Hemashield M002020851270Recalls & reported problems

Intervascular Sas · Hemashield Gold Knitted Microvel Bifurcated Double Velour Vascular Graft with Graft Sizer Accessory · Class 2 · Implant

All models of Hemashield (51)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00384401017691), received between Jan 18, 2019 and Jan 18, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Manufacturing, Packaging or Shipping Problem” (1 report).

6 other recalls cover devices cleared under the same FDA submission (K954848); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Manufacturing, Packaging or Shipping Problem1

Patient problems

  • No Patient Involvement1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 18, 2019Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    STAFF OPENED THE BOX AND NOTICED THAT THE GRAFT INSIDE DID NOT MATCH THE BOX LABELING. THEY THOUGHT THEY WERE OPENING A GRAFT WHOSE DIMENSIONS WERE 12MMX7MMX40CM, INSTEAD THEY FOUND A GRAFT WHOSE DIMENSIONS WERE 14MMX7MMX40CM. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT WAS INVOLVED. IT WAS REPORTED THAT THE EVENT OCCURRED SOME TIME AGO BUT THERE WAS A REPORTING OMISSION. THE EVENT DATE IS UNKNOWN.

    Patient problems: No Patient Involvement

    MAUDE report 8260461 ↗

FDA clearance and classification

Cleared through 510(k) K954848 on Feb 1, 1996 (decision: Substantially Equivalent - Subject to Tracking Reg.). Submitted by Meadox Medicals, Div. Boston Scientific Corp. as “Hemashield Microvel Velour Knitted Vascular Graft”. The FDA review took 101 days. FDA record ↗

The FDA regulates product code MAL (graft, vascular, synthetic/biologic composite) under 21 CFR 870.3450 as a Class II device. 21 CFR 870.3450 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K954848. One submission often covers many models, so check the product and lot details in each record.

  • Z-0729-2022 · Jan 14, 2022 · Terminated
    HEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm; L: 60 cm Indicated for use in the replacement or repair of arteries affected with aneurysmal or occlusive di
  • Z-0403-2022 · Dec 1, 2021 · Terminated
    Hemashield Gold Knitted Bifurcated, 18 mm x 9 mm; reference number M002020851890
  • Z-1860-2020 · Apr 7, 2020 · Terminated
    Hemashield Gold Knitted Microvel Double Velour Vascular Graft-Axillo-bifemoral, Product Code: M002020955880 - Product Usage: for use in the replacement or repair of arteries affect
  • Z-1861-2020 · Apr 7, 2020 · Terminated
    Hemashield Gold Knitted Microvel Double Velour Vascular Graft Axillo-bifemoral, Product Code: M002020955890 - Product Usage: for use in the replacement or repair of arteries affect
  • Z-1859-2020 · Apr 7, 2020 · Terminated
    Hemashield Gold Knitted Microvel Double Velour Vascular Graft-Axillo-bifemoral, Product Code: M002020955010 - Product Usage: For use in the replacement or repair of arteries affect
  • Z-2066-2010 · May 24, 2010 · Terminated
    Hemashield Gold Woven and Knitted Double Vascular Grafts.

Common questions about the Hemashield M002020851270

Has the Hemashield M002020851270 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00384401017691), based on the FDA recall file dated Nov 27, 2026. However, 6 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the Hemashield M002020851270?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Jan 18, 2019. That report lists device problem “Manufacturing, Packaging or Shipping Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Hemashield M002020851270?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Hemashield M002020851270 FDA approved?

It was cleared through 510(k) K954848 on Feb 1, 1996. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Hemashield M002020851270 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.

Who makes the Hemashield M002020851270, and is it still sold?

The FDA's device database lists Intervascular Sas as the company responsible for it (brand Hemashield, model M002020851270). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Graft, Vascular, Synthetic/Biologic Composite devices

All Graft, Vascular, Synthetic/Biologic Composite devices →

Sources and updates