BD PhaSeal Injector Luer (N30C), REF: 515001 recallZ-2453-2025 · Bd Switzerland Sarl · initiated Jun 16, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bd Switzerland Sarl or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Closed system drug transfer devices were shipped to the U.S. market without a U.S. version of the instructions for use, which could be an issue if a healthcare professional is prompted to review it.

FDA root cause category: Error in labeling

What the company is doing

On 6/16/2025, Medical Device Product Advisory notices were mailed and emailed to customers who were asked to do the following: 1) The US IFU to be attached to your product can be accessed and downloaded from the following link by searching under the applicable product code: The US IFU to be attached to your product can be accessed and downloaded from the following link by searching under the applicable product code: eIFU.bd.com 2) Circulate this notice to all those who need to be aware within your organization or to any organization where the potentially affected products have been transferred. 3) Complete the attached Customer Response Form and return via email to [email protected] Distributors were asked to do the following: Identify all customers within your distribution network that purchased any affected product as defined in this notification. Provide a copy of the attached customer letter to all customers to advise them of this product advisory on recalling firm's behalf. Report any complaints experienced with the use of this product to firm via the North American Regional Complaint Center: Phone: 1-844-823-5433 Say "Product Complaints" when prompted Mon-Fri 8:00am to 5:00pm CT, Email: [email protected]

Recall details

Recall number
Z-2453-2025
Status
Open, Classified
Initiated by the company
Jun 16, 2025
Posted by the FDA
Aug 27, 2025
Recalling firm
Bd Switzerland Sarl
Quantity
3,884,670
Product code
ONB
FDA submissions
K123213, K130197, K140591, K181221, K201099
Distribution
US Nationwide distribution including in the states of CA, WA, AL, OH, NC, GA, MA, IL, MO, AZ, PA, MS, MN, MI, TX, VA, IN, NV, CO, FL, WV, WI, NY, NJ, UT, MD, MT, NM, TN, HI, OK, OR, AK, DC, IA, LA, ME, SC.

Product description

BD PhaSeal Injector Luer (N30C), REF: 515001 BD PhaSeal Injector Luer Lock (N35), REF: 515003 BD PhaSeal Injector Luer Lock (N35C), REF: 515004 BD PhaSeal Injector Lu63er Lock (N35C Multi), REF: 515005 BD PhaSeal Protector (P14), REF: 515100 BD PhaSeal Protector (P21), REF: 515102 BD PhaSeal Protector (P21 Multi), REF: 515103 BD PhaSeal Protector (P28), REF: 515104 BD PhaSeal Protector (P50), REF: 515105 BD PhaSeal Protector (P50 Multi), REF: 515106 BD PhaSeal Protector (P53), REF: 515107 BD PhaSeal Protector (P55), REF: 515117 BD PhaSeal Connector Luer Lock (C35), REF: 515200 BD PhaSeal Connector Luer Lock (C45), REF: 515202 BD PhaSeal Y-site Connector (C80), REF: 515304 BD PhaSeal L Connector (C90), REF: 515305 BD PhaSeal Infusion Adapter (C100), REF: 515306 BD PhaSeal Infusion Adapter (C100 Multi), REF: 515307 BD PhaSeal Cap for Injector (M50), REF: 515404 BD PhaSeal Connector Cap (M70), REF: 515406 BD PhaSeal Optima Injector (N35-O), REF: 515052 BD PhaSeal Optima Injector (N35-O Multi), REF: 515053 BD PhaSeal Injector Locking (N40-O), REF: 515056 BD PhaSeal Injector Locking (N40-O Multi), REF: 515057 BD PhaSeal Optima Protector (P13-O), REF: 515060 BD PhaSeal Optima Protector (P13-O Multi), REF: 515062 BD PhaSeal Optima Protector (P28-O), REF: 515063 BD PhaSeal Optima Protector (P20-O), REF: 515064 BD PhaSeal Optima Protector (P20-O Multi), REF: 515065 BD PhaSeal Optima Connector (C35-O), REF: 515070 BD PhaSeal Optima Infusion Adapter (C100-O), 515078 BD PhaSeal Optima Infusion Adapter (C100-O Multi), REF: 515079

Code information (models, lots, serial numbers)

Show the full code information (16k characters)

REF/UDI-DI/Lot(Expiration): 515001/00382905150012. 515003/00382905150036/2402001(31-Jul-26), 2403002(31-Aug-26), 2403003(31-Aug-26), 2404006(30-Sep-26), 2405007(31-Oct-26), 2405014(31-Oct-26), 2408005(31-Jan-27), 2408006(31-Jan-27), 2409005(28-Feb-27), 2403713(31-Aug-26), 2404001(30-Sep-26), 2404002(30-Sep-26), 2404003(30-Sep-26), 2404004(30-Sep-26), 2404005(30-Sep-26), 2410001(31-Mar-27), 2410002(31-Mar-27), 2411001(30-Apr-27), 2411002(30-Apr-27), 2411003(30-Apr-27), 2411709(30-Apr-27), 2412002(31-May-27), 2501001(30-Jun-27), 2501002(30-Jun-27), 2501003(30-Jun-27), 2502001(31-Jul-27), 2502002(31-Jul-27), 2405004(31-Oct-26), 2405008(31-Oct-26), 2407001(31-Dec-26), 2412001(31-May-27), 2503001(31-Aug-27), 2503002(31-Aug-27), 2503003(31-Aug-27), 2504001(30-Sep-27), 2504002(30-Sep-27), 2504003(30-Sep-27); 515004/00382905150043/2405001(31-Oct-26), 2408018(31-Jan-27), 2411708(30-Apr-27), 2411727(30-Apr-27); 515005/00382905150050/2403006(31-Aug-26), 2404008(30-Sep-26), 2404009(30-Sep-26), 2404010(30-Sep-26), 2404011(30-Sep-26), 2404012(30-Sep-26), 2405002(31-Oct-26), 2407010(31-Dec-26), 2409011(28-Feb-27), 2409012(28-Feb-27), 2409016(28-Feb-27), 2410007(31-Mar-27), 2410008(31-Mar-27), 2410010(31-Mar-27), 2411725(30-Apr-27), 2405003(31-Oct-26), 2407011(31-Dec-26), 2407012(31-Dec-26), 2408008(31-Jan-27), 2408009(31-Jan-27), 2409007(28-Feb-27), 2409008(28-Feb-27), 2409009(28-Feb-27), 2409010(28-Feb-27), 2501006(30-Jun-27), 2503738(31-Aug-27), 2504004(30-Sep-27); 515100/00382905151002/2504117(31-Mar-30), 2404103(31-Mar-29), 2404122(31-Mar-29), 2404123(31-Mar-29), 2406112(31-May-29), 2406113(31-May-29), 2404124(31-Mar-29), 2405130(30-Apr-29), 2407112(30-Jun-29), 2409113(31-Aug-29), 2410113(30-Sep-29), 2410114(30-Sep-29), 2411115(31-Oct-29), 2412109(30-Nov-29), 2502116(31-Jan-30), 2503112(28-Feb-30); 515102/00382905151026/2403101(28-Feb-29), 2403102(28-Feb-29), 2403103(28-Feb-29), 2403104(28-Feb-29), 2403141(28-Feb-29), 2403707(28-Feb-29), 2404130(31-Mar-29), 2404131(31-Mar-29), 2404132(31-Mar-29), 2404141(31-Mar-29), 2404143(31-Mar-29), 2409601(31-Aug-29), 2403140(28-Feb-29), 2404125(31-Mar-29), 2404126(31-Mar-29), 2404127(31-Mar-29), 2404128(31-Mar-29), 2404129(31-Mar-29), 2404142(31-Mar-29), 2409108(31-Aug-29), 2409109(31-Aug-29), 2410701(30-Sep-29), 2412108(30-Nov-29), 2412114(30-Nov-29), 2501101(31-Dec-29), 2501102(31-Dec-29), 2501103(31-Dec-29), 2501104(31-Dec-29), 2501105(31-Dec-29), 2504108(31-Mar-30), 2504109(31-Mar-30), 2504119(31-Mar-30. 515103/00382905151033/2404133(31-Mar-29), 2404134(31-Mar-29), 2404140(31-Mar-29), 2410101(30-Sep-29), 2410102(30-Sep-29), 2410103(30-Sep-29), 2410104(30-Sep-29), 2504111(31-Mar-30). 515104/00382905151040/2410119(30-Sep-29), 2410107(30-Sep-29). 515105/00382905151057/2402112(31-Jan-29), 2403122(28-Feb-29), 2403123(28-Feb-29), 2403124(28-Feb-29), 2403125(28-Feb-29), 2403127(28-Feb-29), 2404113(31-Mar-29), 2404115(31-Mar-29), 2404117(31-Mar-29), 2405101(30-Apr-29), 2405102(30-Apr-29), 2405106(30-Apr-29), 2405107(30-Apr-29), 2405108(30-Apr-29), 2405152(30-Apr-29), 2406104(31-May-29), 2406105(31-May-29), 2406106(31-May-29), 2406107(31-May-29), 2406109(31-May-29), 2407103(30-Jun-29), 2407104(30-Jun-29), 2407106(30-Jun-29), 2407107(30-Jun-29), 2407108(30-Jun-29), 2407109(30-Jun-29), 2409103(31-Aug-29), 2409602(31-Aug-29), 2411123(31-Oct-29), 2411125(31-Oct-29), 2411126(31-Oct-29), 2411707(31-Oct-29), 2412105(30-Nov-29), 2403126(28-Feb-29), 2404102(31-Mar-29), 2404110(31-Mar-29), 2404111(31-Mar-29), 2404112(31-Mar-29), 2404114(31-Mar-29), 2404116(31-Mar-29), 2404118(31-Mar-29), 2404119(31-Mar-29), 2405103(30-Apr-29), 2405104(30-Apr-29), 2405105(30-Apr-29), 2407102(30-Jun-29), 2407105(30-Jun-29), 2409104(31-Aug-29), 2409105(31-Aug-29), 2411113(31-Oct-29), 2411124(31-Oct-29), 2501112(31-Dec-29), 2501113(31-Dec-29), 2501114(31-Dec-29), 2501115(31-Dec-29), 2502106(31-Jan-30), 2502107(31-Jan-30), 2501111(31-Dec-29), 2502108(31-Jan-30), 2502109(31-Jan-30), 2502110(31-Jan-30), 2502111(31-Jan-30), 2502112(31-Jan-30), 2503101(28-Feb-30), 2503102(28-Feb-30), 2503103(28-Feb-30), 2503801(28-Feb-30), 2504101(31-Mar-30), 2504102(31-Mar-30), 2504103(31-Mar-30), 2504104(31-Mar-30), 2504105(31-Mar-30), 2504106(31-Mar-30). 515106/00382905151064/2403109(28-Feb-29), 2403110(28-Feb-29), 2403111(28-Feb-29), 2403112(28-Feb-29), 2403113(28-Feb-29), 2403114(28-Feb-29), 2403115(28-Feb-29), 2403116(28-Feb-29), 2403117(28-Feb-29), 2403118(28-Feb-29), 2403119(28-Feb-29), 2403120(28-Feb-29), 2403121(28-Feb-29), 2404101(31-Mar-29), 2404104(31-Mar-29), 2404105(31-Mar-29), 2404106(31-Mar-29), 2404107(31-Mar-29), 2404108(31-Mar-29), 2405119(30-Apr-29), 2405120(30-Apr-29), 2405121(30-Apr-29), 2405122(30-Apr-29), 2405123(30-Apr-29), 2405124(30-Apr-29), 2405125(30-Apr-29), 2405126(30-Apr-29), 2405127(30-Apr-29), 2405128(30-Apr-29), 2405129(30-Apr-29), 2406101(31-May-29), 2406102(31-May-29), 2406103(31-May-29), 2407101(30-Jun-29), 2408101(31-Jul-29), 2409102(31-Aug-29), 2410110(30-Sep-29), 2409101(31-Aug-29), 2410108(30-Sep-29), 2410109(30-Sep-29), 2411101(31-Oct-29), 2411102(31-Oct-29), 2411103(31-Oct-29), 2411104(31-Oct-29), 2412106(30-Nov-29), 2412107(30-Nov-29), 2501110(31-Dec-29), 2502102(31-Jan-30), 2502103(31-Jan-30), 2503107(28-Feb-30), 2503108(28-Feb-30), 2503118(28-Feb-30), 2503119(28-Feb-30), 2504107(31-Mar-30). 515107/00382905151071/2407115(30-Jun-29). 515117/00382905151170/2405131(30-Apr-29), 2405132(30-Apr-29), 2405133(30-Apr-29), 2405134(30-Apr-29), 2407113(30-Jun-29), 2407114(30-Jun-29), 2409115(31-Aug-29), 2409120(31-Aug-29), 2409121(31-Aug-29), 2409122(31-Aug-29), 2411117(31-Oct-29), 2503111(28-Feb-30), 2504115(31-Mar-30), 2504116(31-Mar-30). 515200/00382905152009/2403715(29-Feb-28), 2404208(31-Mar-28), 2405203(30-Apr-28), 2407204(30-Jun-28), 2409202(31-Aug-28), 2411202(31-Oct-28), 2501201(31-Dec-28), 2405204(30-Apr-28), 2407203(30-Jun-28), 2409201(31-Aug-28), 2410201(30-Sep-28), 2410202(30-Sep-28), 2411201(31-Oct-28), 2411203(31-Oct-28), 2501202(31-Dec-28), 2502201(31-Jan-29), 2503201(28-Feb-29), 2503202(28-Feb-29), 2504201(31-Mar-29), 2504202(31-Mar-29). 515202/00382905152023/2404207(31-Mar-28), 2404202(31-Mar-28), 2407201(30-Jun-28), 2412202(30-Nov-28), 2501222(31-Dec-28), 2503206(28-Feb-29), 2504203(31-Mar-29). 515304/00382905153044/2406206(30-Sep-26), 2412212(31-Jan-27), 2503802(30-Apr-27). 515305/00382905153051/2403220(29-Feb-28), 2405221(30-Apr-28), 2405222(30-Apr-28), 2409218(31-Aug-28), 2404201(31-Mar-28), 2404206(31-Mar-28), 2502215(31-Jan-29). 515306/00382905153068/2402228(31-Jan-29), 2404210(31-Mar-28), 2404211(31-Mar-28), 2404212(31-Mar-28), 2404213(31-Mar-28), 2404214(31-Mar-28), 2404215(31-Mar-28), 2404216(31-Mar-28), 2406204(31-May-28), 2410212(30-Sep-28), 2410213(30-Sep-28), 2412216(30-Nov-28), 2404209(31-Mar-28), 2405224(30-Apr-28), 2502213(31-Jan-29), 2502214(31-Jan-29), 2406202(31-May-28), 2406203(31-May-28), 2410217(30-Sep-28), 2411206(31-Oct-28), 2411207(31-Oct-28), 2411721(31-Oct-28), 2411722(31-Oct-28), 2411723(31-Oct-28), 2412205(30-Nov-28), 2412206(30-Nov-28), 2412217(30-Nov-28), 2501213(31-Dec-28), 2501214(31-Dec-28), 2503207(28-Feb-29), 2503208(28-Feb-29), 2503209(28-Feb-29), 2504204(31-Mar-29), 2504205(31-Mar-29), 2504206(31-Mar-29). 515307/00382905153075 2403714(29-Feb-28), 2502212(31-Jan-29), 2403601(28-Feb-26), 2405225(30-Apr-28), 2410216(30-Sep-28). 515404/00382905154041/2404203(31-Mar-29), 2409220(31-Aug-29), 2405201(30-Apr-29), 2405202(30-Apr-29), 2406205(31-May-29). 515406/00382905154065/2404217(31-Mar-29). 515052/00382905150524/2403303(31-Aug-26), 2403304(31-Aug-26), 2403306(31-Aug-26), 2403307(31-Aug-26), 2404301(30-Sep-26), 2406303(30-Nov-26), 2406702(30-Nov-26), 2406801(30-Nov-26), 2408305(31-Jan-27), 2409302(28-Feb-27), 2410304(31-Mar-27), 2410306(31-Mar-27), 2411301(30-Apr-27), 2411302(30-Apr-27), 2412305(31-May-27), 2501303(30-Jun-27), 2501304(30-Jun-27), 2501305(30-Jun-27), 2502303(31-Jul-27), 2502304(31-Jul-27), 2502305(31-Jul-27), 2503301(31-Aug-27), 2405301(31-Oct-26), 2405302(31-Oct-26), 2405303(31-Oct-26), 2406302(30-Nov-26), 2406310(30-Nov-26), 2406803(30-Nov-26), 2407301(31-Dec-26), 2407302(31-Dec-26), 2407303(31-Dec-26), 2408304(31-Jan-27), 2408306(31-Jan-27), 2408307(31-Jan-27), 2408308(31-Jan-27), 2409301(28-Feb-27), 2409303(28-Feb-27), 2409304(28-Feb-27), 2410305(31-Mar-27), 2410307(31-Mar-27), 2503716(31-Aug-27), 2503717(31-Aug-27), 2503721(31-Aug-27), 2503722(31-Aug-27), 2504303(30-Sep-27), 2504304(30-Sep-27), 2504305(30-Sep-27), 2504306(30-Sep-27), 2504307(30-Sep-27). 515053/00382905150531/2404302(30-Sep-26), 2404303(30-Sep-26), 2404304(30-Sep-26), 2409305(28-Feb-27), 2409306(28-Feb-27), 2409307(28-Feb-27), 2405304(31-Oct-26), 2405305(31-Oct-26), 2410308(31-Mar-27), 2412303(31-May-27), 2412304(31-May-27), 2501301(30-Jun-27), 2501302(30-Jun-27), 2406701(30-Nov-26), 2406703(30-Nov-26), 2407304(31-Dec-26), 2407305(31-Dec-26), 2407306(31-Dec-26), 2407307(31-Dec-26), 2407308(31-Dec-26), 2407802(31-Dec-26), 2408309(31-Jan-27), 2408310(31-Jan-27), 2411303(30-Apr-27), 2501601(30-Jun-27), 2502701(31-Jul-27), 2502702(31-Jul-27), 2502703(31-Jul-27), 2502704(31-Jul-27), 2503723(31-Aug-27), 2503724(31-Aug-27), 2503725(31-Aug-27), 2504308(30-Sep-27). 515056/00382905150562/2405306(31-Oct-26), 2405307(31-Oct-26), 2405308(31-Oct-26), 2405312(31-Oct-26), 2408301(31-Jan-27), 2406301(30-Nov-26), 2409310(28-Feb-27), 2410301(31-Mar-27), 2410303(31-Mar-27), 2411304(30-Apr-27), 2411305(30-Apr-27), 2412301(31-May-27), 2501306(30-Jun-27), 2501307(30-Jun-27), 2501308(30-Jun-27), 2502301(31-Jul-27), 2502302(31-Jul-27), 2407309(31-Dec-26), 2407310(31-Dec-26), 2408302(31-Jan-27), 2409309(28-Feb-27), 2410302(31-Mar-27), 2410309(31-Mar-27), 2411306(30-Apr-27), 2503736(31-Aug-27), 2503737(31-Aug-27), 2504301(30-Sep-27), 2504801(30-Sep-27). 515057/00382905150579/2405311(31-Oct-26), 2409308(28-Feb-27), 2411308(30-Apr-27), 2412302(31-May-27), 2501309(30-Jun-27), 2501310(30-Jun-27), 2408303(31-Jan-27), 2411307(30-Apr-27), 2504302(30-Sep-27). 515060/00382905150609/2403413(28-Feb-26), 2403415(28-Feb-26), 2404408(31-Mar-26), 2404409(31-Mar-26), 2405409(30-Apr-26), 2405410(30-Apr-26), 2405412(30-Apr-26), 2405413(30-Apr-26), 2405414(30-Apr-26), 2405415(30-Apr-26), 2406407(31-May-26), 2407805(30-Jun-26), 2407806(30-Jun-26), 2407807(30-Jun-26), 2403414(28-Feb-26), 2404407(31-Mar-26), 2404410(31-Mar-26), 2405411(30-Apr-26), 2407803(30-Jun-26), 2407804(30-Jun-26), 2408406(31-Jul-26), 2408407(31-Jul-26), 2408408(31-Jul-26), 2409411(31-Aug-26), 2409413(31-Aug-26), 2410402(30-Sep-26), 2410408(30-Sep-26), 2410409(30-Sep-26), 2410410(30-Sep-26), 2410416(30-Sep-26), 2410417(30-Sep-26), 2410602(30-Sep-26), 2411413(31-Oct-26), 2411414(31-Oct-26), 2412407(30-Nov-26), 2412408(30-Nov-26), 2412409(30-Nov-26), 2412410(30-Nov-26), 2412411(30-Nov-26), 2412701(30-Nov-26), 2501406(31-Dec-26), 2409412(31-Aug-26), 2409414(31-Aug-26), 2409415(31-Aug-26), 2409416(31-Aug-26), 2412412(30-Nov-26), 2412414(30-Nov-26), 2501407(31-Dec-26), 2501408(31-Dec-26), 2504409(31-Mar-27). 515062/00382905150623. 515063/00382905150630/2502706(31-Mar-27), 2502705(31-Mar-27). 515064/00382905150647/2402405(31-Jan-26), 2403401(28-Feb-26), 2403404(28-Feb-26), 2403405(28-Feb-26), 2403408(28-Feb-26), 2403409(28-Feb-26), 2403410(28-Feb-26), 2403411(28-Feb-26), 2403412(28-Feb-26), 2403703(28-Feb-26), 2403704(28-Feb-26), 2404403(31-Mar-26), 2404404(31-Mar-26), 2404405(31-Mar-26), 2404411(31-Mar-26), 2404412(31-Mar-26), 2404416(31-Mar-26), 2405401(30-Apr-26), 2405403(30-Apr-26), 2405407(30-Apr-26), 2405408(30-Apr-26), 2405417(30-Apr-26), 2406402(31-May-26), 2406403(31-May-26), 2406404(31-May-26), 2406411(31-May-26), 2406412(31-May-26), 2406413(31-May-26), 2406414(31-May-26), 2406415(31-May-26), 2406416(31-May-26), 2407402(30-Jun-26), 2407403(30-Jun-26), 2407404(30-Jun-26), 2408410(31-Jul-26), 2408413(31-Jul-26), 2408414(31-Jul-26), 2405402(30-Apr-26), 2407415(30-Jun-26), 2408403(31-Jul-26), 2408404(31-Jul-26), 2408411(31-Jul-26), 2408412(31-Jul-26), 2409401(31-Aug-26), 2409402(31-Aug-26), 2409403(31-Aug-26), 2409405(31-Aug-26), 2411403(31-Oct-26), 2411404(31-Oct-26), 2411411(31-Oct-26), 2411412(31-Oct-26), 2412401(30-Nov-26), 2412402(30-Nov-26), 2412403(30-Nov-26), 2501403(31-Dec-26), 2501404(31-Dec-26), 2501415(31-Dec-26), 2501416(31-Dec-26), 2502401(31-Jan-27), 2502403(31-Jan-27), 2502404(31-Jan-27), 2502408(31-Jan-27), 2502410(31-Jan-27), 2502707(31-Jan-27), 2410403(30-Sep-26), 2410407(30-Sep-26), 2410413(30-Sep-26), 2410419(30-Sep-26), 2410420(30-Sep-26), 2410601(30-Sep-26), 2411410(31-Oct-26), 2501405(31-Dec-26), 2502708(31-Jan-27), 2502709(31-Jan-27), 2503703(28-Feb-27), 2503705(28-Feb-27), 2503706(28-Feb-27), 2503707(28-Feb-27), 2503708(28-Feb-27), 2504405(31-Mar-27), 2504406(31-Mar-27), 2504407(31-Mar-27), 2504408(31-Mar-27), 2505404(31-Mar-27). 515065/00382905150654/2402404(31-Jan-26), 2402411(31-Jan-26), 2402412(31-Jan-26), 2403403(28-Feb-26), 2403406(28-Feb-26), 2403416(28-Feb-26), 2403417(28-Feb-26), 2404401(31-Mar-26), 2404402(31-Mar-26), 2404406(31-Mar-26), 2405404(30-Apr-26), 2405405(30-Apr-26), 2405406(30-Apr-26), 2406401(31-May-26), 2406405(31-May-26), 2406406(31-May-26), 2406408(31-May-26), 2406409(31-May-26), 2406410(31-May-26), 2407401(30-Jun-26), 2407405(30-Jun-26), 2407406(30-Jun-26), 2407407(30-Jun-26), 2407408(30-Jun-26), 2407409(30-Jun-26), 2407416(30-Jun-26), 2407417(30-Jun-26), 2408401(31-Jul-26), 2408405(31-Jul-26), 2408417(31-Jul-26), 2409404(31-Aug-26), 2409407(31-Aug-26), 2409408(31-Aug-26), 2404417(31-Mar-26), 2408402(31-Jul-26), 2409406(31-Aug-26), 2410406(30-Sep-26), 2411401(31-Oct-26), 2411402(31-Oct-26), 2411418(31-Oct-26), 2412404(30-Nov-26), 2501401(31-Dec-26), 2501413(31-Dec-26), 2501418(31-Dec-26), 2410401(30-Sep-26), 2502405(31-Jan-27), 2410404(30-Sep-26), 2410405(30-Sep-26), 2502409(31-Jan-27), 2502415(31-Jan-27), 2410414(30-Sep-26), 2410415(30-Sep-26), 2410418(30-Sep-26), 2412405(30-Nov-26), 2412601(30-Nov-26), 2501402(31-Dec-26), 2501414(31-Dec-26), 2501417(31-Dec-26), 2502402(31-Jan-27), 2502413(31-Jan-27), 2502414(31-Jan-27), 2503405(28-Feb-27), 2503601(28-Feb-27), 2503701(28-Feb-27), 2503702(28-Feb-27), 2503704(28-Feb-27), 2503709(28-Feb-27), 2503710(28-Feb-27), 2503711(28-Feb-27), 2503712(28-Feb-27), 2503713(28-Feb-27), 2503715(28-Feb-27), 2503741(28-Feb-27), 2503742(28-Feb-27), 2504403(31-Mar-27), 2504404(31-Mar-27), 2504601(31-Mar-27), 2504701(31-Mar-27), 2504702(31-Mar-27). 515070/00382905150708/2403701(28-Feb-26), 2403705(28-Feb-26), 2404507(31-Mar-26), 2404508(31-Mar-26), 2404509(31-Mar-26), 2405505(30-Apr-26), 2405506(30-Apr-26), 2406507(31-May-26), 2407506(30-Jun-26), 2407507(30-Jun-26), 2404506(31-Mar-26), 2406508(31-May-26), 2409501(31-Aug-26), 2409502(31-Aug-26), 2409506(31-Aug-26), 2410505(30-Sep-26), 2410506(30-Sep-26), 2411503(31-Oct-26), 2411504(31-Oct-26), 2412503(30-Nov-26), 2412504(30-Nov-26), 2501506(31-Dec-26), 2501507(31-Dec-26), 2501508(31-Dec-26), 2502501(31-Jan-27), 2501518(31-Dec-26), 2502502(31-Jan-27), 2503726(28-Feb-27), 2503727(28-Feb-27), 2504501(31-Mar-27), 2504502(31-Mar-27), 2504503(31-Mar-27). 515078/00382905150784/2402505(31-Jan-26), 2403505(28-Feb-26), 2403507(28-Feb-26), 2404501(31-Mar-26), 2404502(31-Mar-26), 2405501(30-Apr-26), 2405502(30-Apr-26), 2405503(30-Apr-26), 2405504(30-Apr-26), 2406504(31-May-26), 2406505(31-May-26), 2406506(31-May-26), 2407501(30-Jun-26), 2407502(30-Jun-26), 2407801(30-Jun-26), 2403504(28-Feb-26), 2404505(31-Mar-26), 2408501(31-Jul-26), 2409504(31-Aug-26), 2409505(31-Aug-26), 2410501(30-Sep-26), 2410502(30-Sep-26), 2410507(30-Sep-26), 2411501(31-Oct-26), 2411502(31-Oct-26), 2411508(31-Oct-26), 2412502(30-Nov-26), 2501501(31-Dec-26), 2501502(31-Dec-26), 2501503(31-Dec-26), 2502505(31-Jan-27), 2502506(31-Jan-27), 2502507(31-Jan-27), 2503728(28-Feb-27), 2504506(31-Mar-27), 2504507(31-Mar-27). 515079/00382905150791/2402601(31-Jan-26), 2403702(28-Feb-26), 2403706(28-Feb-26), 2404503(31-Mar-26), 2406501(31-May-26), 2406502(31-May-26), 2406503(31-May-26), 2408502(31-Jul-26), 2408503(31-Jul-26), 2408504(31-Jul-26), 2408505(31-Jul-26), 2410503(30-Sep-26), 2410504(30-Sep-26), 2403508(28-Feb-26), 2409503(31-Aug-26), 2411506(31-Oct-26), 2411507(31-Oct-26), 2412501(30-Nov-26), 2501504(31-Dec-26), 2501505(31-Dec-26), 2502503(31-Jan-27), 2502504(31-Jan-27), 2503504(28-Feb-27), 2503505(28-Feb-27), 2503506(28-Feb-27), 2504504(31-Mar-27), 2504505(31-Mar-27).

Devices named in this recall

32 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page