Skip to content

BD PhaSeal 515106Recalls & reported problems

BD Switzerland Sàrl · Protector (P50 Multi) · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This BD PhaSeal protector is a closed system used to mix and transfer chemotherapy and other hazardous drugs without exposing staff or patients. There are 64 reports naming this device, with no deaths and no injuries; they mostly describe contamination, leaks or splashes, and "coring"—fragments of rubber or membrane breaking off inside drug vials during preparation—though nearly all reports note no harm reached the patient. One open recall began in June 2025 because units were shipped without U.S. instructions for use.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
64
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 64 adverse event reports that list this exact device (DI 00382905151064), received between Mar 13, 2019 and Oct 17, 2025 — classified as 64 malfunctions.

The most frequently reported specific device problem is “Contamination” (20 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 3 reportsMay 2025: 1 reportsJun 2025: 11 reportsJul 2025: 4 reportsAug 2025: 6 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 16, 2025Open, Classified

    Closed system drug transfer devices were shipped to the U.S. market without a U.S. version of the instructions for use, which could be an issue if a healthcare professional is prompted to review it.

    Z-2453-2025 · Root cause: Error in labeling · FDA record ↗

Most reported problems

Device problems

  • Contamination20
  • Device Contamination with Chemical or Other Material19
  • Leak/Splash13
  • Defective Component7
  • Break2
  • Detachment of Device or Device Component1
  • Device Markings/Labelling Problem1
  • Unsealed Device Packaging1

Patient problems

  • No Clinical Signs, Symptoms or Conditions48
  • No Consequences Or Impact To Patient6
  • No Known Impact Or Consequence To Patient5
  • No Patient Involvement5

Recent adverse event reports

The 25 most recent of 64 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 17, 2025Contamination; Contamination

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23322179 ↗
  • MalfunctionAug 25, 2025Contamination; Contamination

    THIS IS A COMPLAINT ABOUT ONIVIDE CORING. OCCURRED DURING ONIVADE PREPARATION. 2 VIALS WERE PREPARED. 1 VIAL WAS RETURNED TO THE INFUSION BAG AFTER VISUAL INSPECTION OF THE INSIDE OF THE VIAL CONFIRMED THAT NO CORING HAD OCCURRED. THE OTHER VIAL (THE ONE SENT TO US) HAD CORING AND WE JUDGED THAT IT WAS NOT POSSIBLE TO COLLECT. THE COMPANY THEN CHECKED THE BAG IN WHICH THE FIRST VIAL WAS RETURNED AND CONFIRMED THAT CORING HAD OCCURRED. AS A RESULT, TWO VIALS WERE DISCARDED. THE INJECTOR AND PROTECTOR WERE CONNECTED. (THE LOT OF INJECTOR 515005 IS 2410010).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22885228 ↗
  • MalfunctionAug 25, 2025Contamination; Contamination

    THIS IS A COMPLAINT ABOUT ONIVIDE CORING. CORING OCCURRED DURING ONIVADE PREPARATION. 2 VIALS WERE PREPARED. 1 VIAL WAS RETURNED TO THE INFUSION BAG AFTER VISUAL INSPECTION OF THE INSIDE OF THE VIAL CONFIRMED THAT NO CORING HAD OCCURRED. THE OTHER VIAL (THE ONE SENT TO US) HAD CORING AND WE JUDGED THAT IT WAS NOT POSSIBLE TO COLLECT. THE COMPANY THEN CHECKED THE BAG IN WHICH THE FIRST VIAL WAS RETURNED AND CONFIRMED THAT CORING HAD OCCURRED. AS A RESULT, TWO VIALS WERE DISCARDED. THE INJECTOR AND PROTECTOR WERE CONNECTED. (THE LOT OF INJECTORS 515005 IS 2410010).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22885202 ↗
  • MalfunctionAug 25, 2025Contamination; Contamination

    THIS IS A COMPLAINT ABOUT ONIVIDE CORING. CORING OCCURRED DURING ONIVADE PREPARATION. 2 VIALS WERE PREPARED. 1 VIAL WAS RETURNED TO THE INFUSION BAG AFTER VISUAL INSPECTION OF THE INSIDE OF THE VIAL CONFIRMED THAT NO CORING HAD OCCURRED. THE OTHER VIAL (THE ONE SENT TO US) HAD CORING AND WE JUDGED THAT IT WAS NOT POSSIBLE TO COLLECT. THE COMPANY THEN CHECKED THE BAG IN WHICH THE FIRST VIAL WAS RETURNED AND CONFIRMED THAT CORING HAD OCCURRED. AS A RESULT, TWO VIALS WERE DISCARDED. THE INJECTOR AND PROTECTOR WERE CONNECTED. (THE LOT OF INJECTORS 515005 IS 2410010).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22885199 ↗
  • MalfunctionAug 25, 2025Contamination; Contamination

    IT WAS REPORTED THAT CORING WAS DISCOVERED DURING PEPCID PREPARATION. THE INJECTOR AND PROTECTOR HAD ALREADY BEEN CONNECTED. COLLECTION WAS INTERRUPTED AND ANOTHER VIAL WAS USED. JUDGING FROM THE COLOR, IT APPEARS TO BE ABRASION OF THE MEMBRANE. (LOT NUMBER FOR INJECTOR 515005 IS 2410010.)

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22885176 ↗
  • MalfunctionAug 7, 2025Contamination; Contamination

    IT IS REPORTED CORING. EVENT DESCRIPTION: THIS IS A COMPLAINT ABOUT BEVACIZUMAB CORING. IT WAS REPORTED THAT CORING OCCURRED INSIDE THE VIAL. CORING WAS DISCOVERED WHEN ATTEMPTING TO WITHDRAW THE DRUG (AFTER CONNECTING THE PROTECTOR AND INJECTOR). THE INJECTOR (515005) IS LOT 2410008.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22739688 ↗
  • MalfunctionAug 7, 2025Contamination; Contamination

    IT IS REPORTED CORING. EVENT DESCRIPTION: THIS IS A COMPLAINT ABOUT CORING IN RITUXIMAB. IT WAS REPORTED THAT CORE FOUND DURING PREPARATION OF RITUXIMAB. CORE FOUND AFTER CONNECTING THE INJECTOR AND PROTECTOR. INJECTOR (515005) LOT2410008.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22739657 ↗
  • MalfunctionJul 24, 2025Contamination; Contamination

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22588126 ↗
  • MalfunctionJul 24, 2025Contamination; Contamination

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22587343 ↗
  • MalfunctionJul 8, 2025Defective Component; Defective Component

    DESCRIPTION: CORING. EVENT DETAILS: IT WAS REPORTED THAT CORING DURING CIRCULISA PREPARATION. CONFIRMED IN VIAL AFTER CONNECTING PROTECTOR AND INJECTOR. THE SIZE SUGGESTS THAT IT WAS CAUSED BY PASSING THROUGH THE AIR NEEDLE OF THE PROTECTOR. INJECTOR LOT IS 2409011 AND PART NUMBER IS 515005.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22441316 ↗
  • MalfunctionJul 8, 2025Defective Component; Defective Component

    DESCRIPTION: CORING. EVENT DETAILS: IT WAS REPORTED THAT CORE FORMED DURING PREPARATION OF ENDOXAN. CONFIRMED IN VIAL AFTER CONNECTING PROTECTOR AND INJECTOR. BASED ON SIZE, IT IS THOUGHT TO HAVE BEEN CAUSED BY THE PROTECTOR'S AIR NEEDLE PASSING THROUGH. THE LOT NUMBER OF THE INJECTOR IS 2409011 AND THE PART NUMBER IS 515005. CAN WE VERIFY IF THE PROTECTOR IS ASSEMBLED MANUALLY OR USING THE ASSEMBLY FIXTURE? THE PROTECTOR IS ASSEMBLED USING THE ASSEMBLY FIXTURE TO VERIFY, THE PARTICLE WAS OBSERVED BEFORE ENGAGING THE INJECTOR OR AFTER? CORING WAS DISCOVERED WHEN THE PROTECTOR AND INJECTOR WERE CONNECTED, AND THE DRUG WAS BEING COLLECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22440615 ↗
  • MalfunctionJun 26, 2025Contamination; Contamination

    FOREIGN OBJECT FOUND IN BAG CONTAINING PACLITAXEL. EVENT DETAILS: ONE VIAL OF PACLITAXEL 100ML AND ONE VIAL OF PACLITAXEL 30ML WERE PREPARED AND A FOREIGN BODY WAS FOUND IN THE BAG. AFTER PREPARING 5 VIALS OF 30ML PACLITAXEL NK AND 2 VIALS OF 100ML PACLITAXEL, THE DRUG SOLUTION WAS COLLECTED FROM EACH VIAL AND INFUSED INTO AN INFUSION BAG. (MINUSCULE). THE PRODUCT THAT WAS CONNECTED TO THE INFUSION BAG WAS 515309.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22336145 ↗
  • MalfunctionJun 25, 2025Contamination; Contamination

    THIS IS A COMPLAINT ABOUT FOREIGN MATTER FOUND IN ABRAXANE BAG. EVENT DETAILS: IT WAS REPORTED THAT THREE VIALS OF ABRAXANE WERE PREPARED, THE LIQUID WAS COLLECTED FROM EACH VIAL, INFUSED INTO THE INFUSION BAG, AND BEFORE THE INFUSION BAG WAS HANDED OVER TO THE NURSE, THE PERSON IN CHARGE CHECKED THE BAG AND FOUND FOREIGN MATTER. WE WOULD LIKE TO REQUEST AN ANALYSIS OF THE SIZE OF THE CORE. (FOREIGN OBJECT MIXED INSIDE THE BAG) THE LOT NUMBER OF THE INJECTOR IS UNKNOWN. THE INCIDENT OCCURRED AFTER CONNECTING THE PROTECTOR AND INJECTOR. THE PRODUCT THAT WAS CONNECTED TO THE INFUSION BAG WAS 515309.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22332404 ↗
  • MalfunctionJun 25, 2025Contamination; Contamination

    FOREIGN OBJECT FOUND IN KEYTRUDA BAG. EVENT DETAILS: AFTER PREPARING 4 VIALS OF KEYTRUDA AND COLLECTING DRUG SOLUTION FROM EACH VIAL, INFUSE THE SOLUTION INTO AN INFUSION BAG, AND BEFORE DISPENSING THE INFUSION BAG TO THE NURSE, THE AUDITOR CHECKED INSIDE THE BAG AND FOUND A SMALL FOREIGN OBJECT. WE WOULD LIKE AN ANALYSIS OF THE SIZE OF THE CORING. (IT IS MIXED IN THE BAG.) A LOT OF THE INJECTOR IS UNKNOWN. THE EVENT OCCURRED AFTER THE PROTECTOR AND INJECTOR WERE CONNECTED. THE PRODUCT THAT WAS CONNECTED TO THE INFUSION BAG WAS 515309.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330553 ↗
  • MalfunctionJun 25, 2025Contamination; Contamination

    FOREIGN OBJECT IN BAG. EVENT DETAILS: PREPARED 1 VIAL OF 100 ML PACLITAXEL AND 1 VIAL OF 30 ML PACLITAXEL, THEN CHECKED THE BAG AND FOUND A SINGLE FOREIGN OBJECT. EIGHT VIALS OF PACLITAXEL NK 30 ML WERE PREPARED, AND AFTER COLLECTING THE LIQUID FROM EACH VIAL, IT WAS INFUSED INTO THE INFUSION BAG. BEFORE THE INFUSION BAG WAS HANDED OVER TO THE NURSE, THE PERSON IN CHARGE CHECKED THE BAG AND FOUND MULTIPLE FOREIGN OBJECTS. THE LOT NUMBER OF THE INJECTOR IS UNKNOWN. THE INCIDENT OCCURRED AFTER CONNECTING THE PROTECTOR AND INJECTOR. THE CONNECTION WAS MADE APPROXIMATELY 10 TIMES. THE PRODUCT THAT WAS CONNECTED TO THE INFUSION BAG WAS 515309.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22329829 ↗
  • MalfunctionJun 25, 2025Contamination; Contamination

    IT WAS REPORTED THAT FOREIGN OBJECT FOUND IN KEYTRUDA BAG. EVENT DETAILS: TWO VIALS OF KEYTRUDA WERE USED TO PREPARE THE DRUG, AND AFTER COLLECTING THE DRUG SOLUTION FROM EACH VIAL, IT WAS INFUSED INTO AN INFUSION BAG, AND BEFORE THE INFUSION BAG WAS DISPENSED TO THE NURSE, THE AUDITOR CHECKED INSIDE THE BAG AND FOUND ONE SMALL FOREIGN OBJECT. WE WOULD LIKE AN ANALYSIS OF THE SIZE OF THE CORING. (IT IS MIXED IN THE BAG.) A LOT OF THE INJECTOR IS UNKNOWN. THE EVENT OCCURRED AFTER THE PROTECTOR AND INJECTOR WERE CONNECTED. THE PRODUCT THAT WAS CONNECTED TO THE INFUSION BAG WAS 515309.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22329105 ↗
  • MalfunctionJun 25, 2025Contamination; Contamination

    FOREIGN MATTER IN BAG. EVENT DETAILS: IT WAS REPORTED THAT TWO VIALS OF 30 ML OF PACLITAXEL NK AND ONE VIAL OF 100 ML WERE PREPARED, AND AFTER COLLECTING THE LIQUID FROM EACH VIAL, IT WAS INFUSED INTO THE INFUSION BAG. BEFORE THE INFUSION BAG WAS HANDED OVER TO THE NURSE, THE PERSON IN CHARGE CHECKED THE BAG AND FOUND A FOREIGN OBJECT. WE REQUEST AN ANALYSIS OF THE SIZE OF THE FOREIGN OBJECT (PRESENT INSIDE THE BAG). THE LOT NUMBER OF THE INJECTOR IS UNKNOWN. THE INCIDENT OCCURRED AFTER CONNECTING THE PROTECTOR AND INJECTOR. THE CONNECTION WAS MADE APPROXIMATELY SIX TIMES. THE PRODUCT THAT WAS CONNECTED TO THE INFUSION BAG WAS 515309.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22329092 ↗
  • MalfunctionJun 25, 2025Contamination; Contamination

    FOREIGN MATTER FOUND IN KEYTRUDA BAG. EVENT DETAILS: IT WAS REPORTED THAT TWO VIALS OF KEYTRUDA WERE PREPARED, THE LIQUID WAS COLLECTED FROM EACH VIAL AND INFUSED INTO THE INFUSION BAG. BEFORE THE INFUSION BAG WAS HANDED OVER TO THE NURSE, THE PERSON IN CHARGE CHECKED THE BAG AND FOUND FOREIGN MATTER. ONE PARTICLE WAS FOUND BY VISUAL INSPECTION. PLEASE ANALYZE THE SIZE OF THE CORE. (IT IS MIXED IN THE BAG.) THE LOT OF THE INJECTOR IS UNKNOWN. THE INCIDENT OCCURRED AFTER CONNECTING THE PROTECTOR AND INJECTOR. THE PRODUCT THAT WAS CONNECTED TO THE INFUSION BAG WAS 515309.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22329090 ↗
  • MalfunctionJun 25, 2025Contamination; Contamination

    IT WAS REPORTED FOREIGN OBJECT WAS FOUND. EVENT DETAILS: 30ML OF PACLITAXEL NK WAS PREPARED USING 4 VIALS, AND AFTER COLLECTING THE LIQUID FROM EACH VIAL, IT WAS INJECTED INTO AN INFUSION BAG. WHEN THE AUDITOR CHECKED THE INSIDE OF THE BAG BEFORE DISPENSING THE INFUSION BAG TO THE NURSE, A FOREIGN OBJECT WAS FOUND. WE WOULD LIKE YOU TO ANALYZE THE SIZE OF THE CORING. (CONTAMINATED INSIDE THE BAG) THE LOT OF THE INJECTOR IS UNKNOWN. THE INCIDENT OCCURRED AFTER THE PROTECTOR AND INJECTOR WERE CONNECTED. THE NUMBER OF CONNECTIONS WAS ABOUT 9 TIMES. 4 VIALS AND 1 INFUSION BAG WILL BE RETURNED. ADDITIONAL INFORMATION: 1: THIS IS THE SAME INJECTOR. 515005 LOT IS UNKNOWN. 2: THE L CONNECTOR IS CONNECTED, AND THE COLLECTED MEDICATION IS INJECTED FROM THE L CONNECTOR INTO THE INFUSION BAG.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22328029 ↗
  • MalfunctionJun 25, 2025Contamination; Contamination

    FOREIGN OBJECT FOUND IN PACLITAXEL BAG. EVENT DETAILS: 11 VIALS OF 30ML PACLITAXEL NK WERE USED TO PREPARE THE DRUG, AND AFTER COLLECTING THE DRUG SOLUTION FROM EACH VIAL, IT WAS INJECTED INTO AN INFUSION BAG. WHEN CHECKED THE INSIDE OF THE BAG BEFORE DISPENSING THE INFUSION BAG, A FOREIGN OBJECT WAS FOUND. (PERHAPS BECAUSE SOME TIME HAS PASSED, THERE APPEARS TO BE CRYSTALLIZED DRUG FLOATING IN THE BAG, SEPARATE FROM THE CORING.) THE INJECTOR'S LOT IS UNKNOWN. THE INCIDENT OCCURRED AFTER THE PROTECTOR AND INJECTOR WERE CONNECTED. THE DRUG ADDED WAS PACLITAXEL. ADDITIONAL INFORMATION: THE PRODUCT THAT WAS CONNECTED TO THE INFUSION BAG WAS 515309. 1: PACLITAXEL IS A DRUG THAT IS SAID TO HAVE THE POTENTIAL TO CRYSTALLIZE (THIS IS ALSO WRITTEN IN THE PACKAGE INSERT), BUT THE PROBLEM HERE IS NOT THAT PART, BUT THE CORING THAT IS MIXED IN THE INFUSION BAG. IT IS MENTIONED TO PREVENT THE CRYSTALS FROM GETTING IN THE WAY AND PREVENTING THE CORING FROM BEING FOUND. 2: IT'S HIGHLY LIKELY THAT IT'S THE SAME INJECTOR. IN THAT CASE, IT MAY HAVE MADE 22 ROUND TRIPS. 515005 LOT IS UNKNOWN. 3: THE L CONNECTOR IS CONNECTED, AND THE COLLECTED DRUG IS INJECTED INTO THE INFUSION BAG FROM THE L CONNECT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22327651 ↗
  • MalfunctionJun 25, 2025Contamination; Contamination

    FOREIGN OBJECT FOUND IN KEYTRUDA BAG. EVENT DETAILS: KEYTRUDA WAS PREPARED USING TWO VIALS, AND THE DRUG SOLUTION WAS COLLECTED FROM EACH VIAL AND INJECTED INTO AN INFUSION BAG. WHEN CHECKED THE BAG BEFORE DISPENSING THE INFUSION BAG TO THE NURSE, A VERY SMALL FOREIGN OBJECT WAS FOUND. WE WOULD LIKE AN ANALYSIS OF THE SIZE OF THE CORING. (CONTAMINATED IN THE BAG) THE INJECTOR'S LOT IS UNKNOWN. THE INCIDENT OCCURRED AFTER THE PROTECTOR AND INJECTOR WERE CONNECTED. ADDITIONAL INFORMATION: THE PRODUCT THAT WAS CONNECTED TO THE INFUSION BAG WAS 515309 1: THIS IS THE SAME INJECTOR. 515005 LOT IS UNKNOWN. 2: THE L CONNECTOR IS CONNECTED, AND THE COLLECTED MEDICATION IS INJECTED FROM THE L CONNECTOR INTO THE INFUSION BAG.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22327629 ↗
  • MalfunctionJun 25, 2025Contamination; Contamination

    IT IS REPORTED, FOREIGN OBJECT FOUND. EVENT DETAILS: KEYTRUDA WAS PREPARED USING 4 VIALS, AND THE DRUG SOLUTION WAS COLLECTED FROM EACH VIAL AND INJECTED INTO AN INFUSION BAG. WHEN CHECKED THE BAG BEFORE DISPENSING THE INFUSION BAG, MULTIPLE SMALL FOREIGN OBJECTS WERE FOUND. THE INJECTOR'S LOT IS UNKNOWN. THE INCIDENT OCCURRED AFTER THE PROTECTOR AND INJECTOR WERE CONNECTED. THE DRUG ADDED WAS KEYTRUDA. WE WOULD LIKE AN ANALYSIS OF THE SIZE OF THE CORING (CONTAMINATED IN THE BAG). ADDITIONAL INFORMATION: THE PRODUCT THAT WAS CONNECTED TO THE INFUSION BAG WAS 515309 CAN WE VERIFY- WAS IT THE SAME INJECTOR USED TO COLLECT FROM THE FOUR VIALS? THIS IS THE SAME INJECTOR. 515005 LOT IS UNKNOWN. WAS AN INFUSION ADAPTER CONNECTED TO THE IV BAG? THE L CONNECTOR IS CONNECTED, AND THE COLLECTED MEDICATION IS INJECTED FROM THE L CONNECTOR INTO THE INFUSION BAG.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22327118 ↗
  • MalfunctionMay 22, 2025Defective Component; Defective Component

    VERBATIM: THIS IS A COMPLAINT ABOUT CORING. IT WAS REPORTED THAT AFTER PREPARING NINE DOSES OF PACLITAXEL, CORE WAS OBSERVED WHEN THE DRUG WAS RETURNED TO THE BAG. THE CORE WAS SMALL, SUGGESTING THAT IT ORIGINATED FROM THE INJECTOR. ADDITIONAL INFORMATION COULD YOU PLEASE CONFIRM WHETHER CORING OCCURRED AFTER THE INJECTOR INSTALLED? IF YES, COULD YOU PLEASE CONFIRM THE PRODUCT CODE AND LOT NUMBER OF THE INJECTOR? AFTER INSTALLING THE INJECTOR, CUSTOMER CONNECTED IT TO THE PROTECTOR AND CONFIRMED THE OCCURRENCE WHEN ATTEMPTING TO COLLECT THE LIQUID. THE LOT OF THE INJECTOR IS UNKNOWN. INJECTOR'S PRODUCT CODE IS 515005.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22069894 ↗
  • MalfunctionApr 16, 2025Defective Component; Defective Component

    IT WAS REPORTED THAT CORING OCCURRED DURING PREPARATION OF ALKERAN. NO PATIENT HARM. M12 ASSEMBLY FIXTURE WAS USED. A: OBSERVATION OF THE VIAL ONLY. A: AFTER THE INJECTOR WAS ASSEMBLED. A: INJECTOR LOT NUMBER 2408008 (PROBABLY) A: PROTECTOR LOT NUMBER (PROBABLY 2409102)

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21849377 ↗
  • MalfunctionApr 2, 2025Defective Component; Defective Component

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21753424 ↗

Other Closed antineoplastic and hazardous drug reconstitution and transfer system devices

All Closed antineoplastic and hazardous drug reconstitution and transfer system devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.