Skip to content

StatSensor/StatStrip 1.75 (REF 53402)Recalls & reported problems

Nova Biomedical Corporation · StatStrip StatSensor Docking Station · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00385480534024), received between Dec 22, 2020 and Dec 22, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Melted” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Melted1
  • Overheating of Device1
  • Smoking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 22, 2020Melted; Overheating of Device; Smoking; Overheating of Device

    THE CUSTOMER REPORTED THAT THE STATSTRIP WIRELESS METER (1.75) S/N: (B)(4) HAD BEEN SITTING IN THE DOCKING STATION S/N: (B)(4) FOR APPROXIMATELY ONE WEEK. WHEN THE METER WAS REMOVED FROM THE DOCKING STATION, THERE WAS A SMALL PUFF OF BLACK SMOKE, THE PLASTIC CASE WAS SLIGHTLY MELTED AND THE CONNECTION PINS APPEARED RED. THE BATTERY WAS REMOVED AND THE DOCKING STATION UNPLUGGED. THERE WAS NO PATIENT HARM OR MEDICAL INTERVENTION REPORTED. THIS IS REPORT 1 OF 2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11059362 ↗

FDA clearance and classification

The FDA regulates product code CGA (glucose oxidase, glucose) under 21 CFR 862.1345 as a Class II device. 21 CFR 862.1345 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the StatSensor/StatStrip 1.75

Has the StatSensor/StatStrip 1.75 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00385480534024), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the StatSensor/StatStrip 1.75?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Dec 22, 2020. That report lists device problems “Melted” (1), “Overheating of Device” (1) and “Smoking” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the StatSensor/StatStrip 1.75?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the StatSensor/StatStrip 1.75, and is it still sold?

The FDA's device database lists Nova Biomedical Corporation as the company responsible for it (brand StatSensor/StatStrip, model 1.75). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other GLUCOSE OXIDASE, GLUCOSE devices

All GLUCOSE OXIDASE, GLUCOSE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page