Aptima Combo 2® KIT, BX 1 (REFRIG) AC2, 100T, AS, IVDRecalls & reported problems
Hologic, Inc. · Aptima Combo 2® Assay, 100 test, box 1 of 2 (2°C to 8°C box). The Aptima Combo 2® Assay is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection and differentiation of ribosomal RNA (rRNA) from Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) to aid in the diagnosis of chlamydial and/or gonococcal disease using the Panther® System as specified. · Class 2
All models of Aptima Combo 2 (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 15420045505735), received between Sep 12, 2017 and Jan 9, 2020 — classified as 2 injuries.
The most frequently reported specific device problem is “Contamination” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Contamination1
- Incorrect, Inadequate or Imprecise Result or Readings1
- No Apparent Adverse Event1
Patient problems
- No Code Available1
- Unspecified Infection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 9, 2020No Apparent Adverse Event; No Apparent Adverse Event
FINAL REPORT: A (B)(6) PREVALENCE STUDY WAS INITIATED BY THE (B)(6) IN CONJUNCTION WITH THE (B)(6). ROUTINE SAMPLES THAT TESTED [?]EQUIVOCAL' FOR (B)(6) BY AC2 WERE RE-TESTED BY THE (B)(6) WHICH USES A DIFFERENT DETECTION TARGET (16S RRNA). SAMPLES THAT TESTED ACT (B)(6) WERE FORWARDED TO NMLFOR SEQUENCING TO EXAMINE IF MUTATION(S) IN THE 23S RRNA REGION PRODUCED (B)(6) VARIANT STRAINS. NML IDENTIFIED THREE (B)(6) VARIANT SAMPLES IN OVER 100,000 APTIMA COMBO2 SAMPLES SCREENED; 2 SAMPLES CARRIED THE C1514T MUTATION AND 1 SAMPLE CARRIED THE G1523A MUTATION. THESE MUTATIONS WERE PREVIOUSLY IDENTIFIED IN EUROPE AT VERY LOW PREVALENCE (ROBERTS ET AL., EUROSURVEILLANCE, 2019). THE NUMBER OF AC2 SAMPLES SCREENED WAS NEEDED TO ESTIMATE THE PREVALENCE OF THESE 3 (B)(6) VARIANTS IN (B)(6). DUE TO THE DIFFICULTY EXTRACTING THE NUMBERS FROM EACH LABORATORY'S INFORMATION SYSTEM, HOLOGIC USED CUSTOMER ORDERING DATA (DURING THE (B)(6) SURVEILLANCE PERIOD (JUNE 2019 TO OCTOBER 2019)) TO DETERMINE A VOLUME OF APPROXIMATELY 115,000 SCREENED AC2 SAMPLES. THE ESTIMATED PREVALENCE OF (B)(6) VARIANTS IDENTIFIED THUS FAR IN (B)(6) IS 0.0026%. THIS PREVALENCE RATE IS SIMILAR TO THE 0.003% PREVALENCE SEEN
Patient problems: Unspecified Infection
MAUDE report 9572514 ↗InjurySep 12, 2017Contamination; Incorrect, Inadequate or Imprecise Result or Readings
(B)(6) REPORTED THAT THEY HAD ONE AC2 ASSAY RUN WITH MORE POSITIVES THAN NORMAL (GC POSITIVE RATE WAS 1%). THE RUN (CONSISTING OF BOTH SWAB AND URINE SAMPLES) DONE ON THE PANTHER ON (B)(6) 2017 WAS A VALID RUN AND THE LABORATORY SAID THAT THEY DIDN'T IDENTIFY A PATTERN TO THE RUN. THE LABORATORY RELEASED THE RESULTS. A PHYSICIAN QUESTIONED FOUR OF THE RESULTS FOR SWAB AND URINE SAMPLES AND SENT FOR NEW PATIENT COLLECTIONS FOR THESE FOUR QUESTIONABLE RESULTS. ALL FOUR NEWLY COLLECTED SAMPLES WERE TESTED AS NEGATIVE. WHEN LABORATORY HEARD FROM THE PHYSICIAN, THEY CONTACTED HOLOGIC. HOLOGIC REVIEWED THE INSTRUMENT LOGS FOR THE RUN WITH THE HIGHER POSITIVITY RATE WHICH SHOWED NO INSTRUMENT OR REAGENT PREPARATION RELATED ISSUE. THERE WAS NO SIGN OF GROSS CONTAMINATION. HOLOGIC REQUESTED THE SAMPLES BACK (BOTH THE INITIAL SAMPLE WHICH TESTED POSITIVE AND THE SAMPLE TAKEN FROM THE RE-TEST) AND THEY WERE NOT AVAILABLE. HOLOGIC COULD NOT CONFIRM THE INITIAL OR SECONDARY SAMPLE TEST RESULTS. AN INCORRECT RESULT COULD BE DUE TO A KIT OR SAMPLE MISHANDLING ISSUE OR THE SAMPLES COULD BE LOW TITER SAMPLES. HOLOGIC RECOMMENDED CUSTOMER TO DISCARD ALL OPEN KITS AND CLEAN THE LAB. HOLOGIC INVESTIGA
Patient problems: No Code Available
MAUDE report 6858600 ↗
Other DNA-REAGENTS, NEISSERIA devices
- BD COR™ MX 443989Becton, Dickinson And Company · DNA-REAGENTS, NEISSERIA1 recall4 reports
- Aptima Combo 2® KIT APTIMA COMBO 2 TIGRIS-REF (IVD)-CEHologic, Inc. · DNA-REAGENTS, NEISSERIA39 reports
- BD CTGCTV2 for BD MAX™ System 443904Becton, Dickinson And Company · DNA-REAGENTS, NEISSERIA35 reports
- BD ProbeTec™ Chlamydia trachomatis (CT) Qx Amplified DNA Assay 441126Becton, Dickinson And Company · DNA-REAGENTS, NEISSERIA5 reports
- Aptima Combo 2® KIT,APTIMACOMBO2,100T,REF,CEIVD-UPDATEDHologic, Inc. · DNA-REAGENTS, NEISSERIA3 reports
- BD Viper™ System 441091Becton, Dickinson And Company · DNA-REAGENTS, NEISSERIA3 reports
- Aptima® APTIMA VAGINAL SWAB COLLECTION KITHologic, Inc. · DNA-REAGENTS, NEISSERIA2 reports
- BD Viper™ System Remanufactured 442843Becton, Dickinson And Company · DNA-REAGENTS, NEISSERIA2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page