Skip to content

BD Multi-Check 349704Recalls & reported problems

Becton, Dickinson And Company · BD Multi-Check™ CD4 Low Control · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00382903497041), received between Jul 20, 2023 and Jul 20, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Oct 5, 2020Terminated

    Recent product complaints showing an unexpected cell population exhibiting dim CD3 expression that may result in inaccurate results gating.

    Z-0562-2021 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 20, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE WITH THE BD MULTI-CHECK CONTROL THERE WAS INCORRECT LABEL INFORMATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE THE INITIAL REPORTER: IT WAS REPORTED BY THE CUSTOMER THAT WHEN THE R&D MULTI-CHECK CONTROLS (349701 AND 349704) COME DIRECTLY FROM R&D, THE 2D BARCODE CAN BE READ BY THE FACSDUET. BUT WHEN THE R&D CONTROLS ARE REPACKAGED FOR BDB THE QR BARCODE ON THE TUBE IS NOT COMPATIBLE WITH THE DUET.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17359067 ↗

Other Mixture, hematology quality control devices

All Mixture, hematology quality control devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page