Hemashield M002020951080Recalls & reported problems
Intervascular Sas · Hemashield Gold Knitted Microvel Straight Double Velour Vascular Graft with Graft Sizer Accessory · Class 2 · Implant
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00384401017363), received between Apr 15, 2016 and Feb 6, 2017 — classified as 2 injuries.
The most frequently reported specific device problem is “Manufacturing, Packaging or Shipping Problem” (2 reports).
6 other recalls cover devices cleared under the same FDA submission (K954848); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Manufacturing, Packaging or Shipping Problem2
Patient problems
- Blood Loss2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 6, 2017Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
DURING A CARDIAC SURGERY PERFORMED ON (B)(6) 2017, THE INVOLVED GRAFT WAS USED TO CANNULATE THE AXILLARY ARTERY FOR CARDIOPULMONARY BYPASS. THE GRAFT WAS SOAKED WITH RIFAMPICIN BEFORE THE CASE FOR ABOUT 10 MINUTES. IT WAS REPORTED THAT EXCESSIVE WEEPING/BLEEDING THROUGH THE GRAFT WAS NOTICED DURING THE PROCEDURE.
Patient problems: Blood Loss
MAUDE report 6303630 ↗InjuryApr 15, 2016Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
DURING SURGERY PERFORMED ON (B)(6) 2016, THE GRAFT WAS USED AS A CONDUIT FOR A BYPASS FOR AORTIC ROOT REPLACEMENT. THE GRAFT WAS REPORTED TO WEEP, AND SURGICAL GLUE WAS USED TO STOP THE WEEPING. NO ADDITIONAL INFORMATION COULD BE OBTAINED AT THE DATE OF THIS REPORT.
Patient problems: Blood Loss
MAUDE report 5580347 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K954848. One submission often covers many models, so check the product and lot details in each record.
- Z-0729-2022 · Jan 14, 2022 · TerminatedHEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm; L: 60 cm Indicated for use in the replacement or repair of arteries affected with aneurysmal or occlusive
- Z-0403-2022 · Dec 1, 2021 · TerminatedHemashield Gold Knitted Bifurcated, 18 mm x 9 mm; reference number M002020851890
- Z-1859-2020 · Apr 7, 2020 · TerminatedHemashield Gold Knitted Microvel Double Velour Vascular Graft-Axillo-bifemoral, Product Code: M002020955010 - Product Usage: For use in the replacement or repair of arteries affect
- Z-1861-2020 · Apr 7, 2020 · TerminatedHemashield Gold Knitted Microvel Double Velour Vascular Graft Axillo-bifemoral, Product Code: M002020955890 - Product Usage: for use in the replacement or repair of arteries affect
- Z-1860-2020 · Apr 7, 2020 · TerminatedHemashield Gold Knitted Microvel Double Velour Vascular Graft-Axillo-bifemoral, Product Code: M002020955880 - Product Usage: for use in the replacement or repair of arteries affect
- Z-2066-2010 · May 24, 2010 · TerminatedHemashield Gold Woven and Knitted Double Vascular Grafts.
Other Graft, Vascular, Synthetic/Biologic Composite devices
- Hemashield M00202166168P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite1 recall3 reports
- Hemashield M002020851890Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite1 recall2 reports
- Hemashield M002020952070Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite1 recall0 reports
- Hemashield M00202175928P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite10 reports
- Hemashield M00202175208P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite8 reports
- Hemashield M00202175210P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite8 reports
- AlboGraft Polyester Vascular Graft AMC1407Lemaitre Vascular, Inc. · Graft, Vascular, Synthetic/Biologic Composite6 reports
- Hemashield M00202175430P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite6 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.