Aptima Combo 2® KIT,APTIMACOMBO2,100T,REF,CEIVD-UPDATEDRecalls & reported problems

Hologic, Inc. · Aptima Combo 2® assay, 100 test, box 1 of 2 (2°C to 8°C box). The Aptima Combo 2® assay is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection and differentiation of ribosomal RNA (rRNA) from Chlamydia trachomatis (CT) and/or Neisseria gonorrhoeae (GC) to aid in the diagnosis of chlamydial and/or gonococcal disease using the Panther® system as specified. · Class 2

All models of Aptima Combo 2 (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 15420045512894), received between Oct 21, 2022 and Oct 17, 2025 — classified as 3 injuries.

The most frequently reported specific device problem is “Off-Label Use” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Off-Label Use2
  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 17, 2025Off-Label Use; Off-Label Use

    ON (B)(6) 2025, A CUSTOMER REPORTED A DISCREPANT RESULT USING THE APTIMA COMBO 2 (AC2) ASSAY ON THEIR PANTHER INSTRUMENT. ON (B)(6) 2025, SAMPLE ID: (B)(6) RESULTED AS GC POSITIVE, CT NEGATIVE WITH MASTER LOT 922619 ON WORKLIST (B)(4). CUSTOMER REPORTED THEY HAVE MADE IT THEIR LAB POLICY TO RETEST ANY POSITIVE RESULT WITH AN RLU VALUE BELOW 1000. ON (B)(6) 2025, CUSTOMER RERAN THE SAMPLE USING THE SAME MASTER LOT ON WORKLIST (B)(4) WHERE IT RESULTED AS DOUBLE CT NEGATIVE / GC NEGATIVE. CUSTOMER RELAYED THE INITIAL POSITIVE RESULT WAS HELD BACK, BUT DUE TO HOW THEIR LIS WAS PROGRAMMED, THE NEGATIVE RESULT WAS TRANSMITTED TO THE CAREGIVER. HOLOGIC HAS NOT LEARNED OF ANY ADVERSE PATIENT OUTCOMES DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23324704 ↗
  • InjuryOct 17, 2025Off-Label Use; Off-Label Use

    ON (B)(6) 2025, A CUSTOMER REPORTED A DISCREPANT RESULT USING THE APTIMA COMBO 2 (AC2) ASSAY ON THEIR PANTHER INSTRUMENT. ON (B)(6) 2025, SAMPLE ID (B)(6) RESULTED AS CT POSITIVE, GC NEGATIVE WITH MASTER LOT 915869 ON WORKLIST (B)(4). CUSTOMER REPORTED THEY HAVE MADE IT THEIR LAB POLICY TO RETEST ANY POSITIVE RESULT WITH AN RLU VALUE BELOW 1000. ON (B)(6) 2025, CUSTOMER RERAN THE SAMPLE USING MASTER LOT 922619 ON WORKLIST (B)(4) WHERE IT RESULTED AS DOUBLE CT NEGATIVE / GC NEGATIVE. CUSTOMER RELAYED THE INITIAL POSITIVE RESULT WAS HELD BACK, BUT DUE TO HOW THEIR LIS WAS PROGRAMMED, THE NEGATIVE RESULT WAS TRANSMITTED TO THE CAREGIVER. HOLOGIC HAS NOT LEARNED OF ANY ADVERSE PATIENT OUTCOMES DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23324701 ↗
  • InjuryOct 21, 2022No Apparent Adverse Event; No Apparent Adverse Event

    CUSTOMER CONTACTED HOLOGIC ON (B)(6) 2022 QUESTIONING AC2 ASSAY RESULT FROM LOT 327730. CUSTOMER WAS RUNNING THE UPDATED APTIMA COMBO 2 ASSAY. CUSTOMER PROVIDED SAMPLES TO HOLOGIC FOR EVALUATION. HOLOGIC PERFORMED TESTING OF THE SAMPLES USING THE UPDATED APTIMA COMBO 2 ASSAY AND CONFIRMED THE POSITIVE RESULTS. THE RESULTS ARE CONSISTENT WITH A CT VARIANT, ALTHOUGH SEQUENCING HAS NOT YET BEEN PERFORMED TO CONFIRM IF THE SAMPLE ARE A CT VARIANT(S).

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15655759 ↗

Other DNA-REAGENTS, NEISSERIA devices

All DNA-REAGENTS, NEISSERIA devices →