The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00389701007243), received between Mar 26, 2018 and Jan 3, 2020 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Difficult to Advance” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Difficult to Advance1
- Difficult to Remove1
- Kinked1
- Material Deformation1
Patient problems
- No Consequences Or Impact To Patient2
- Blood Loss1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 3, 2020Difficult to Remove; Difficult to Advance; Material Deformation; Difficult to Remove; Difficult to Advance; Material Deformation
THE USER FACILITY REPORTED THAT AFTER ACCESSING THE FEMORAL VEIN, THE DOCTOR TRIED TO REPOSITION THE WIRE, HOWEVER IT WOULD NEITHER ADVANCE NOR PULL BACK. INSPECTION OF WIRE SHOWED THAT THE WIRE COIL SURROUNDING IT HAD BROKEN AND THE WIRE PULLED APART. THE DOCTOR WAS PRESENT TO SAFELY REMOVE THE WIRE. THE WIRE DID NOT BREAK APART; HOWEVER, IT WAS STRETCHED. THE WIRE WAS INTACT. ADDITIONAL INFORMATION WAS RECEIVED ON 04DEC2019. THE TYPE OF PROCEDURE THAT WAS BEING PERFORMED WAS AN EP STUDY- FLUTTER ABLATION. THE PATIENT CONDITION WAS STABLE. THE PROCEDURE WAS ABLE TO BE COMPLETED. A SURGICAL INTERVENTION WAS NOT REQUIRED. HOWEVER, THE VASCULAR SURGEON WAS REQUESTED TO COME IN TO ATTEMPT TO SNARE THE WIRE, HOWEVER, WAS UNSUCCESSFUL. HE THEN OPTED TO REMOVE THE INTACT WIRE MANUALLY WITH SOMEONE READY TO APPLY PRESSURE IF NECESSARY. NO CALCIUM OR TORTUOSITY WAS NOTED IN THE VENOUS STICK. AN 18G THIN WALL COOK NEEDLE WAS USED. THE WIRE WAS NEVER BROKEN IN PIECES, HOWEVER, THE RISK OF THAT HAPPENING WAS THE CONCERN AND WHY THEY HAD THE VASCULAR SURGEON REMOVE IT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9546517 ↗MalfunctionMar 26, 2018Kinked; Kinked
THE USER FACILITY REPORTED THAT THE 5 FR. PINNACLE SHEATH KINKED IN AN ARTERY DURING A LEFT HEART CATH. IT WAS A SINGLE STICK. THE ENTRY ANGLE WAS STEEPER THAN 45%, WHICH PROBABLY CONTRIBUTED TO THE KINK. THE PHYSICIAN HAD DIFFICULTY ADVANCING AND TORQUEING THE CATHETERS. THE SHEATH WAS SWITCHED OUT FOR AN UNKINKED SHEATH AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO INJURY TO PATIENT. ADDITIONAL INFORMATION WAS RECEIVED ON 3/7/18: IT WAS REPORTED THAT THERE WAS A MINIMAL BLOOD LOSS WHEN GETTING ACCESS.
Patient problems: Blood Loss; No Consequences Or Impact To Patient
MAUDE report 7369827 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page