Skip to content

FLIXENE SLIDER GDS 25137Recalls & reported problems

Atrium Medical Corporation · FLIXENE 4-6MMX45CM GWT-GW, W/DBL GDS · Class 2 · Implant

All models of FLIXENE SLIDER GDS (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00650862251371), received between Dec 21, 2018 and Mar 18, 2026 — classified as 3 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Moisture or Humidity Problem” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 3, 2024Open, Classified

    Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.

    Z-2016-2024 · Root cause: Equipment maintenance · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Moisture or Humidity Problem2
  • Complete Blockage1
  • Device Damaged by Another Device1
  • Material Disintegration1
  • Material Fragmentation1
  • Material Integrity Problem1
  • Material Too Rigid or Stiff1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Occlusion2
  • Aneurysm1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 18, 2026Moisture or Humidity Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24632652 ↗
  • MalfunctionMar 18, 2026Moisture or Humidity Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24626674 ↗
  • MalfunctionJun 26, 2025Material Too Rigid or Stiff; Material Too Rigid or Stiff

    IT WAS REPORTED THAT UPON OPENING THE FLIXENE GRAFT THE CUSTOMER EXPERIENCED AN UNUSUAL HARDNESS COMPARED TO PREVIOUSLY USED PRODUCTS. THE SURGEON DID NOT FEEL COMFORTABLE USING THE DEVICE SO HE SWITCHED THE DEVICE TO A STRAIGHT 6MM GRAFT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22341497 ↗
  • MalfunctionJun 27, 2024Material Fragmentation; Material Fragmentation

    DR. (B)(6) REPORTED THAT HE HAD EXPERIENCED SEVERAL CASES WHERE THE VASCULAR MARKINGS HAD DISAPPEARED AND THE OUTER WALL HAD GOTTEN PEELED OFF AFTER IMPLANTING A FLIXENE VASCULAR GRAFT USING A SHEATH-TYPE TUNNELER. (IT WAS CONFIRMED THAT THERE WERE NO PROBLEMS BEFORE TUNNELING.) AT PRESENT, THERE HAVE BEEN NO PROBLEMS AFTER SURGERY, AND THE PROGRESS IS GOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19630576 ↗
  • InjuryAug 22, 2021Material Disintegration; Material Integrity Problem; Material Integrity Problem

    REPORT RECEIVED STATED THAT A FLIXENE GRAFT THAT WAS PLACED APPROXIMATELY 12 MONTHS AGO IN CONJUNCTION WITH A HERO GRAFT WAS EXPLANTED. THE PHYSICIAN BELIEVES THE GRAFT HAS DISINTEGRATED. ADDITIONAL INFORMATION RECEIVED STATED THE FLIXENE GRAFT WAS REMOVED DUE TO FALSE ANEURYSMS FORMING ALONG THE GRAFT.

    Patient problems: Aneurysm

    MAUDE report 12352032 ↗
  • MalfunctionMay 4, 2021Device Damaged by Another Device; Adverse Event Without Identified Device or Use Problem; Device Damaged by Another Device; Adverse Event Without Identified Device or Use Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11768758 ↗
  • InjuryOct 22, 2019Complete Blockage

    IT WAS REPORTED THAT TWO DAYS AFTER IMPLANT THE GRAFT BECAME OCCLUDED. THE GRAFT HAD TO BE EXPLANTED 3 WEEKS AFTER IMPLANT.

    Patient problems: Occlusion

    MAUDE report 9219651 ↗
  • InjuryDec 21, 2018Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS IMPLANTED THE FLIXENE (SHORT-TAPERED) DEVELOPED VASCULAR OCCLUSION AND HAS BEEN RECEIVING PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY REPEATEDLY. THE FIRST VASCULAR OCCLUSION OCCURRED AT THE ANASTOMOSIS AREA ON THE ARTERIAL SIDE, AND THEN, ALSO OCCURRED AT THE ANASTOMOSIS AREA ON THE VENOUS SIDE. THIS IS THE FIRST AVG FOR THE PATIENT.

    Patient problems: Occlusion

    MAUDE report 8192389 ↗

Other PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER devices

All PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER devices →

Sources and updates