BD COR™ MX 443989Recalls & reported problems
Becton, Dickinson And Company · BD COR™ MX · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00382904439897), received between Mar 25, 2025 and Mar 13, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Oct 31, 2024Open, Classified
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Z-0588-2025 · Root cause: Software design · FDA record ↗
Most reported problems
Device problems
- False Positive Result4
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 13, 2026False Positive Result
IT WAS REPORTED AFTER USING BD COR MX INSTRUMENT, THERE WAS ONE FALSE POSITIVE RESULT ON PATIENT SAMPLE. THE INSTRUMENT REPORTED POSITIVE FOR CHLAMYDIA, BUT THE SAMPLE WAS RETESTED THE NEXT DAY WITH A NEGATIVE RESULT. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24597677 ↗MalfunctionJun 10, 2025False Positive Result; False Positive Result
IT WAS REPORTED DURING USE OF BD COR MX INSTRUMENT THERE WAS A FALSE POSITIVE RESULT FOR ALL MM2 TARGETS ON THE COR VAGINAL PANEL ASSAY. CONFIRMATORY TESTING WAS PERFORMED ON BD MAX INSTRUMENT AND INDICATED THAT ONLY ONE TARGET WAS POSITIVE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22181467 ↗MalfunctionApr 16, 2025False Positive Result; False Positive Result
IT WAS REPORTED THAT DURING USE OF THE BD COR¿ MX, 8 SAMPLES WERE FALSE POSITIVE FOR BOTH CHLAMYDIA TRACHOMATIS AND NEISSERIA GONORRHOEAE. AFTER INSPECTING REAGENTS AND INSTRUMENT, A "BLOODY SLIME STRING" FROM A SAMPLE WAS FOUND DRAGGED OVER PROCESS AREA AND CONTAMINATED SAMPLES. CLEANING AND ENVIRONMENTAL TESTING WERE PERFORMED. SAMPLES WERE RERUN AND WERE NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21848839 ↗MalfunctionMar 25, 2025False Positive Result; False Positive Result
IT WAS REPORTED AFTER USING BD COR MX INSTRUMENT, THERE WAS 1 FALSE POSITIVE RESULT FOR THE TRICHOMONAS VAGINALIS PANEL. CONFIRMATORY TESTING WAS PERFORMED ON THE BD MAX, WHERE THE RESULT WAS NEGATIVE. AFTER OBSERVING THE BD COR GRAPHS, THE CUSTOMER DISCOVERED THAT THE SAMPLE HAD JUST CROSSED THE THRESHOLD TO BE CALLED POSITIVE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21691589 ↗
Other DNA-REAGENTS, NEISSERIA devices
- Aptima Combo 2® KIT APTIMA COMBO 2 TIGRIS-REF (IVD)-CEHologic, Inc. · DNA-REAGENTS, NEISSERIA39 reports
- BD CTGCTV2 for BD MAX™ System 443904Becton, Dickinson And Company · DNA-REAGENTS, NEISSERIA35 reports
- BD ProbeTec™ Chlamydia trachomatis (CT) Qx Amplified DNA Assay 441126Becton, Dickinson And Company · DNA-REAGENTS, NEISSERIA5 reports
- BD Viper™ System 441091Becton, Dickinson And Company · DNA-REAGENTS, NEISSERIA3 reports
- Aptima Combo 2® KIT,APTIMACOMBO2,100T,REF,CEIVD-UPDATEDHologic, Inc. · DNA-REAGENTS, NEISSERIA3 reports
- BD Viper™ System Remanufactured 442843Becton, Dickinson And Company · DNA-REAGENTS, NEISSERIA2 reports
- Control Set for the BD ProbeTec™ (CT/GC) 441125Becton, Dickinson And Company · DNA-REAGENTS, NEISSERIA2 reports
- Aptima Combo 2® KIT, BX 1 (REFRIG) AC2, 100T, AS, IVDHologic, Inc. · DNA-REAGENTS, NEISSERIA2 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.