Skip to content

STATSTRIP 1.0 (REF 47488)Recalls & reported problems

Nova Biomedical Corporation · StatStrip LAC Meter with Battery and Quick Reference Guide · Class 1

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00385480474887), received between Sep 29, 2021 and Sep 29, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Imprecise or Erratic Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Imprecise or Erratic Results1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2021Imprecise or Erratic Results; Imprecise or Erratic Results

    ON AUGUST 30, 2021, NOVA BIOMEDICAL CANADA WAS MADE AWARE OF A POTENTIAL ISSUE REGARDING IMPRECISE LACTATE RESULTS ON A FETAL PATIENT WHILE USING A STAT STRIP LACTATE METER, SERIAL NUMBER (B)(6), AND LACTATE TEST STRIPS, LOT 2620345249. THE USER QUESTIONED THE RESULTS DURING OBSTETRIC SURGERY AND FOUND MORE BIAS IN THE METER RESULTS COMPARED TO THOSE FROM A BLOOD GAS ANALYZER. IT WAS LATER COMMUNICATED A CESAREAN SECTION (C-SECTION) WAS PERFORMED BECAUSE OF THE UNCERTAINTY OF THE RESULTS FROM THE METER IN AN EFFORT TO MITIGATE POTENTIAL HARM TO FETUS OR THE MOTHER. THE USER STATED THE C-SECTION MAY HAVE BEEN UNNECESARY BUT PROCEEDED BASED OFF THE RESULTS FROM THE METER. NO ADDITIONAL PATIENT INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 12547983 ↗

Other ACID, LACTIC, ENZYMATIC METHOD devices

All ACID, LACTIC, ENZYMATIC METHOD devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.