NIPRO SET BLOOD TUBING SET WITH TRANSDUCER PROTECTORS AND PRIMING SET A223D/V809DRecalls & reported problems
Nipro Medical Corporation · PUMP SEGMENT 8.0 mm PRIMING VOLUME 141 ML EQUIPMENT FRESENIUS 2008 · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00383790004176), received between May 9, 2018 and May 13, 2026 — classified as 2 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak4
- Material Puncture/Hole4
- Device Operational Issue1
- Obstruction of Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Hemorrhage/Blood Loss/Bleeding2
- Diaphoresis1
- Syncope1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 13, 2026Fluid/Blood Leak
THE CUSTOMER ADVISED THAT THE MACHINES ARE DETECTING CONTINUOUS AIR & BLOOD GOING INTO THE TRANSDUCER. NO ONE WAS INJURED. THE ISSUE OCCURRED DURING THE 1ST AND 2ND SHIFTS ONLY. AFTER REMOVING THE REPORTED LOT NUMBER, NO FURTHER ISSUES WERE NOTED. THE CLINIC EXPERIENCED CLOTTED SYSTEMS AND HAD TO RESTRING THE MACHINES USING BLOODLINES FROM A DIFFERENT LOT NUMBER. ALL PATIENTS WERE ABLE TO COMPLETE THEIR TREATMENTS WITHOUT FURTHER COMPLICATIONS. SOME PATIENTS DID LOSE BLOOD; HOWEVER, THEY ARE UNABLE TO CONFIRM HOW MANY PATIENTS WERE AFFECTED OR THE AMOUNT OF BLOOD LOST. THEY ALSO CANNOT CONFIRM WHETHER HGB WAS CHECKED FOR ALL PATIENTS.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 25164933 ↗MalfunctionMay 13, 2026Fluid/Blood Leak
THE CUSTOMER ADVISED THAT THE MACHINES ARE DETECTING CONTINUOUS AIR & BLOOD GOING INTO THE TRANSDUCER. NO ONE WAS INJURED. THE ISSUE OCCURRED DURING THE 1ST AND 2ND SHIFTS ONLY. AFTER REMOVING THE REPORTED LOT NUMBER, NO FURTHER ISSUES WERE NOTED. THE CLINIC EXPERIENCED CLOTTED SYSTEMS AND HAD TO RESTRING THE MACHINES USING BLOODLINES FROM A DIFFERENT LOT NUMBER. ALL PATIENTS WERE ABLE TO COMPLETE THEIR TREATMENTS WITHOUT FURTHER COMPLICATIONS. SOME PATIENTS DID LOSE BLOOD; HOWEVER, THEY ARE UNABLE TO CONFIRM HOW MANY PATIENTS WERE AFFECTED OR THE AMOUNT OF BLOOD LOST. THEY ALSO CANNOT CONFIRM WHETHER HGB WAS CHECKED FOR ALL PATIENTS.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 25164220 ↗MalfunctionJan 13, 2023Material Puncture/Hole; Material Puncture/Hole
DURING THE 1ST MINUTE OF TREATMENT, A PHYSICAL LEAK WAS NOTED ON THE BLOOD PUMP, TREATMENT STOPPED, BLOOD WAS NOT RETURNED. NEW SET UP DONE USING SAME LOT #, DIFFERENT MACHINE AND FINISHED TREATMENT WITH NO ISSUES. PATIENT LOST 200ML OF BLOOD. NO HOSPITALIZATION. HBG RECHECKED, NO DATA ON BASELINE HGB AND NO RESULT YET. IV ANTIBIOTICS STARTED AS A PROPHYLAXIS. THE BIOMED ALSO CHECKED THE MACHINE AND ORDERED SOME PARTS OF THE BLOOD PUMP SEGMENT. THE MANAGER ALSO MENTIONED THAT THEY JUST HAD A RECALL OF SOME OF THE MACHINES BECAUSE OF ISSUES WITH THE BLOOD PUMP SEGMENT. OTHER DEVICES USED: FRESENIUS 2008T.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16164989 ↗MalfunctionJan 13, 2023Material Puncture/Hole; Material Puncture/Hole
DURING THE 1ST MINUTE OF TREATMENT, A PHYSICAL LEAK WAS NOTED ON THE BLOOD PUMP, TREATMENT STOPPED, BLOOD WAS NOT RETURNED. NEW SET UP DONE USING SAME LOT #, DIFFERENT MACHINE AND FINISHED TREATMENT WITH NO ISSUES. PATIENT LOST 200ML OF BLOOD. NO HOSPITALIZATION. HBG RECHECKED, NO DATA ON BASELINE HGB AND NO RESULT YET. IV ANTIBIOTICS STARTED AS A PROPHYLAXIS. THE BIOMED ALSO CHECKED THE MACHINE AND ORDERED SOME PARTS OF THE BLOOD PUMP SEGMENT. THE MANAGER ALSO MENTIONED THAT THEY JUST HAD A RECALL OF SOME OF THE MACHINES BECAUSE OF ISSUES WITH THE BLOOD PUMP SEGMENT. OTHER DEVICES USED: FRESENIUS 2008T.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16164902 ↗MalfunctionJan 13, 2023Material Puncture/Hole; Material Puncture/Hole
DURING THE 1ST MINUTE OF TREATMENT, A PHYSICAL LEAK WAS NOTED ON THE BLOOD PUMP, TREATMENT STOPPED, BLOOD WAS NOT RETURNED. NEW SET UP DONE USING SAME LOT #, DIFFERENT MACHINE AND FINISHED TREATMENT WITH NO ISSUES. PATIENT LOST 200ML OF BLOOD. NO HOSPITALIZATION. HBG RECHECKED, NO DATA ON BASELINE HGB AND NO RESULT YET. IV ANTIBIOTICS STARTED AS A PROPHYLAXIS. THE BIOMED ALSO CHECKED THE MACHINE AND ORDERED SOME PARTS OF THE BLOOD PUMP SEGMENT. THE MANAGER ALSO MENTIONED THAT THEY JUST HAD A RECALL OF SOME OF THE MACHINES BECAUSE OF ISSUES WITH THE BLOOD PUMP SEGMENT. OTHER DEVICES USED: FRESENIUS 2008T.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16164893 ↗MalfunctionJan 13, 2023Material Puncture/Hole; Material Puncture/Hole
DURING THE 1ST MINUTE OF TREATMENT, A PHYSICAL LEAK WAS NOTED ON THE BLOOD PUMP, TREATMENT STOPPED, BLOOD WAS NOT RETURNED. NEW SET UP DONE USING SAME LOT #, DIFFERENT MACHINE AND FINISHED TREATMENT WITH NO ISSUES. PATIENT LOST 200ML OF BLOOD. NO HOSPITALIZATION. HBG RECHECKED, NO DATA ON BASELINE HGB AND NO RESULT YET. IV ANTIBIOTICS STARTED AS A PROPHYLAXIS. THE BIOMED ALSO CHECKED THE MACHINE AND ORDERED SOME PARTS OF THE BLOOD PUMP SEGMENT. THE MANAGER ALSO MENTIONED THAT THEY JUST HAD A RECALL OF SOME OF THE MACHINES BECAUSE OF ISSUES WITH THE BLOOD PUMP SEGMENT. OTHER DEVICES USED: FRESENIUS 2008T.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16164561 ↗InjurySep 15, 2022Fluid/Blood Leak; Fluid/Blood Leak
PATIENT WAS 3 HOUR 15 MINUTES INTO A 4 HOUR HD TREATMENT WHEN STAFF STATED THEY OBSERVED BLOOD LEAKING FROM THE BLOOD PUMP SEGMENT ON THE MACHINE. REPORTED THAT HD TREATMENT WAS STOPPED AND THE BLOODLINE AND DIALYZER WAS THROWN AWAY. BLOOD WAS NOT RETURNED TO THE PATIENT. SHE REPORTED THAT THE PATIENT WAS STABLE DURING AND AFTER THE EVENT. PATIENT DID NOT REQUIRE ANY FURTHER MEDICAL ATTENTION RELATED TO THIS ISSUE. STAFF DID NOT RETAIN THE BLOODLINE FOR RETURN TO US.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15428111 ↗MalfunctionSep 15, 2022Fluid/Blood Leak; Fluid/Blood Leak
PATIENT WAS 3 HOUR 15 MINUTES INTO A 4 HOUR HD TREATMENT WHEN STAFF STATED THEY OBSERVED BLOOD LEAKING FROM THE BLOOD PUMP SEGMENT ON THE MACHINE. REPORTED THAT HD TREATMENT WAS STOPPED AND THE BLOODLINE AND DIALYZER WAS THROWN AWAY. BLOOD WAS NOT RETURNED TO THE PATIENT. SHE REPORTED THAT THE PATIENT WAS STABLE DURING AND AFTER THE EVENT. PATIENT DID NOT REQUIRE ANY FURTHER MEDICAL ATTENTION RELATED TO THIS ISSUE. STAFF DID NOT RETAIN THE BLOODLINE FOR RETURN TO US.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15427956 ↗InjuryMay 9, 2018Obstruction of Flow; Device Operational Issue; Obstruction of Flow; Device Operational Issue
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
Patient problems: Diaphoresis; Syncope
MAUDE report 7499625 ↗
Other SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE devices
- Hemodialysis Bloodlines SL-2010M2096B. Braun Medical Inc. · Set, tubing, blood, with and without anti-regurgitation valve3 recalls3 deaths1,000 reports
- Hemodialysis Bloodlines SL-2010M2096AB. Braun Medical Inc. · Set, tubing, blood, with and without anti-regurgitation valve2 recalls1 death483 reports
- Hemodialysis Bloodlines SL-2000M2095DAB. Braun Medical Inc. · Set, tubing, blood, with and without anti-regurgitation valve2 recalls1 death214 reports
- CombiSet 03-2722-9Fresenius Medical Care Holdings, Inc. · SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE8 deaths311 reports
- Hemodialysis Bloodlines SL-2000M2095B. Braun Medical Inc. · Set, tubing, blood, with and without anti-regurgitation valve1 recall219 reports
- CombiSet 03-2522-1Fresenius Medical Care Holdings, Inc. · Set, Tubing, Blood, With And Without Anti-Regurgitation Valve1 recall214 reports
- Cartridge Blood Set 101025Gambro Renal Products, S.A. de C.V. · SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE3 recalls80 reports
- Hemodialysis Bloodlines SL-2000M2095DB. Braun Medical Inc. · Set, tubing, blood, with and without anti-regurgitation valve1 recall158 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page