BD MAX™ Enteric Parasite Panel 442960Recalls & reported problems

Becton, Dickinson And Company · Kit BD MAX Enteric Parasite Panel · Class 2

Quick summary

This BD MAX Enteric Parasite Panel is a lab test kit that detects gut parasites such as Giardia and Cryptosporidium in stool samples. The 70 reports mostly describe results that came back positive when later reference lab testing found no parasite, with no deaths and no injuries. Reports rose from 8 in the prior 12 months to 20 in the most recent 12 months. One closed recall, started July 2019, involved reagent extraction tube foil bags that may not have been sealed properly.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
70
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 70 adverse event reports that list this exact device (DI 00382904429607), received between Aug 4, 2021 and Sep 3, 2026 — classified as 70 malfunctions.

The most frequently reported specific device problem is “False Positive Result” (62 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 3 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 5 reportsJan 2026: 3 reportsFeb 2026: 9 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 17, 2019Terminated

    The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.

    Z-2329-2019 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • False Positive Result62
  • False Negative Result4
  • Incorrect, Inadequate or Imprecise Result or Readings4

Patient problems

  • No Clinical Signs, Symptoms or Conditions70

Recent adverse event reports

The 25 most recent of 70 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 3, 2026False Positive Result

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50589667 ↗
  • MalfunctionMar 3, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A FALSE POSITIVE GIARDIA PATIENT RESULT WAS OBTAINED. SAMPLE WAS SENT TO A REFERENCE LAB FOR CONFIRMATION TESTING, AND WAS GIARDIA NEGATIVE USING AN UNSPECIFIED ASSAY. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24498658 ↗
  • MalfunctionMar 3, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A FALSE POSITIVE GIARDIA PATIENT RESULT WAS OBTAINED. SAMPLE WAS SENT TO A REFERENCE LAB FOR CONFIRMATION TESTING, AND WAS GIARDIA NEGATIVE USING AN UNSPECIFIED ASSAY. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24498656 ↗
  • MalfunctionFeb 24, 2026False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A FALSE POSITIVE CRYPTOSPORIDIUM PATIENT RESULT WAS OBTAINED. SAMPLE WAS SENT TO A REFERENCE LABORATORY AND WAS CRYPTOSPORIDIUM NEGATIVE. SAMPLE WAS ALSO TESTED WITH QUICKCHEK ASSAY AND WAS ALSO CRYPTOSPORIDIUM NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24428688 ↗
  • MalfunctionFeb 23, 2026False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A FALSE POSITIVE CRYPTOSPORIDIUM PATIENT RESULT WAS OBTAINED. SAMPLE WAS SENT TO A REFERENCE LABORATORY AND WAS CRYPTOSPORIDIUM NEGATIVE. SAMPLE WAS ALSO TESTED WITH QUICKCHEK ASSAY AND WAS ALSO CRYPTOSPORIDIUM NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24415544 ↗
  • MalfunctionFeb 23, 2026False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A FALSE POSITIVE CRYPTOSPORIDIUM PATIENT RESULT WAS OBTAINED. SAMPLE WAS SENT TO A REFERENCE LABORATORY AND WAS CRYPTOSPORIDIUM NEGATIVE. SAMPLE WAS ALSO TESTED WITH QUICKCHEK ASSAY AND WAS ALSO CRYPTOSPORIDIUM NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24415543 ↗
  • MalfunctionFeb 23, 2026False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A FALSE POSITIVE CRYPTOSPORIDIUM PATIENT RESULT WAS OBTAINED. SAMPLE WAS SENT TO A REFERENCE LABORATORY AND WAS CRYPTOSPORIDIUM NEGATIVE. SAMPLE WAS ALSO TESTED WITH QUICKCHEK ASSAY AND WAS ALSO CRYPTOSPORIDIUM NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24415542 ↗
  • MalfunctionFeb 23, 2026False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A FALSE POSITIVE CRYPTOSPORIDIUM PATIENT RESULT WAS OBTAINED. SAMPLE WAS SENT TO A REFERENCE LABORATORY AND WAS CRYPTOSPORIDIUM NEGATIVE. SAMPLE WAS ALSO TESTED WITH QUICKCHEK ASSAY AND WAS ALSO CRYPTOSPORIDIUM NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24415541 ↗
  • MalfunctionFeb 23, 2026False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A FALSE POSITIVE CRYPTOSPORIDIUM PATIENT RESULT WAS OBTAINED. SAMPLE WAS SENT TO A REFERENCE LABORATORY AND WAS CRYPTOSPORIDIUM NEGATIVE. SAMPLE WAS ALSO TESTED WITH QUICKCHEK ASSAY AND WAS ALSO CRYPTOSPORIDIUM NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24415538 ↗
  • MalfunctionFeb 23, 2026False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A FALSE POSITIVE CRYPTOSPORIDIUM PATIENT RESULT WAS OBTAINED. SAMPLE WAS SENT TO A REFERENCE LABORATORY AND WAS CRYPTOSPORIDIUM NEGATIVE. SAMPLE WAS ALSO TESTED WITH QUICKCHEK ASSAY AND WAS ALSO CRYPTOSPORIDIUM NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24415537 ↗
  • MalfunctionFeb 23, 2026False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A FALSE POSITIVE CRYPTOSPORIDIUM PATIENT RESULT WAS OBTAINED. SAMPLE WAS SENT TO A REFERENCE LABORATORY AND WAS CRYPTOSPORIDIUM NEGATIVE. SAMPLE WAS ALSO TESTED WITH QUICKCHEK ASSAY AND WAS ALSO CRYPTOSPORIDIUM NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24415534 ↗
  • MalfunctionFeb 23, 2026False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A FALSE POSITIVE CRYPTOSPORIDIUM PATIENT RESULT WAS OBTAINED. SAMPLE WAS SENT TO A REFERENCE LABORATORY AND WAS CRYPTOSPORIDIUM NEGATIVE. SAMPLE WAS ALSO TESTED WITH QUICKCHEK ASSAY AND WAS ALSO CRYPTOSPORIDIUM NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24415525 ↗
  • MalfunctionJan 6, 2026False Positive Result

    REPORT 3 OF 3: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A FALSE POSITIVE GIARDIA PATIENT RESULT WAS OBTAINED. TEST WAS REPEATED WITH A NEW REAGENT KIT LOT AND WAS GIARDIA POSITIVE. TEST WAS THEN REPEATED WITH THE SAME REAGENT KIT LOT AND WAS GIARDIA NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23979887 ↗
  • MalfunctionJan 6, 2026False Positive Result

    REPORT 2 OF 3: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A FALSE POSITIVE GIARDIA PATIENT RESULT WAS OBTAINED. TEST WAS REPEATED WITH A NEW REAGENT KIT LOT AND WAS GIARDIA POSITIVE. TEST WAS THEN REPEATED WITH THE SAME REAGENT KIT LOT AND WAS GIARDIA NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23979866 ↗
  • MalfunctionJan 6, 2026False Positive Result

    REPORT 1 OF 3: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A FALSE POSITIVE GIARDIA PATIENT RESULT WAS OBTAINED. TEST WAS REPEATED WITH A NEW REAGENT KIT LOT AND WAS GIARDIA POSITIVE. TEST WAS THEN REPEATED WITH THE SAME REAGENT KIT LOT AND WAS GIARDIA NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23979865 ↗
  • MalfunctionDec 17, 2025False Positive Result; False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A PATIENT SAMPLE TESTED DUAL POSITIVE FOR GIARDIA AND CRYPTOSPORIDIUM. SAMPLE WAS REPEATED ON MAX AND WAS NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23828871 ↗
  • MalfunctionDec 17, 2025False Positive Result; False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A PATIENT SAMPLE TESTED DUAL POSITIVE FOR GIARDIA AND CRYPTOSPORIDIUM. SAMPLE WAS REPEATED ON MAX AND WAS NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23828508 ↗
  • MalfunctionDec 17, 2025False Positive Result; False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A PATIENT SAMPLE TESTED DUAL POSITIVE FOR GIARDIA AND CRYPTOSPORIDIUM. SAMPLE WAS REPEATED ON MAX AND WAS NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23828497 ↗
  • MalfunctionDec 17, 2025False Positive Result; False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A PATIENT SAMPLE TESTED DUAL POSITIVE FOR GIARDIA AND CRYPTOSPORIDIUM. SAMPLE WAS REPEATED ON MAX AND WAS NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23828491 ↗
  • MalfunctionDec 17, 2025False Positive Result; False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A PATIENT SAMPLE TESTED DUAL POSITIVE FOR GIARDIA AND CRYPTOSPORIDIUM. SAMPLE WAS REPEATED ON MAX AND WAS NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23828490 ↗
  • MalfunctionOct 21, 2025False Positive Result; False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC PARASITE PANEL, A FALSE POSITIVE CRYPTOSPORIDIUM PATIENT RESULT WAS OBTAINED. TEST WAS REPEATED ON MAX AND WAS NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23350918 ↗
  • MalfunctionSep 3, 2025False Positive Result; False Positive Result

    IT WAS REPORTED WHILE USING THE BD MAX¿ ENTERIC PARASITE PANEL, THERE WAS A FALSE POSITIVE RESULT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22951755 ↗
  • MalfunctionSep 3, 2025False Positive Result; False Positive Result

    IT WAS REPORTED WHILE USING THE BD MAX¿ ENTERIC PARASITE PANEL, THERE WAS A FALSE POSITIVE RESULT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22951741 ↗
  • MalfunctionSep 3, 2025False Positive Result; False Positive Result

    IT WAS REPORTED WHILE USING THE BD MAX¿ ENTERIC PARASITE PANEL, THERE WAS A FALSE POSITIVE RESULT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22951736 ↗
  • MalfunctionApr 24, 2025False Positive Result; False Positive Result

    IT WAS REPORTED THAT DURING USE OF THE BD MAX¿ ENTERIC PARASITE PANEL, A GIARDIA LAMBLIA FALSE POSITIVE PATIENT RESULT WITH AN NOISY PCR CURVE WAS OBTAINED. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21899820 ↗

Other Gastrointestinal pathogen panel multiplex nucleic acid-based assay system devices

All Gastrointestinal pathogen panel multiplex nucleic acid-based assay system devices →