Hemashield M002020851260Recalls & reported problems
Intervascular Sas · Hemashield Gold Knitted Microvel Bifurcated Double Velour Vascular Graft with Graft Sizer Accessory · Class 2 · Implant
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00384401017684), received between Mar 1, 2019 and Jun 18, 2019 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Flaked” (1 report).
6 other recalls cover devices cleared under the same FDA submission (K954848); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Flaked1
- Manufacturing, Packaging or Shipping Problem1
Patient problems
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 18, 2019Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
IT WAS REPORTED THAT THE GRAFT INSIDE THE PRODUCT BOX DID NOT MATCH THE BOX LABELING. THE PRODUCT WAS NEVER OPENED OR IMPLANTED.
Patient problems: No Patient Involvement
MAUDE report 8707774 ↗MalfunctionMar 1, 2019Flaked; Flaked
TWO VASCULAR GRAFTS WERE SUTURED TOGETHER AND WERE USED ON THE SAME PATIENT (SEE MANUFACTURER REPORT 1640201-2019-00013 FOR THE OTHER INVOLVED DEVICE). WHEN SUTURING THE GRAFTS TO BE IMPLANTED, IT WAS NOTICED THAT THE COLLAGEN COATING WAS CRUMBLING LIKE BREAD CRUMBS. THE SURGEON WET THE GRAFTS WITH STERILE SALINE AND NO MORE CRUMBLING WAS NOTICED. THE GRAFTS WERE IMPLANTED IN THE ABDOMINAL AREA WITH NO OTHER SIGNS OF ISSUES OR COMPLICATIONS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8382221 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K954848. One submission often covers many models, so check the product and lot details in each record.
- Z-0729-2022 · Jan 14, 2022 · TerminatedHEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm; L: 60 cm Indicated for use in the replacement or repair of arteries affected with aneurysmal or occlusive
- Z-0403-2022 · Dec 1, 2021 · TerminatedHemashield Gold Knitted Bifurcated, 18 mm x 9 mm; reference number M002020851890
- Z-1859-2020 · Apr 7, 2020 · TerminatedHemashield Gold Knitted Microvel Double Velour Vascular Graft-Axillo-bifemoral, Product Code: M002020955010 - Product Usage: For use in the replacement or repair of arteries affect
- Z-1861-2020 · Apr 7, 2020 · TerminatedHemashield Gold Knitted Microvel Double Velour Vascular Graft Axillo-bifemoral, Product Code: M002020955890 - Product Usage: for use in the replacement or repair of arteries affect
- Z-1860-2020 · Apr 7, 2020 · TerminatedHemashield Gold Knitted Microvel Double Velour Vascular Graft-Axillo-bifemoral, Product Code: M002020955880 - Product Usage: for use in the replacement or repair of arteries affect
- Z-2066-2010 · May 24, 2010 · TerminatedHemashield Gold Woven and Knitted Double Vascular Grafts.
Other Graft, Vascular, Synthetic/Biologic Composite devices
- Hemashield M00202166168P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite1 recall3 reports
- Hemashield M002020851890Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite1 recall2 reports
- Hemashield M002020952070Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite1 recall0 reports
- Hemashield M00202175928P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite10 reports
- Hemashield M00202175208P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite8 reports
- Hemashield M00202175210P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite8 reports
- AlboGraft Polyester Vascular Graft AMC1407Lemaitre Vascular, Inc. · Graft, Vascular, Synthetic/Biologic Composite6 reports
- Hemashield M00202175430P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite6 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.