Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00382903497034), received between Oct 23, 2020 and Oct 23, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Oct 5, 2020Terminated

    Recent product complaints showing an unexpected cell population exhibiting dim CD3 expression that may result in inaccurate results gating.

    Z-0562-2021 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 23, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE WITH A BD MULTI-CHECK CD4 LOW CTRL 1X2.5ML US-IVD RESULTS SHOW EXTRA POPULATION IN LOW CONTROL. THERE WAS NO IMPACT TO PATIENT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTING EXTRA POPULATION PRESENT IN LOW CONTROL ONLY. NORMAL CONTROL LOOKS AS EXPECTED. REQUIRES A GATE ADJUSTMENT TO CORRECT FOR THIS. CUSTOMER HAS FOLLOWED UP THEY ORDERED 2 VIALS OF CATALOG 349703 LOT 1020 THAT SHOW THIS EXTRA POPULATION. THEY HAVE CONFIRMED THIS DID NOT AFFECT PATIENT DATA AS THEY NOTICED THE POPULATION AND WERE ABLE TO GATE THEM OUT. QUALITY IS AWARE OF THIS ISSUE AND IS INVESTIGATING. ADDITIONALLY, ON 2020-10-07 THE FSE PROVIDED THE FOLLOWING ADDITIONAL INFORMATION: CUSTOMER HAS FOLLOWED UP THEY ORDERED 2 VIALS OF CATALOG 349703 LOT 1020 THAT SHOW THIS EXTRA POPULATION. THEY HAVE CONFIRMED THIS DID NOT AFFECT PATIENT DATA AS THEY NOTICED THE POPULATION AND WERE ABLE TO GATE THEM OUT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10726257 ↗

Other Mixture, hematology quality control devices