Skip to content

Hemashield M002020952160Recalls & reported problems

Intervascular Sas · Hemashield Gold Knitted Microvel Straight Double Velour Vascular Graft with Graft Sizer Accessory · Class 2 · Implant

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00384401017516), received between Aug 1, 2025 and Aug 1, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Handling Problem” (1 report).

6 other recalls cover devices cleared under the same FDA submission (K954848); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Handling Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 1, 2025Device Handling Problem; Device Handling Problem

    IT WAS REPORTED TO INTERVASCULAR THAT ON (B)(6) 2025 DURING A REPAIR OF RUPTURED MYCOTIC ANEURYSM OF THORACIC AND ABDOMINAL AORTA, USING 16MM RIFAMPIN SOAKED HEMAHSIELD GRAFT, THE CONCERNED GRAFT WAS INCORRECTLY OPENED AND HANDLED ONTO THE STERILE FIELD WITH THE NON-STERILE PACKAGING, DUE TO MISINTERPRETATION OF PACKAGING. THE SEQUENCE OF EVENTS WAS CONFIRMED AS FOLLOWS: "THE CLEAR PLASTIC ANTIMOISTURE POUCH THAT HAS THE YELLOW WARNING LABEL "POUCH IS NOT A STERILE BARRIER" WAS OPENED OUTSIDE THE STERILE FIELD. HOWEVER, THE NON-STERILE OUTER TRAY PACKAGE WAS PLACED IN THE STERILE FIELD. THE STERILE INNER TRAY PACKAGE WAS PLACED IN THE STERILE FIELD AND GRAFT ITSELF WAS STERILE." THIS PATIENT WAS ALREADY UNDER ANTIBIOTICS TREATMENT AND NEEDED TO CONTINUE ANTIBIOTICS DUE TO THE PREOPERATIVE DIAGNOSIS OF MYCOTIC ANEURYSM REGARDLESS OF THE CONTAMINATION ERROR MADE BY THE NURSE. NO PROCEDURAL DELAY WAS REPORTED. THE PATIENT¿S MEDICAL FOLLOW-UP IS PLANNED AT 1, 6 AND 12 MONTHS. AS OF TODAY, NO HARM TO THE PATIENT WAS REPORTED. THE CUSTOMER IS A LONG TIME USER OF INTERVASCULAR PRODUCTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22684576 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K954848. One submission often covers many models, so check the product and lot details in each record.

  • Z-0729-2022 · Jan 14, 2022 · Terminated
    HEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm; L: 60 cm Indicated for use in the replacement or repair of arteries affected with aneurysmal or occlusive
  • Z-0403-2022 · Dec 1, 2021 · Terminated
    Hemashield Gold Knitted Bifurcated, 18 mm x 9 mm; reference number M002020851890
  • Z-1859-2020 · Apr 7, 2020 · Terminated
    Hemashield Gold Knitted Microvel Double Velour Vascular Graft-Axillo-bifemoral, Product Code: M002020955010 - Product Usage: For use in the replacement or repair of arteries affect
  • Z-1861-2020 · Apr 7, 2020 · Terminated
    Hemashield Gold Knitted Microvel Double Velour Vascular Graft Axillo-bifemoral, Product Code: M002020955890 - Product Usage: for use in the replacement or repair of arteries affect
  • Z-1860-2020 · Apr 7, 2020 · Terminated
    Hemashield Gold Knitted Microvel Double Velour Vascular Graft-Axillo-bifemoral, Product Code: M002020955880 - Product Usage: for use in the replacement or repair of arteries affect
  • Z-2066-2010 · May 24, 2010 · Terminated
    Hemashield Gold Woven and Knitted Double Vascular Grafts.

Other Graft, Vascular, Synthetic/Biologic Composite devices

All Graft, Vascular, Synthetic/Biologic Composite devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.