PhaSeal Optima 515063Recalls & reported problems

Becton Dickinson Pen Limited · PhaSeal Optima Protector (P28-O) · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00382905150630), received between Oct 11, 2023 and Oct 11, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 16, 2025Open, Classified

    Closed system drug transfer devices were shipped to the U.S. market without a U.S. version of the instructions for use, which could be an issue if a healthcare professional is prompted to review it.

    Z-2453-2025 · Root cause: Error in labeling · FDA record ↗

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 11, 2023Break; Break

    IT WAS REPORTED WHILE USING BD PHASEAL¿ OPTIMA PROTECTOR P28-O THE DEVICE WAS DAMAGED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: AND WE HAD ANOTHER INCIDENT WHERE THE LARGER PROTECTORS ARE NOT EXPANDING. THIS TIME THE PROTECTOR ACTUALLY CRACKED WHERE THE PROTECTOR AND SYRINGE CONNECT AND SPRAYED CISPLATIN INTO THE HOOD BECAUSE THE AIR WENT INTO THE VIAL AND NOT THE PROTECTOR. BEAR IN MIND THIS IS ONLY HAPPENING WITH THE LARGER PROTECTORS AND IT'S NOT HAPPENING EVERY TIME WE UTILIZE THEM. HERE ARE THE STEPS WE ARE TAKING: DRAWING BACK TO ANYWHERE BETWEEN 30ML TO 40ML OF AIR LEAVING THE VIAL SITTING IN THE HOOD, DIRECTING AIR INTO THE PROTECTOR BALLOON DOES NOT EXPAND, THEREFORE GIVING US "PUSH BACK" AGAINST THE SYRINGE THIS TIME THE PUSH BACK CAUSED THE SNAPPING OF THE PROTECTOR AND SYRINGE AND SPRAYING CISPLATIN UNTIL THE VIAL EQUALIZED PRESSURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17916786 ↗