER REPORTED, PRIOR TO INJECTION, CLEAR LIQUID CAME OUT OF THE PATIENT END AND WENT INTO HER PEN CAUSING THE INSULIN TO BECOME "CLOUDY". (B)(4) PEN NEEDLE AFFECTED STATED, AFTER HER INJECTION WITH A SECOND PEN NEEDLE FROM THE SAME BOX, SHE SAW A SMALL AMOUNT OF "BLOOD" IN HER INSULIN. NO INJURY TO REPORT. DID NOT SEEK MEDICAL INTERVENTION AND HER BLOOD SUGAR WAS NOT AFFECTED AFTER THE INJECTIONS. (B)(4) PEN NEEDLE AFFECTED. LOT: 4038407. CATALOG: 320119. DATE OF EVENT: 2024-17-10. SAMPLES: YES CL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20669758 ↗MalfunctionOct 30, 2024Failure to Deliver; Failure to Deliver
FRI 9/27/2024 5:26 PM. SUBJECT: NOVOLOG FLEXPEN COMPATIBILITY. HELLO, I HAD BEEN USING THE HUMALOG FLEXPEN WITH BD ULTRA-FINE PEN NEEDLES MINI 5MMX1G FOR SOME TIME, BUT MY INSURANCE COMPANY RECENTLY STOPPED COVERING HUMALOG AND SWITCHED ME TO THE NOVOLOG FLEXPEN. A COUPLE OF MONTHS AGO, I RAN OUT OF HUMALOG AND STARTED USING NOVOLOG, BUT I'VE NOTICED THAT IT'S MUCH HARDER TO INJECT, AS THE MECHANISM ISN'T SMOOTH. I CONTACTED NOVOLOG, AND THEY INFORMED ME THAT THEY DON'T TEST FOR NEEDLES AND RECOMMENDED I PURCHASE NOVOLOG NEEDLES (NDC #0169-1851-89, 32G, 6MM). HOWEVER, I'M UNSURE WHETHER MY INSURANCE WILL COVER THEM OR IF MY PHARMACY WILL BE ABLE TO PROVIDE THEM. MY QUESTION FOR YOU IS: DO YOU TEST YOUR BD NEEDLES WITH NOVOLOG PRODUCTS? I'VE USED BD NEEDLES FOR YEARS WITH VARIOUS INSULIN PRODUCTS AND NEVER ENCOUNTERED THIS ISSUE BEFORE. I WOULD APPRECIATE YOUR GUIDANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20567965 ↗MalfunctionOct 24, 2024Misconnection; Off-Label Use; Failure to Deliver; Misconnection; Off-Label Use; Failure to Deliver
CONSUMER REPORTED NO INSULIN FLOW WHEN TAKING INJECTION STATED, HE HAS TO PUSH DOWN REALLY HARD WHEN TAKING HIS INJECTION. STATED, HE IS CONCERNED THE PEN NEEDLE MAY BREAK WHEN TAKING INJECTION. DOES NOT PRIME PEN NEEDLES BEFORE INJECTION. LOT: 4051681 CATALOG: 320119 DATE OF EVENT: UNKNOWN SAMPLES: YES CL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20526769 ↗MalfunctionOct 10, 2024Failure to Deliver; Failure to Deliver
CONSUMER REPORTED FOUND 2 PEN NEEDLES FROM THIS SUPPLY THAT CLOGGED DURING THE FLOW CHECK. DC. LOT #4044088. CATALOG# 320119. DATE OF EVENT 09.16.2024 = 1 PEN NEEDLE - AWAITING SAMPLE. DATE OF EVENT UNKNOWN = 1 PEN NEEDLE DISCARD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20417770 ↗MalfunctionOct 10, 2024Failure to Deliver; Failure to Deliver
CONSUMER REPORTED FOUND 2 PEN NEEDLES FROM THIS SUPPLY THAT CLOGGED DURING THE FLOW CHECK. DC LOT #4044088 CATALOG# 320119 DATE OF EVENT 09.16.2024 = 1 PEN NEEDLE - AWAITING SAMPLE DATE OF EVENT UNKNOWN = 1 PEN NEEDLE DISCARD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20417747 ↗MalfunctionSep 30, 2024Failure to Deliver; Failure to Deliver
CONSUMER REPORTED, NO INSULIN FLOW WHEN TAKING INJECTION. STATED, SHE DOES NOT PRIME BEFORE TAKING INJECTION. LOT: 4016306. CATALOG: 320119. DATE OF EVENT: UNKNOWN. SAMPLES: YES, CL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20333063 ↗MalfunctionSep 26, 2024Failure to Deliver; Failure to Deliver
NEEDLE CLOG DURING PRIMING. CONSUMER STATED THAT THERE IS NO INSULIN FLOW. CONSUMER DOES NOT RE-USE. LOT #: 4030377. CATALOG #: 320119. DATE OF EVENT: UNKNOWN. SAMPLES AVAILABLE: SENDING MAIL KIT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20311504 ↗MalfunctionSep 13, 2024Misconnection; Misconnection
CONSUMER REPORTED FIRST TIME USING THE PEN NEEDLE SYSTEM WITH NOVO FIASP. HAVING 4 PEN NEEDLES THAT WOULD NOT ATTACH TO HIS PEN. CALLER HAS VISION ISSUES. INFORMED CALLER TO PLACE THE NON PATIENT END STRAIGHT ONTO PEN AND TURN UNTIL IT STOPS. CALLER WAS ABLE TO ATTACH A NEW PEN NEEDLE WITH MY DIRECTION. DC ITEM # 320119 LOT # 40110761 EVENT DATE AUGUST 21 2024 STATUS AWAITING SAMPLE DC.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20212592 ↗MalfunctionSep 13, 2024Failure to Deliver; Failure to Deliver
CONSUMER REPORTED, NO INSULIN FLOW WHEN TAKING INJECTION. 12 PEN NEEDLES AFFECTED. STATED, DOES NOT PRIME BEFORE ATTEMPTING INJECTION BECAUSE SHE DOES NOT WANT TO WASTE THE MEDICATION. STATED, ONCE, THE PATIENT END ON THE PEN NEEDLE WAS MISSING PRIOR TO INJECTION. LOT: 3304314, CATALOG: 320119, DATE OF EVENT: UNKNOWN, SAMPLES: NO CL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20212544 ↗MalfunctionAug 23, 2024Failure to Deliver; Failure to Deliver
CONSUMER REPORTED NEEDLE CLOG DURING INJECTION, SHE STATED THAT THERE IS NO INSULIN FLOW. CONSUMER DOES NOT RE-USE. LOT #: 3283936. CATALOG #: 320119. DATE OF EVENT: UNKNOWN. SAMPLES: DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20050082 ↗MalfunctionAug 23, 2024Failure to Deliver; Failure to Deliver
CONSUMER REPORTED WHEN INJECTING HER BASLAGAR INSULIN SHE RECEIVED HALF HER DOSE. THE PEN NEEDLE CLOGGED. PROVIDED CONSUMER LILLY PHONE NUMBER. FIRST TIME INJECTING WITH PENS. INFORMED PROPER PLACEMENT OF NON PATIENT END AND TO COMPLETE A FLOW CHECK. DC LOT # 0070045. CATALOG#320119. DATE OF EVENT JULY 28, 2024. SAMPLE STATUS DISCARD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20050075 ↗MalfunctionJul 24, 2024Packaging Problem; Packaging Problem
CONSUMER REPORTED FOUND APROX 20 PEN NEEDLES WITH THE TEAR TAB LOOSE ON THE NON PATIENT END OF NEEDLE. DISCARDED THE SAMPLES -DC LOT # 3297041 CATALOG# 320119 DATE OF EVENT UNKNOWN -APRIL 2024 SAMPLE STATUS DISCARD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19818089 ↗MalfunctionJul 24, 2024Failure to Deliver; Failure to Deliver
CONSUMER LEFT VOICEMAIL REGARDING MINI PEN NEEDLES. I RETURNED CONSUMER'S CALL AND SHE STATED THAT THERE IS NO INSULIN FLOW. CONSUMER DOES NOT RE-USE. CONSUMER DOES NOT PRIME. LOT #: 2159715, CATALOG #: 320119, DATE OF EVENT: UNKNOWN, SAMPLES: DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19817905 ↗MalfunctionJun 13, 2024Failure to Deliver; Failure to Deliver
CONSUMER REPORTED FINDING THE NEEDLES TO CLOG DURING INJECTION. DOES NOT ALWAYS COMPLETE A PROPER FLOW CHECK. INFORMED CALLER OF PROPER PLACEMENT OF NON PATIENT END. LOT # 2172143. CATALOG# 320119. DATE OF EVENT UNKNOWN . SAMPLE STATUS DISCARD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19533263 ↗MalfunctionMay 13, 2024Break; Break
CALLER REPORTED NEEDLE BREAK ON BEHALF OF HER GRANDMOTHER WHO IS THE CONSUMER. SHE STATED THAT HER GRANDMOTHER VISITED THE EMERGENCY ROOM DUE TO A FALL. WHILE AT THE EMERGENCY ROOM, THEY DID A CT SCAN OF HER ABDOMEN AND FOUND THAT THERE WAS A NEEDLE LODGED IN HER STOMACH. THE VISIT TO THE EMERGENCY ROOM WAS UNRELATED. SHE WAS ADVISED TO CONTACT HER PRIMARY CARE DOCTOR FOR ADVICE ON WHETHER OR NOT SHE WOULD NEED TO SURGICALLY REMOVE THE NEEDLE. CALLER STATED THAT THERE WAS NOT PAIN OR DISCOMFORT AND SHE DOES NOT KNOW WHEN THE NEEDLE BROKE OFF. LOT #: 3269032. CATALOG #: 320119. DATE OF EVENT: UNKNOWN. SAMPLES: DISCARDED.
Patient problems: Foreign Body In Patient
MAUDE report 19305672 ↗
FDA clearance and classification
Cleared through 510(k) K213478 on Jun 28, 2022 (decision: Substantially Equivalent). Submitted by Becton, Dickinson and Company as “BD Pen Needle”. The FDA review took 242 days. FDA record ↗
The FDA regulates product code FMI (needle, hypodermic, single lumen) under 21 CFR 880.5570 as a Class II device. 21 CFR 880.5570 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K213478. One submission often covers many models, so check the product and lot details in each record.
- Z-3180-2026 · Jul 26, 2023 · Open, ClassifiedEWO_02513 (Catalog number 320122) For Investigational Use Only Nano Un-Cannulated Demo Samples Quantity 100
Common questions about the BD Ultra-Fine 320119
Has the BD Ultra-Fine 320119 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00382903201198), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.
What problems have been reported with the BD Ultra-Fine 320119?
The FDA's MAUDE database lists 162 adverse event reports for this device, received between Oct 26, 2017 and Jul 17, 2025. In those reports, the most common specific device problems are “Failure to Deliver” (70), “Misconnection” (19) and “Break” (13); the most common patient problems are “Foreign Body In Patient” (4), “Hyperglycemia” (4) and “Needle Stick/Puncture” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the BD Ultra-Fine 320119?
None of the reports are classified as deaths, 8 as injuries and 154 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the BD Ultra-Fine 320119 FDA approved?
It was cleared through 510(k) K213478 on Jun 28, 2022. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the BD Ultra-Fine 320119, and is it still sold?
The FDA's device database lists Becton Dickinson Pen Limited as the company responsible for it (brand BD Ultra-Fine, model 320119). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Needle, hypodermic, single lumen devices
- BD Nano 2nd Gen 320550Becton Dickinson Pen Limited · Needle, hypodermic, single lumen1,892 reports
- BD SafetyGlide 305916Becton, Dickinson And Company · Needle, hypodermic, single lumen1,768 reports
- BD Nano Ultra-Fine 320122Becton Dickinson Pen Limited · Needle, hypodermic, single lumen566 reports
- N/A 305106Becton, Dickinson And Company · Needle, hypodermic, single lumen1 recall465 reports
- N/A 305110Becton, Dickinson And Company · Needle, hypodermic, single lumen418 reports
- BD Safe-Clip 328235Becton, Dickinson and Company Inc · Needle, hypodermic, single lumen268 reports
- BD Ultra-Fine 320109Becton Dickinson Pen Limited · Needle, hypodermic, single lumen208 reports
- BD AutoShield Duo 329515Becton, Dickinson And Company · Needle, hypodermic, single lumen200 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page