Skip to content

BD Multi-Check recallZ-0562-2021 · Becton, Dickinson and Company, BD Biosciences · initiated Oct 5, 2020

Class II Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Becton, Dickinson and Company, BD Biosciences or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Recent product complaints showing an unexpected cell population exhibiting dim CD3 expression that may result in inaccurate results gating.

FDA root cause category: Process control

What the company is doing

On 10/13/2020, the firm notified it customer via telephone and sent an "URGENT MEDICAL DEVICE RECALL" Notification via email and/or FedEx to inform them that complaints have been report regarding their Low Control product which shows an unexpected cell population exhibiting dim CD3 expression that may result in inaccurate gating with the potential for delayed patient management. Customer are instructed to: 1. Immediately review their inventory for the specific Catalog and lot numbers listed. Destroy all product subject to this recall following their institutions process for destruction. 2. Share this recall notification with all users of the product within their facility to ensure that they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the Recalling Firm contact noted on the form. Even if customer do not have any of the affected lots in their inventory, they are asked to complete the Customer Recall Response Form indicating they have zero (0) quantity and return as indicated. Return Customer Response Form via email [email protected] or fax 312-949-0371 Actions the Recalling Firm is taking: An investigation has been initiated to identify root cause and corrective actions to prevent recurrence. An questions, contact Customer/Technical Support at 877-232-8995 (prompt 2,2)

Recall details

Recall number
Z-0562-2021
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 5, 2020
Recalling firm
Becton, Dickinson and Company, BD Biosciences
Quantity
619 vials
Product code
JPK
FDA submissions
K150815
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, CA, CO, CT, DE, FL, GA, IL, IN, KY, MA, MD, MI, NC, NJ, NV, NY, OH, OK, PA, PR, RI, SD, TN, TX, and VA. The countries of Canada, Belgium, Korea, India, Ecuador, Japan, Costa Rica, Australia, Vietnam, Pakistan, Singapore, Thailand, and Peru.

Product description

BD" Multi-Check CD4 Low Control 2.5 mL IVD - Product Usage: intended as a complete process control for immunophenotyping by flow cytometry.

Code information (models, lots, serial numbers)

Lot Number: BM1020L  BD Multi-Check CD4 Low Control (US Distribution) Catalog Number UDI 349703 00382903497034 US 349704 00382903497041 US 349705 00382903497058 US  BD" Multi-Check CD4 Low Control Whole Blood Control for Lymphocyte Subset Enumeration OUS Distribution Catalog Number UDI 340914 00382903409143 OUS only 340915 00382903409150 OUS only 340916 00382903409167 OUS only

Devices named in this recall

6 devices in the FDA device registry whose identifier appears in the recall's code information.