Hemashield M002020852010Recalls & reported problems
Intervascular Sas · Hemashield Gold Knitted Microvel Bifurcated Double Velour Vascular Graft with Graft Sizer Accessory · Class 2 · Implant
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00384401017776), received between Apr 12, 2019 and Sep 28, 2026 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Nonstandard Device” (2 reports).
6 other recalls cover devices cleared under the same FDA submission (K954848); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Nonstandard Device2
- Device Contaminated During Manufacture or Shipping1
- Stretched1
Patient problems
- Hemorrhage/Blood Loss/Bleeding2
- Insufficient Information1
- No Patient Involvement1
- Thrombosis/Thrombus1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Stretched
It was reported to Intervascular that on (b)(6) 2021, an abdominal aortic graft replacement surgery was performed using Hemashield.A 20mm portion of the 20mm x 10mm graft was used to perform an I-shaped replacement in the area from the infrarenal artery to the terminal aorta. The distal end of the I-shaped graft was anastomosed to the left common iliac artery, and a Propaten 8mm ring, end-to-end anastomosed to the I-shaped portion, was anastomosed to the right external iliac artery. One week after the replacement surgery, on (b)(6) 2021, the graft had expanded to 28mm (the entire 20mm graft was expanded), and a large amount of mural thrombus was observed. Further expansion continued thereafter, reaching 32.4mm at 5 years post-surgery. Thrombus was prominently observed distal to the Propaten bifurcation.Complaint # (b)(4).
Patient problems: Insufficient Information; Thrombosis/Thrombus
MAUDE report 50911465 ↗InjuryJan 4, 2022Nonstandard Device; Nonstandard Device
IT WAS REPORTED : "PROBLEM: BLOOD LEAKS FROM THE CROTCH OF THE Y TUBE. I WAS CLAMPING BECAUSE I WAS ANASTOMOSING THE CENTRAL SIDE, BUT BLOOD WAS FLOWING FROM THE CROTCH PART OF THE LEG. AN ADDITIONAL ANASTOMOSIS IS PERFORMED ON THE ARTIFICIAL BLOOD VESSEL, LEADING TO HEMOSTASIS. IN ANOTHER CASE, THE SAME THING HAPPENED AROUND (B)(6) 2021 ((B)(6), SO THIS TIME, WE WILL CONTACT YOU ABOUT THE PROBLEM. REQUEST FOR FUTURE RESPONSE: THE ACTUAL PRODUCT IS USED AND IS IN THE PATIENT'S BODY. WE CANNOT ANALYZE THE PRODUCT, BUT WE REQUEST A LETTER TO THE CUSTOMER AFTER CONFIRMING WHETHER A SIMILAR EVENT HAS OCCURRED OR NOT AT THE SAME LOT."
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 13152177 ↗InjuryDec 23, 2021Nonstandard Device; Nonstandard Device
COMPLAINT#: (B)(4). SEE MFG REPORT(S) 1640201-2021-00041.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 13076690 ↗MalfunctionApr 12, 2019Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
THE HOSPITAL REPORTED THAT DURING PREPARATION FOR A PROCEDURE, UPON OPENING THE GRAFT, IT APPEARED TO BE A BROWN SPOT ON THE GRAFT. THE GRAFT WAS NEVER IMPLANTED. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO PHYSICIAN ASSOCIATED. NO PATIENT INVOLVEMENT. THE DATE OF THE EVENT IS UNKNOWN.
Patient problems: No Patient Involvement
MAUDE report 8510183 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K954848. One submission often covers many models, so check the product and lot details in each record.
- Z-0729-2022 · Jan 14, 2022 · TerminatedHEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm; L: 60 cm Indicated for use in the replacement or repair of arteries affected with aneurysmal or occlusive
- Z-0403-2022 · Dec 1, 2021 · TerminatedHemashield Gold Knitted Bifurcated, 18 mm x 9 mm; reference number M002020851890
- Z-1861-2020 · Apr 7, 2020 · TerminatedHemashield Gold Knitted Microvel Double Velour Vascular Graft Axillo-bifemoral, Product Code: M002020955890 - Product Usage: for use in the replacement or repair of arteries affect
- Z-1860-2020 · Apr 7, 2020 · TerminatedHemashield Gold Knitted Microvel Double Velour Vascular Graft-Axillo-bifemoral, Product Code: M002020955880 - Product Usage: for use in the replacement or repair of arteries affect
- Z-1859-2020 · Apr 7, 2020 · TerminatedHemashield Gold Knitted Microvel Double Velour Vascular Graft-Axillo-bifemoral, Product Code: M002020955010 - Product Usage: For use in the replacement or repair of arteries affect
- Z-2066-2010 · May 24, 2010 · TerminatedHemashield Gold Woven and Knitted Double Vascular Grafts.
Other Graft, Vascular, Synthetic/Biologic Composite devices
- Hemashield M00202166168P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite1 recall3 reports
- Hemashield M002020851890Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite1 recall2 reports
- Hemashield M002020952070Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite1 recall0 reports
- Hemashield M00202175928P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite10 reports
- Hemashield M00202175208P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite8 reports
- Hemashield M00202175210P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite8 reports
- AlboGraft Polyester Vascular Graft AMC1407Lemaitre Vascular, Inc. · Graft, Vascular, Synthetic/Biologic Composite6 reports
- Hemashield M00202175430P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite6 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.