BD MAX™ Enteric Bacterial Panel 442963Recalls & reported problems

Becton, Dickinson And Company · Kit BD MAX Enteric Bacterial Panel · Class 2

Quick summary

This BD MAX enteric bacterial panel is a gastrointestinal pathogen assay kit made by BD. The 97 reports mostly describe false positive or false negative results, with 3 injuries and no deaths. It has 2 recalls, both closed: one for foil bags that may not have been sealed properly and one for false negative results for ETEC and Plesiomonas. Reports fell from 23 in the previous 12 months to 10 in the last 12.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
102
Deaths reported
0
Injuries reported
3
Recalls naming it
2

Summary

The FDA's MAUDE database contains 102 adverse event reports that list this exact device (DI 00382904429638), received between Sep 18, 2020 and Sep 29, 2026 — classified as 3 injuries, 99 malfunctions.

The most frequently reported specific device problem is “False Positive Result” (69 reports).

2 FDA recall records name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 8 reportsJan 2024: 0 reportsFeb 2024: 3 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 10 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 10 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 7 reportsJul 2025: 3 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 5 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 4 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 17, 2019Terminated

    The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.

    Z-2330-2019 · Root cause: Process control · FDA record ↗

  2. Jan 14, 2019Terminated

    False negative results with the BD MAX Extended Enteric Bacterial Panel when testing specifically for Enterotoxigenic Escherichia coli (ETEC) and Plesiomonas shigelloides targets of the BD MAX Extended Enteric Bacterial Panel assay.

    Z-0981-2019 · Root cause: Nonconforming Material/Component · FDA record ↗

Most reported problems

Device problems

  • False Positive Result69
  • False Negative Result17
  • Incorrect, Inadequate or Imprecise Result or Readings15
  • Missing Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions101
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 102 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026False Positive Result

    It was reported that during use of BD MAX¿ Enteric Bacterial Panel, a False Positive Campylobacter patient result was obtained.There was no report of patient impact.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50931720 ↗
  • MalfunctionSep 18, 2026False Positive Result

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820934 ↗
  • MalfunctionSep 15, 2026False Negative Result

    Report 2 of 2:It was reported that during use of BD MAX¿ Bacterial Panel, a False Negative patient result was obtained.Specimen was retested using same assay and again produced a Negative Salmonella result.The specimen was also tested by culture and Salmonella was recovered.Repeat culture of specimen again yielded Salmonella.There was no report of patient impact.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50773744 ↗
  • MalfunctionSep 15, 2026False Negative Result

    Report 1 of 2:It was reported that during use of BD MAX¿ Bacterial Panel, a False Negative patient result was obtained.Specimen was retested using same assay and again produced a Negative Salmonella result.The specimen was also tested by culture and Salmonella was recovered.Repeat culture of specimen again yielded Salmonella.There was no report of patient impact.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50773382 ↗
  • MalfunctionAug 11, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50207422 ↗
  • MalfunctionMay 6, 2026False Negative Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ BACTERIAL PANEL, A FALSE NEGATIVE SALMONELLA PATIENT RESULT WAS OBTAINED. THE BD MAX REPORTED A NEGATIVE RESULT FOR SALMONELLA SPP. TWICE, ALTHOUGH SALMONELLA GREW IN CULTURE. AGGLUTINATION IDENTIFIED SEROGROUP D, AND FURTHER DIAGNOSTIC TESTING CONFIRMED THE ISOLATE AS SALMONELLA ENTERITIDIS. THERE WAS NO REPORT OF PATIENT IMPACT. THIS IS REPORT 2 OF 2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25104820 ↗
  • MalfunctionMay 6, 2026False Negative Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ BACTERIAL PANEL, A FALSE NEGATIVE SALMONELLA PATIENT RESULT WAS OBTAINED. THE BD MAX REPORTED A NEGATIVE RESULT FOR SALMONELLA SPP. TWICE, ALTHOUGH SALMONELLA GREW IN CULTURE. AGGLUTINATION IDENTIFIED SEROGROUP D, AND FURTHER DIAGNOSTIC TESTING CONFIRMED THE ISOLATE AS SALMONELLA ENTERITIDIS. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25096683 ↗
  • MalfunctionFeb 23, 2026False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC BACTERIAL PANEL, A FALSE POSITIVE CAMPYLOBACTER PATIENT RESULT WITH AN UNUSUAL PCR CURVE WAS OBTAINED. TEST WAS REPEATED AND WAS CAMPYLOBACTER NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24416042 ↗
  • MalfunctionFeb 23, 2026False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC BACTERIAL PANEL, A FALSE POSITIVE CAMPYLOBACTER PATIENT RESULT WITH AN UNUSUAL PCR CURVE WAS OBTAINED. TEST WAS REPEATED AND WAS CAMPYLOBACTER NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24416027 ↗
  • MalfunctionFeb 23, 2026False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC BACTERIAL PANEL, A FALSE POSITIVE CAMPYLOBACTER PATIENT RESULT WITH AN UNUSUAL PCR CURVE WAS OBTAINED. TEST WAS REPEATED AND WAS CAMPYLOBACTER NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24416024 ↗
  • MalfunctionDec 3, 2025False Negative Result; False Negative Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC BACTERIAL PANEL, A FALSE NEGATIVE SALMONELLA PATIENT RESULT WAS OBTAINED. A HEKTOEN PLATE WAS SEEDED IN PARALLEL WITH THE BD MAX TEST. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23707797 ↗
  • MalfunctionDec 3, 2025False Negative Result; False Negative Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC BACTERIAL PANEL, A FALSE NEGATIVE SALMONELLA PATIENT RESULT WAS OBTAINED. A HEKTOEN PLATE WAS SEEDED IN PARALLEL WITH THE BD MAX TEST. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23707777 ↗
  • MalfunctionDec 3, 2025False Negative Result; False Negative Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC BACTERIAL PANEL, A FALSE NEGATIVE SALMONELLA PATIENT RESULT WAS OBTAINED. A HEKTOEN PLATE WAS SEEDED IN PARALLEL WITH THE BD MAX TEST. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23707772 ↗
  • MalfunctionDec 3, 2025False Negative Result; False Negative Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC BACTERIAL PANEL, A FALSE NEGATIVE SALMONELLA PATIENT RESULT WAS OBTAINED. A HEKTOEN PLATE WAS SEEDED IN PARALLEL WITH THE BD MAX TEST. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23707555 ↗
  • MalfunctionDec 3, 2025False Negative Result; False Negative Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC BACTERIAL PANEL, A FALSE NEGATIVE SALMONELLA PATIENT RESULT WAS OBTAINED. THE FIRST MAX TEST WAS SALMONELLA NEGATIVE, AND THE REPEAT MAX TEST WAS SALMONELLA POSITIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23706462 ↗
  • MalfunctionAug 28, 2025False Positive Result; False Positive Result

    IT WAS REPORTED WHILE USING BD MAX¿ ENTERIC BACTERIAL PANEL, THERE WAS A FALSE POSITIVE RESULT. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22912622 ↗
  • MalfunctionJul 31, 2025False Positive Result; False Positive Result

    REPORT 1 OF 3: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC BACTERIAL PANEL, A CAMPYLOBACTER FALSE POSITIVE PATIENT RESULT WAS OBTAINED. TEST WAS REPEATED TWICE ON BD MAX¿ AND WAS NEGATIVE, AND COULD NOT BE SUBCULTURED. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22677997 ↗
  • MalfunctionJul 31, 2025False Positive Result; False Positive Result

    REPORT 3 OF 3: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC BACTERIAL PANEL, A CAMPYLOBACTER FALSE POSITIVE PATIENT RESULT WAS OBTAINED. TEST WAS REPEATED TWICE ON BD MAX¿ AND WAS NEGATIVE AND COULD NOT BE SUBCULTURED. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22677993 ↗
  • MalfunctionJul 31, 2025False Positive Result; False Positive Result

    REPORT 2 OF 3: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC BACTERIAL PANEL, A CAMPYLOBACTER FALSE POSITIVE PATIENT RESULT WAS OBTAINED. TEST WAS REPEATED TWICE ON BD MAX¿ AND WAS NEGATIVE AND COULD NOT BE SUBCULTURED. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22677992 ↗
  • MalfunctionJun 18, 2025False Positive Result; False Positive Result

    REPORT 2 OF 6: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC BACTERIAL PANEL, A FALSE POSITIVE SALMONELLA PATIENT RESULT WAS OBTAINED. MAX TEST WAS REPEATED AND WAS SALMONELLA NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22287249 ↗
  • MalfunctionJun 18, 2025False Positive Result; False Positive Result

    REPORT 5 OF 6: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC BACTERIAL PANEL, A FALSE POSITIVE SALMONELLA PATIENT RESULT WAS OBTAINED. MAX TEST WAS REPEATED AND WAS SALMONELLA NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22287248 ↗
  • MalfunctionJun 18, 2025False Positive Result; False Positive Result

    REPORT 6 OF 6: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC BACTERIAL PANEL, A FALSE POSITIVE SALMONELLA PATIENT RESULT WAS OBTAINED. MAX TEST WAS REPEATED AND WAS SALMONELLA NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22287246 ↗
  • MalfunctionJun 18, 2025False Positive Result; False Positive Result

    REPORT 4 OF 6: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC BACTERIAL PANEL, A FALSE POSITIVE SALMONELLA PATIENT RESULT WAS OBTAINED. MAX TEST WAS REPEATED AND WAS SALMONELLA NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22287243 ↗
  • MalfunctionJun 18, 2025False Positive Result; False Positive Result

    REPORT 1 OF 6: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC BACTERIAL PANEL, A FALSE POSITIVE SALMONELLA PATIENT RESULT WAS OBTAINED. MAX TEST WAS REPEATED AND WAS SALMONELLA NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22287242 ↗
  • MalfunctionJun 18, 2025False Positive Result; False Positive Result

    REPORT 3 OF 6: IT WAS REPORTED THAT DURING USE OF BD MAX¿ ENTERIC BACTERIAL PANEL, A FALSE POSITIVE SALMONELLA PATIENT RESULT WAS OBTAINED. MAX TEST WAS REPEATED AND WAS SALMONELLA NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22287241 ↗

Other Gastrointestinal pathogen panel multiplex nucleic acid-based assay system devices

All Gastrointestinal pathogen panel multiplex nucleic acid-based assay system devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.