Trilogy Trilogy200 Ventilator, GermanyRecalls & reported problems

Respironics, Inc. · Trilogy 200 Ventilator, Germany · Class 2

All models of Philips Trilogy Ventilator (90) · Trilogy (46)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00606959022690), received between Mar 9, 2016 and Feb 23, 2021 — classified as 15 malfunctions.

The most frequently reported specific device problem is “Device Operational Issue” (3 reports).

5 other recalls cover devices cleared under the same FDA submission (K093416); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available3
  • Device Operational Issue3
  • Failure to Charge2
  • Incorrect Or Inadequate Test Results2
  • Adverse Event Without Identified Device or Use Problem1
  • Computer Software Problem1
  • Device Displays Incorrect Message1
  • Failure to Recalibrate1

Patient problems

  • No Consequences Or Impact To Patient11
  • No Clinical Signs, Symptoms or Conditions2
  • No Patient Involvement2

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 23, 2021Mechanical Problem; Mechanical Problem

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR ROUTINE PREVENTATIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE'S KEYPAD WAS FOUND TO BE STUCK. THE DEVICE'S KEYPAD WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11364251 ↗
  • MalfunctionFeb 8, 2021Failure to Recalibrate; Failure to Recalibrate

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE FAILED A STEP DURING TESTING. THE DEVICE'S ACTIVE EXHALATION CONTROL MODULE WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11294046 ↗
  • MalfunctionApr 9, 2019Computer Software Problem; Unexpected Shutdown; Computer Software Problem; Unexpected Shutdown

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED. THE MANUFACTURER'S INVESTIGATION IS CURRENTLY ON-GOING AND A FOLLOW-UP REPORT WILL BE FILED WHEN THE INVESTIGATION IS COMPLETE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8495247 ↗
  • MalfunctionDec 13, 2017Tidal Volume Fluctuations; Tidal Volume Fluctuations

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR WAS NOT ABLE TO DELIVER THE SET VOLUME. THERE WAS NO HARM OR INJURY REPORTED. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVIC'S FILTER WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7112291 ↗
  • MalfunctionAug 9, 2017Failure to Charge; Failure to Charge

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR FAILED TO CHARGE ITS INTERNAL BATTERY. THERE WAS NO HARM OR INJURY REPORTED. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S INTERNAL BATTERY WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6776667 ↗
  • MalfunctionDec 21, 2016Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR ROUTINE PREVENTIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SENSOR BOARD WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6193036 ↗
  • MalfunctionOct 12, 2016Incorrect Or Inadequate Test Results; Incorrect Or Inadequate Test Results

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE FAILED A STEP DURING TESTING. THE DEVICE'S SENSOR BOARD WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6021607 ↗
  • MalfunctionOct 5, 2016Device Operational Issue; Device Operational Issue

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S INTERNAL BATTERY WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5999877 ↗
  • MalfunctionSep 14, 2016Device Operational Issue; Device Operational Issue

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S DETACHABLE BATTERY CABLE WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5950112 ↗
  • MalfunctionSep 1, 2016Device Operational Issue; Device Operational Issue

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S DETACHABLE BATTERY CABLE WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5922330 ↗
  • MalfunctionAug 10, 2016Power Problem; Power Problem

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR'S POWER CONNECTOR WAS FAULTY. THERE WAS NO HARM OR INJURY REPORTED. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S POWER CORD WAS REPLACED TO ADDRESS THE ISSUE. THE DEVICE HAD PHYSICAL DAMAGE CONSISTENT WITH BEING DROPPED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5865670 ↗
  • MalfunctionJul 20, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR ROUTINE PREVENTATIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SENSOR BOARD WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Patient Involvement

    MAUDE report 5808553 ↗
  • MalfunctionApr 27, 2016Failure to Charge; Device Displays Incorrect Message; Failure to Charge; Device Displays Incorrect Message

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A "SERVICE REQUIRED" ALARM CONDITION OCCURRED AND A FAILURE TO CHARGE THE INTERNAL BATTERY. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S DETACHABLE BATTERY CONNECTOR ASSEMBLY WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5611336 ↗
  • MalfunctionApr 13, 2016Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER'S SERVICE CENTER FOR EVALUATION. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S BATTERY CONNECTOR CABLE WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5574229 ↗
  • MalfunctionMar 9, 2016Incorrect Or Inadequate Test Results; Appropriate Device Problem Term/Code Not Available; Incorrect Or Inadequate Test Results; Appropriate Device Problem Term/Code Not Available

    A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR ROUTINE PREVENTIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE FAILED A STEP DURING TESTING. THE DEVICE'S ACTIVE EXHALATION CONTROL MODULE WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: No Patient Involvement

    MAUDE report 5489281 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K093416. One submission often covers many models, so check the product and lot details in each record.

  • Z-0883-2023 · Dec 7, 2022 · Open, Classified
    Philips Trilogy 200 Ventilator, Model Numbers 1040005, U1040005, 1040005B, U1040005B, CA1032800B, CA1032800, KO1032800, 1032800, R1032804TP, R1032804B, R1032804TPV, 1032804B
  • Z-1956-2021 · Jun 14, 2021 · Open, Classified
    Trilogy 100, Trilogy 200, Garbin Plus, Aeris, LifeVent Continuous Ventilators.
  • Z-1065-2016 · Jan 18, 2016 · Terminated
    Trilogy 100, Trilogy 200, Trilogy O2, Trilogy 202, Trilogy EC, Garbin, Garbin Plus Trilogy 100, Trilogy 200, Garbin, Garbin plus provides continuous or intermittent ventilatory su
  • Z-1143-2014 · Feb 11, 2014 · Terminated
    Trilogy Ventilators The Respironics Trilogy 100 system provides continuous or intermittent ventilator support for the care of individuals who require mechanical ventilation. Trilog
  • Z-2024-2012 · Apr 27, 2012 · Terminated
    Respironics Trilogy 100, 200, and 202 Ventilators The Respironics Trilogy systems provides continuous or intermittent ventilatory support for the care of individuals who require m