Philips Trilogy VentilatorRecalls & reported problems across all models

Respironics, Inc. · 90 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

22 open recalls name at least one Philips Trilogy Ventilator model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
36,113
Deaths reported
316
Injuries reported
641
Recalls
22

We found 90 models of Philips Trilogy Ventilator by Respironics, Inc. in the FDA's device registry (GUDID). Together they are named in 36,113 adverse event reports in the FDA's MAUDE database, received between Sep 23, 2015 and Sep 30, 2026 — classified as 316 deaths, 641 injuries, 35,140 malfunctions.

The most frequently reported specific device problem is “Circuit Failure” (5,246 reports).

22 FDA recall records name at least one of these models by its device identifier, 22 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 332 reportsDec 2023: 382 reportsJan 2024: 654 reportsFeb 2024: 494 reportsMar 2024: 379 reportsApr 2024: 349 reportsMay 2024: 370 reportsJun 2024: 421 reportsJul 2024: 522 reportsAug 2024: 385 reportsSep 2024: 573 reportsOct 2024: 910 reportsNov 2024: 759 reportsDec 2024: 356 reportsJan 2025: 643 reportsFeb 2025: 482 reportsMar 2025: 659 reportsApr 2025: 399 reportsMay 2025: 630 reportsJun 2025: 337 reportsJul 2025: 393 reportsAug 2025: 241 reportsSep 2025: 280 reportsOct 2025: 271 reportsNov 2025: 265 reportsDec 2025: 418 reportsJan 2026: 303 reportsFeb 2026: 318 reportsMar 2026: 663 reportsApr 2026: 1,073 reportsMay 2026: 532 reportsJun 2026: 332 reportsJul 2026: 14 reportsAug 2026: 19 reportsSep 2026: 165 reportsOct 2026: 0 reports2024202520261,073
Last 36 months. The most recent month may be incomplete.

Philips Trilogy Ventilator recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Oct 7, 2024Open, Classified

    Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.

    Z-0293-2025 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  2. Oct 7, 2024Open, Classified

    Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.

    Z-0292-2025 · Root cause: Device Design · Recall details · FDA record ↗

  3. Oct 7, 2024Open, Classified

    Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.

    Z-0291-2025 · Root cause: Device Design · names 18 of these models · Recall details · FDA record ↗

  4. Oct 7, 2024Open, Classified

    Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.

    Z-0290-2025 · Root cause: Device Design · names 11 of these models · Recall details · FDA record ↗

  5. Jul 16, 2024Open, Classified

    Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls.

    Z-2944-2024 · Root cause: Device Design · names 28 of these models · Recall details · FDA record ↗

  6. Jul 16, 2024Open, Classified

    Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls.

    Z-2943-2024 · Root cause: Device Design · Recall details · FDA record ↗

  7. Jul 16, 2024Open, Classified

    Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls.

    Z-2942-2024 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  8. Mar 19, 2024Open, Classified

    Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.

    Z-1505-2024 · Root cause: Device Design · Recall details · FDA record ↗

  9. Mar 19, 2024Open, Classified

    Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.

    Z-1504-2024 · Root cause: Device Design · names 17 of these models · Recall details · FDA record ↗

  10. Mar 19, 2024Open, Classified

    Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.

    Z-1503-2024 · Root cause: Device Design · names 6 of these models · Recall details · FDA record ↗

  11. Mar 19, 2024Open, Classified

    Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.

    Z-1502-2024 · Root cause: Device Design · names 7 of these models · Recall details · FDA record ↗

  12. Nov 27, 2023Open, Classified

    The following was missing from the Contraindications Statement: The AVAPS-AE therapy mode is contraindicated for invasive use and patients less than 10 kg. The AVAPS feature is contraindicated for patients less than 10 kg. Potential harms if devices are used on contraindicated patients include barotrauma, hypoventilation/hypercapnia, and rebreathing of excessive CO2.

    Z-0638-2024 · Root cause: Device Design · names 24 of these models · Recall details · FDA record ↗

  13. May 25, 2023Open, Classified

    There are errors present in the Korean Instructions For Use where the word "invasive," was incorrectly translated to "noninvasive".

    Z-2106-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  14. May 1, 2023Open, Classified

    The Trilogy Evo Universal ventilator provides continuous or intermittent positive pressure ventilation for the care of individuals who require mechanical ventilation. Trilogy Evo Universal is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and pulse rate data when integrated with the appropriate accessories. The ventilator is suitable for use in institutional and hospital settings and non-emergency transport settings; for example, wheelchair. It may be used for both invasive and non-invasive ventilation.

    Z-1652-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  15. May 1, 2023Open, Classified

    Environmental debris may accumulate on the internal machine flow sensor causing partial occlusion which may impact accurate delivery of pressure, volume, or flow. A malfunction caused by this issue can result in patient harm up to hypoxemia if not addressed by the care provider.

    Z-1651-2023 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

  16. May 1, 2023Open, Classified

    Environmental debris may accumulate on the internal machine flow sensor causing partial occlusion which may impact accurate delivery of pressure, volume, or flow. A malfunction caused by this issue can result in patient harm up to hypoxemia if not addressed by the care provider.

    Z-1650-2023 · Root cause: Under Investigation by firm · names 5 of these models · Recall details · FDA record ↗

  17. May 1, 2023Open, Classified

    Environmental debris may accumulate on the internal machine flow sensor causing partial occlusion which may impact accurate delivery of pressure, volume, or flow. A malfunction caused by this issue can result in patient harm up to hypoxemia if not addressed by the care provider.

    Z-1649-2023 · Root cause: Under Investigation by firm · names 7 of these models · Recall details · FDA record ↗

  18. Mar 1, 2023Open, Classified

    The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually delivering. This may lead to under delivery of oxygen.

    Z-1289-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  19. Mar 1, 2023Open, Classified

    The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually delivering. This may lead to under delivery of oxygen.

    Z-1288-2023 · Root cause: Under Investigation by firm · names 3 of these models · Recall details · FDA record ↗

  20. Mar 1, 2023Open, Classified

    The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually delivering. This may lead to under delivery of oxygen.

    Z-1287-2023 · Root cause: Under Investigation by firm · names 10 of these models · Recall details · FDA record ↗

  21. Dec 7, 2022Open, Classified

    The replacement silicone sound abatement foam installed into some Trilogy 100 and Trilogy 200 devices may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially block air inlet, which could result in a reduction in delivered therapy volume or pressure and could cause the device to alarm.

    Z-0882-2023 · Root cause: Process design · Recall details · FDA record ↗

  22. Dec 22, 2021Open, Classified

    Philips has identified a specific lot of non-conforming material manufactured by one of its suppliers. A Philips supplier incorrectly used polyester-based polyurethane (PE-PUR) sound abatement foam, a non-conforming material in the devices and repair kits. PE-PUR foam may degrade into particles which may enter the device s air pathway and be ingested or inhaled by the user. The foam degradation may be exacerbated by use of unapproved cleaning methods such as unapproved ozone cleaners, and the PE-PUR foam may off-gas certain chemicals during initial use of the device.

    Z-0493-2022 · Root cause: Mixed-up of materials/components · names 2 of these models · Recall details · FDA record ↗

Most reported Philips Trilogy Ventilator problems

Device problems

  • Circuit Failure5,246
  • Electrical /Electronic Property Problem4,737
  • Battery Problem4,649
  • Failure to Recalibrate3,991
  • Mechanical Problem3,980
  • Failure to Charge2,783
  • Failure to Calibrate2,419
  • Device Displays Incorrect Message2,120

Patient problems

  • No Clinical Signs, Symptoms or Conditions25,054
  • No Patient Involvement4,989
  • No Consequences Or Impact To Patient4,877
  • Unspecified Respiratory Problem354
  • No Known Impact Or Consequence To Patient309
  • Headache225
  • Dizziness203
  • Asthma147

Philips Trilogy Ventilator models

90 models, most reported first. Each model has its own page with its reports, recalls and device details.