Skip to content

Cardioroot HEWROOT0028Recalls & reported problems

Intervascular Sas · Cardioroot Woven Collagen Coated Vascular Graft · Class 2 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00384401013907), received between Jan 13, 2016 and May 19, 2025 — classified as 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Material Frayed” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 12, 2026Open, Classified

    Product failed to meet the specified acceptance criteria for Integral Water Permeability (IWP).

    Z-3258-2026 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Material Frayed2
  • Product Quality Problem2
  • Device Damaged Prior to Use1
  • Flaked1
  • Manufacturing, Packaging or Shipping Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Failure to Anastomose2
  • Hemorrhage/Blood Loss/Bleeding2
  • Blood Loss1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 19, 2025Material Frayed; Material Frayed

    IT WAS REPORTED TO INTERVASCULAR THAT UPON TAKEN THE GRAFT OUT OF THE PACKAGING WITHOUT PRODUCT HANDLING PRIOR USE, IT WAS NOTICED THAT PART OF THE COLLAR OF THE GRAFT WAS FRAYED. THE PROSTHESIS WAS IMPLANTED AND THE OPERATION WENT WELL. NO INJURIES WERE REPORTED ON THE PATIENT. COMPLAINT#: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22043701 ↗
  • MalfunctionJan 31, 2025Flaked; Material Frayed; Flaked; Material Frayed

    IT WAS REPORTED TO INTERVASCULAR THAT DURING THE SURGICAL PROCEDURE, WHEN THE CARDIOROOT GRAFT WAS GOING TO BE SEWN IN AFTER CUTTING THE NECK AREA AFTER THE BULGE, THE SURGEON NOTICED CLUMPS OF THE GRAFT COATING PEELING FROM THE GRAFT ONTO GLOVES. THE GRAFT WAS TAKEN OFF THE FIELD AND NOT IMPLANTED PER SURGEON DECISION. THE PROCEDURE WAS SLIGHTLY DELAYED TO GET ANOTHER GRAFT (A PLATINUM 28MM STRAIGHT GRAFT) TO FINISH SURGERY. NO HARM TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21282059 ↗
  • InjuryOct 13, 2022Product Quality Problem; Product Quality Problem

    IT WAS REPORTED TO INTERVASCULAR FROM THE HEALTH FACILITY THAT: "THERE WAS AN INCIDENT DURING A BENTALL PROCEDURE: A CARDIOROOT PROSTHESIS WAS PUT IN PLACE. THE PROXIMAL PART OF THE TUBE WHERE THE CORONARIES ARE REIMPLANTED WAS DEEMED TO BE TOO POROUS AND FRAGILE, WHICH LED TO SMALL TEARS IN THE WIRES. HEMOSTASIS WAS VERY COMPLICATED AT THE END OF THE PROCEDURE WITH A REFITTING OF THE CEC, RE-CLAMPING OF THE AORTA AND THE MULTIPLE USE OF COAGULATION FACTORS AND BIOGLUE IN PARTICULAR. ON (B)(6) 2022, THE INSTITUTION INDICATED THAT THE PATIENT HAS BEEN ADMITTED TO THE INTENSIVE CARE UNIT. THE DEVICE WILL NOT BE AVAILABLE FOR EXPERTISE (IT REMAINED ON THE PATIENT)." COMPLAINT #(B)(4).

    Patient problems: Failure to Anastomose; Hemorrhage/Blood Loss/Bleeding

    MAUDE report 15596615 ↗
  • InjuryOct 10, 2022Product Quality Problem; Product Quality Problem

    IT WAS REPORTED TO INTERVASCULAR FROM THE HEALTH FACILITY THAT: "THERE WAS AN INCIDENT DURING A BENTALL PROCEDURE: A CARDIOROOT PROSTHESIS WAS PUT IN PLACE. THE PROXIMAL PART OF THE TUBE WHERE THE CORONARIES ARE REIMPLANTED WAS DEEMED TO BE TOO POROUS AND FRAGILE, WHICH LED TO SMALL TEARS IN THE WIRES. HEMOSTASIS WAS VERY COMPLICATED AT THE END OF THE PROCEDURE WITH A REFITTING OF THE CEC, RE-CLAMPING OF THE AORTA AND THE MULTIPLE USE OF COAGULATION FACTORS AND BIOGLUE IN PARTICULAR. ON (B)(6) 2022, THE INSTITUTION INDICATED THAT THE PATIENT HAS BEEN ADMITTED TO THE INTENSIVE CARE UNIT. THE DEVICE WILL NOT BE AVAILABLE FOR EXPERTISE (IT REMAINED ON THE PATIENT)." (B)(4).

    Patient problems: Failure to Anastomose; Hemorrhage/Blood Loss/Bleeding

    MAUDE report 15571350 ↗
  • MalfunctionMay 3, 2021Device Damaged Prior to Use; Device Damaged Prior to Use

    THIS IS A CASE OF A PRODUCT DAMAGED DURING TRANSPORTATION. THE CARRIER (DHL) PUT A BOX WITH WATER (WHO HAD LEECHES INSIDE) IN THE SAME TRANSPORTATION OF OUR PRODUCT. THIS BOX BROKE OPEN AND ALL THE LIQUID CAME ON THE CARDIOROOT PRODUCT. THEREFORE THE PRODUCT WAS DAMAGED WHEN IT ARRIVED AT THE END CUSTOMER. THE PRODUCT HAS THEN BEEN RETURNED TO THE MANUFACTURER. COMPLAINT #(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11764682 ↗
  • InjuryJan 13, 2016Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    DURING AN AORTIC REPAIR VALVE REPLACEMENT PERFORMED ON (B)(6) 2015, SUTURE HOLE BLEEDING WAS NOTICED, WHICH REQUIRED THE USE OF HEMOSTATIC AGENT. THE GRAFT REMAINED IMPLANTED. THE PATIENT WAS REPORTED AS PROCESSING WELL.

    Patient problems: Blood Loss

    MAUDE report 5363396 ↗

Other Prosthesis, Vascular Graft, Of 6mm And Greater Diameter devices