Skip to content

FLIXENE SLIDER GDS 25058Recalls & reported problems

Atrium Medical Corporation · 6MMX50CM FLIXENE GRAFT, GW, W/GDS · Class 2 · Implant

All models of FLIXENE SLIDER GDS (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
13
Deaths reported
0
Injuries reported
7
Recalls naming it
1

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00650862250589), received between Jan 31, 2017 and Mar 17, 2026 — classified as 7 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Complete Blockage” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 3, 2024Open, Classified

    Complaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.

    Z-2002-2024 · Root cause: Equipment maintenance · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Appropriate Device Problem Term/Code Not Available2
  • Complete Blockage2
  • Obstruction of Flow2
  • Activation, Positioning or Separation Problem1
  • Improper or Incorrect Procedure or Method1
  • Kinked1
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Insufficient Information2
  • Blister1
  • Embolism1
  • Failure to Anastomose1
  • Local Reaction1
  • Thrombosis/Thrombus1
  • Thrombus1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 17, 2026Moisture or Humidity Problem

    IT WAS REPORTED THAT FLIXENE GRAFT HAD HUMIDITY LEVELS EXCEEDED 80% FOR MORE THAN 24 HOURS AND REACHED UP TO 100%. DURING INSPECTION, BUBBLING WAS OBSERVED ON THE PRODUCT LABELING, AND ALL PRODUCTS FELT COLD AS IF LIKE THEY WERE DEFROSTED. THERE WAS NO PATIENT INVOLVED OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24619952 ↗
  • MalfunctionNov 18, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE PHYSICIAN DID NOT PERCEIVE THE GRADUATED WALL FEATURE ON THE FLIXENE GRAFT, AS THE WALL THICKNESS DID NOT APPEAR REDUCED TOWARD THE END DURING HANDLING. THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23580161 ↗
  • MalfunctionNov 18, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23579682 ↗
  • MalfunctionJun 27, 2024Material Fragmentation; Material Fragmentation

    (B)(6) REPORTED THAT HE HAD EXPERIENCED SEVERAL CASES WHERE THE VASCULAR MARKINGS HAD DISAPPEARED AND THE OUTER WALL HAD GOTTEN PEELED OFF AFTER IMPLANTING A FLIXENE VASCULAR GRAFT USING A SHEATH-TYPE TUNNELER. (IT WAS CONFIRMED THAT THERE WERE NO PROBLEMS BEFORE TUNNELING.) AT PRESENT, THERE HAVE BEEN NO PROBLEMS AFTER SURGERY, AND THE PROGRESS IS GOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19630594 ↗
  • MalfunctionApr 11, 2024Material Separation; Material Separation

    DURING THE PROCEDURE THE DOCTOR WAS SUTURING IN FLIXENE GRAFT WHEN HE MENTIONED THAT THE OUTER LAYER OF THE GRAFT HAD APPEARED TO HAVE SEPARATED. HE ALSO COMMENTED THAT THE FLIXENE GRAFT HE IMPLANTED THE WEEK PRIOR ALSO HAD THE SAME ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19088535 ↗
  • MalfunctionAug 29, 2023Material Puncture/Hole; Material Puncture/Hole

    GOT A CALL TODAY FROM CUSTOMER AT (B)(6) STATING A FLIXENE GRAFT HAD HOLES IN IT AND LEAKED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17647558 ↗
  • InjuryNov 19, 2021Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12843490 ↗
  • InjuryApr 6, 2021Patient-Device Incompatibility; Patient-Device Incompatibility

    N/A.

    Patient problems: Blister; Local Reaction

    MAUDE report 11625546 ↗
  • InjuryFeb 17, 2021Complete Blockage; Adverse Event Without Identified Device or Use Problem

    N/A.

    Patient problems: Insufficient Information

    MAUDE report 11337265 ↗
  • InjuryFeb 17, 2021Obstruction of Flow; Adverse Event Without Identified Device or Use Problem

    N/A.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 11337262 ↗
  • InjuryFeb 13, 2021Adverse Event Without Identified Device or Use Problem; Obstruction of Flow; Adverse Event Without Identified Device or Use Problem

    N/A.

    Patient problems: Thrombus

    MAUDE report 11325950 ↗
  • InjuryJan 31, 2021Adverse Event Without Identified Device or Use Problem; Complete Blockage; Adverse Event Without Identified Device or Use Problem

    APPROXIMATELY 45 DAYS AFTER FLIXENE WAS INSERTED, BLOOD CLOTS WERE FORMED IN THE VEINS OF AUTOGENOUS VASCULAR WHICH CAUSED BLOCKAGE. THE PATIENT RETURNED TO THE EMERGENCY ROOM OF THE HOSPITAL FOR REVASCULARIZATION. THE BLOCKAGE OCCURRED AGAIN SHORTLY AFTER THE REVASCULARIZATION. THUS, THE PATIENT RETURNED TO THE HOSPITAL AGAIN.

    Patient problems: Embolism

    MAUDE report 11256702 ↗
  • InjuryJan 31, 2017Kinked; Improper or Incorrect Procedure or Method

    REPORT RECEIVED STATED THE FLIXENE GRAFT WAS PLACED FOR DIALYSIS IN A PATIENT BUT THE ANASTOMOSIS WAS NOT DONE CORRECTLY. THE GRAFT WAS CUT TOO SHORT AND KINKED. PATIENT WAS GIVEN A CATHETER FOR DIALYSIS.

    Patient problems: Failure to Anastomose

    MAUDE report 6289951 ↗

Other PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER devices

All PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER devices →