GLIDESHEATH SLENDER 80-1065Recalls & reported problems

Terumo Medical Corporation · GLIDESHEATH SLENDER Stainless Steel Kit · Class 2

All models of GLIDESHEATH SLENDER (13)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00389701012070), received between Jul 10, 2019 and Mar 30, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash3
  • Difficult to Advance1
  • Material Deformation1

Patient problems

  • Blood Loss1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Information1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 30, 2024Leak/Splash; Leak/Splash

    THE USER FACILITY REPORTED THAT THE INVOLVED GLIDESHEATH SLENDER DEVICE WAS SUCCESSFULLY INTRODUCED INTO THE PATIENT'S RADIAL ARTERY BUT LEAKED BLOOD THROUGHOUT THE CASE FROM ONE (1) WAY VALVE. THE ESTIMATED BLOOD LOSS LESS THAN 250 CC. THE EVENT OCCURRED INTRA-OPERATIVE. THERE IS NOT A DIRECT ALLEGATION THAT THE REPORTED DEVICE CAUSED OR CONTRIBUTED TO PATIENT INJURY AND/OR NEED FOR MEDICAL INTERVENTION. THE PATIENT WAS NOT INJURED DURING THE EVENT AND MEDICAL/SURGICAL INTERVENTION WAS NOT NEEDED. ADDITIONAL INFORMATION WAS RECEIVED ON 27 FEB 2024: PROCEDURE WAS A LEFT HEART CATHETERIZATION. THE BLOOD LEAKAGE WAS CONFIRMED TO BEING LEAKING FROM THE CC VALUE. THE ACCOUNT CONTINUED THE CASE AS IT WAS A NORMAL CATH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19009045 ↗
  • MalfunctionFeb 28, 2020Leak/Splash; Leak/Splash

    THE USER FACILITY REPORTED THAT THE GLIDESHEATH SLENDER WAS INSERTED INTO RIGHT RADIAL ARTERY WITHOUT INCIDENT. A 5FR TIG CATHETER WAS INSERTED THROUGH THE SHEATH OVER A .035 J WIRE TO THE DESCENDING AORTA WITHOUT INCIDENT. THE WIRE WAS REMOVED AND SUBSEQUENT IMAGING OF BOTH RIGHT AND LEFT CORONARY ARTERIES WAS ACCOMPLISHED AGAIN WITHOUT INCIDENT. AFTER WHICH IT WAS NOTICED THAT THE SHEATH WAS LEAKING AT THE JUNCTION OF THE HUB AND THE SHEATH TUBING MATERIAL. THE LEAKING DID NOT INVOLVE THE VALVE. THE BLOOD LOSS WAS LESS THAN 10CC. THE SHEATH WAS REMOVED, AND A TR BAND WAS PLACED OVER THE PUNCTURE SITE WITHOUT INCIDENT. PATIENT CONDITION WAS REPORTED TO BE EXCELLENT, NO ISSUE. THE PROCEDURE OUTCOME WAS EXCELLENT. THERE WAS NO PATIENT INJURY/MEDICAL OR SURGICAL INTERVENTION REQUIRED. ADDITIONAL INFORMATION WAS RECEIVED ON 06FEB2020. THE LOCATION WAS ON THE SHEATH TUBING JUST BELOW THE VALVE/HUB AREA. THE PATIENT WAS IN STABLE CONDITION. THE PROCEDURE WAS A LEFT HEART CATHETERIZATION VIA A RIGHT RADIAL APPROACH.

    Patient problems: Blood Loss

    MAUDE report 9771318 ↗
  • MalfunctionSep 4, 2019Difficult to Advance; Material Deformation; Difficult to Advance; Material Deformation

    THE USER FACILITY REPORTED AN ISSUE WITH THE GLIDESHEATH SLENDER, THEY HAD TROUBLE GETTING THE SHEATH WIRE TO ADVANCE THROUGH THE SHEATH ITSELF. IT APPEARED TO HAVE A "BULB SHAPE" AT THE SOFT TIP. THEY TRIED THE DISTAL END OF THE WIRE AND IT ADVANCED WITH NO ISSUE. THE PATIENT WAS IN STABLE CONDITION. THEY WERE ABLE TO USE THE REPORTED DEVICE AND NO OTHER EQUIPMENT OR DEVICES WERE USED WITH THE REPORTED PRODUCT. THE PROCEDURE OUTCOME WAS A SUCCESS. ADDITIONAL INFORMATION WAS RECEIVED 19AUGUST2019: THE PROCEDURE WAS A DIAGNOSTIC LEFT HEART CATH.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8960003 ↗
  • MalfunctionJul 10, 2019Leak/Splash; Leak/Splash

    THE USER FACILITY REPORTED THAT THE INVOLVED GLIDE SHEATH SLENDER SHEATH WAS DAMAGED AND THE VALVE WAS LEAKING THROUGHOUT THE PROCEDURE.

    Patient problems: No Information

    MAUDE report 8778211 ↗