DESTINATION 54-86501Recalls & reported problems
Terumo Medical Corporation · DESTINATION Peripheral Guiding Sheath · Class 2
All models of DESTINATION (21)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00389701009438), received between Mar 8, 2021 and Apr 29, 2021 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Deformation4
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 29, 2021Material Deformation; Material Deformation
THE USER FACILITY REPORTED THAT WHEN THE DESTINATION WAS EXTRACTED FROM THE PACKAGE IT APPEARED CRUSHED. ADDITIONAL INFORMATION WAS RECEIVED ON 15APRIL2021: THEY RESOLVED THE ISSUE WITH THE DEVICE, TO CONTINUE THE PROCEDURE BY USING A SECOND DEVICE. THE PACKAGING WAS NOT DAMAGED. THE PATIENT'S HEALTH WAS NOT AFFECTED, THE DEVICE DID NOT GET TO TOUCH THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11749040 ↗MalfunctionApr 29, 2021Material Deformation; Material Deformation
THE USER FACILITY REPORTED THAT THE DESTINATION DESILET FOLDED, WHICH MADE IT IMPOSSIBLE TO MOUNT THE DILATOR. THIS HAPPENED WITH TWO DEVICES FROM THIS LOT. A TOTAL OF THREE DEVICES WERE USED FOR THIS PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11745199 ↗MalfunctionApr 29, 2021Material Deformation; Material Deformation
THE USER FACILITY REPORTED THAT THE DESTINATION DESILET FOLDED, WHICH MADE IT IMPOSSIBLE TO MOUNT THE DILATOR. THIS HAPPENED WITH TWO DEVICES FROM THIS LOT. A TOTAL OF THREE DEVICES WERE USED FOR THIS PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11745197 ↗MalfunctionMar 8, 2021Material Deformation; Material Deformation
THE USER FACILITY REPORTED THAT THERE WAS DIFFICULTY PLACING THE DESTINATION DILATOR INTO THE SHEATH. UPON INSPECTION IT WAS NOTED THAT PART OF THE SHEATH MIDWAY WAS CRUSHED. SURGEONS PUT THE SHEATH ASIDE FOR RETURN. THERE WERE NO ISSUES WITH THE REPLACEMENT SHEATH. THE VASCULAR SURGERY PROCEDURE WAS N ATHERECTOMY AND STENT OF THE SUPERFICIAL FEMORAL ARTERY (SFA). ADDITIONAL INFORMATION WAS RECEIVED ON 15FEB2021. THE CUSTOMER SWAPPED OUT TO A NEW TERUMO DESTINATION SHEATH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11440464 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page