Hemashield M002020952100Recalls & reported problems
Intervascular Sas · Hemashield Gold Knitted Microvel Straight Double Velour Vascular Graft with Graft Sizer Accessory · Class 2 · Implant
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00384401017622), received between Mar 14, 2022 and Mar 17, 2022 — classified as 2 injuries.
The most frequently reported specific device problem is “Off-Label Use” (2 reports).
6 other recalls cover devices cleared under the same FDA submission (K954848); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Off-Label Use2
Patient problems
- Hemorrhage/Blood Loss/Bleeding2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMar 17, 2022Off-Label Use; Off-Label Use
SEE MFG INITIAL REPORT #1640201-2022-00007. COMPLAINT #(B)(4).
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 13792981 ↗InjuryMar 14, 2022Off-Label Use; Off-Label Use
THIS COMPLAINT INFORMATION WAS RECEIVED THROUGH A FDA MEDWATCH REPORT AND WAS FORWARDED TO THE MANUFACTURER ON 21 FEB 2022. ACCORDING TO THE MEDWATCH REPORT, COMPLAINT IS AS FOLLOWS: "DESCRIBE THE EVENT OR PROBLEM: HEMASHIELD GOLD KNITTED MICROVEL STRAIGHT DOUBLE VELOUR VASCULAR GRAFT FAILED WHILE CANNULATING FOR VENO-ARTERIAL EXTRACORPOREAL MEMBRANE OXYGENATION (VA-ECMO). REPLACED WITH NEW GRAFT. WHAT WAS THE ORIGINAL INTENDED PROCEDURE?: VENO-ARTERIAL EXTRACORPOREAL MEMBRANE OXYGENATION (VA-ECMO) SUPPORT. WHAT PROBLEM DID THE USER HAVE (CHECK ALL THAT APPLY): DEVICE FAILED (E.G. BROKE, COULDN'T GET IT TO WORK OR STOPPED WORKING);" TO BE NOTED, THE USE OF HEMASHIELD GOLD KNITTED GRAFT FOR CANNULATION IN VENO-ARTERIAL EXTRACORPOREAL MEMBRANE OXYGENATION (VA-ECMO) SUPPORT IS NOT AN APPROVED INDICATION. AFTER CONTACTING THE INITIAL REPORTER, WE OBTAINED THE FOLLOWING INFORMATION ABOUT THE EVENT: ACCORDING TO THE MEDICAL RECORD, THE PATIENT EXPERIENCE BLEEDING FROM THE RIGHT AXILLARY ARTERIAL GRAFT SITE. THE MEDICAL RECORD DOES NOT INDICATE THE AMOUNT OF BLOOD LOSS. IT IS UNKNOWN IF THE PROCEDURE WAS SIGNIFICANTLY PROLONGED. ACCORDING TO THE MEDICAL RECORD, THE PATIENT WAS ON BRILINTA
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 13755715 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K954848. One submission often covers many models, so check the product and lot details in each record.
- Z-0729-2022 · Jan 14, 2022 · TerminatedHEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm; L: 60 cm Indicated for use in the replacement or repair of arteries affected with aneurysmal or occlusive
- Z-0403-2022 · Dec 1, 2021 · TerminatedHemashield Gold Knitted Bifurcated, 18 mm x 9 mm; reference number M002020851890
- Z-1859-2020 · Apr 7, 2020 · TerminatedHemashield Gold Knitted Microvel Double Velour Vascular Graft-Axillo-bifemoral, Product Code: M002020955010 - Product Usage: For use in the replacement or repair of arteries affect
- Z-1861-2020 · Apr 7, 2020 · TerminatedHemashield Gold Knitted Microvel Double Velour Vascular Graft Axillo-bifemoral, Product Code: M002020955890 - Product Usage: for use in the replacement or repair of arteries affect
- Z-1860-2020 · Apr 7, 2020 · TerminatedHemashield Gold Knitted Microvel Double Velour Vascular Graft-Axillo-bifemoral, Product Code: M002020955880 - Product Usage: for use in the replacement or repair of arteries affect
- Z-2066-2010 · May 24, 2010 · TerminatedHemashield Gold Woven and Knitted Double Vascular Grafts.
Other Graft, Vascular, Synthetic/Biologic Composite devices
- Hemashield M00202166168P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite1 recall3 reports
- Hemashield M002020851890Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite1 recall2 reports
- Hemashield M002020952070Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite1 recall0 reports
- Hemashield M00202175928P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite10 reports
- Hemashield M00202175208P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite8 reports
- Hemashield M00202175210P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite8 reports
- AlboGraft Polyester Vascular Graft AMC1407Lemaitre Vascular, Inc. · Graft, Vascular, Synthetic/Biologic Composite6 reports
- Hemashield M00202175430P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite6 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.