Trilogy 00 (REF RSL1054260)Recalls & reported problems
Respironics, Inc. · Trilogy100 Ventilator, Rental, U.S. version · Class 2
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00606959026322), received between Sep 7, 2016 and Sep 25, 2026 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Battery Problem” (6 reports).
8 other recalls cover devices cleared under the same FDA submission (K083526); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Battery Problem6
- Electrical /Electronic Property Problem3
- Device Operational Issue2
- Mechanical Problem2
- Contamination1
- Degraded1
- Failure to Charge1
- Failure to Power Up1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- No Patient Involvement3
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 25, 2026Failure to Power Up
A Trilogy 100 was returned to a third-party service center for evaluation. There was no harm or injury reported. There was no evidence that the device was in patient use. During the evaluation the device was noted to have no power. There was no documentation of any components replaced to address this issue. Additionally, there were no visible foam particles noted. No repairs were performed. The device was scrapped.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50900865 ↗MalfunctionJan 19, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR FOAM REMEDIATION. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE HAS BEEN RETURNED TO THE MANUFACTURER, BUT REPAIR HAS NOT YET BEGUN. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18542421 ↗MalfunctionDec 15, 2023Battery Problem; Battery Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING THE DEVICE DID NOT MEET THE ACCEPTANCE CRITERIA OF THE EVALUATION PROCESS FOR THE FOLLOWING CONDITIONS: CRITICAL ERRORS (193, 290). ERROR 193 INTERNAL LI-ION BATTERY PERMANENT FAILURE. THE INTERNAL BATTERY IS THE FINAL POWER SOURCE. FAILURE OF THE INTERNAL BATTERY MAY IMPACT THERAPY. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18334403 ↗MalfunctionDec 11, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
DURING THE EVALUATION OF THE TRILOGY100 AT THE MANUFACTURER'S SERVICE CENTER, IT WAS CONFIRMED THAT THE DEVICE DID NOT MEET CRITERIA OF EVALUATION PROCESS FOR "CAN'T SEE ANYTHING ON THE SCREEN." THERE WAS NO HARM OR INJURY REPORTED. THE DEVICE WAS SCRAPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18300588 ↗MalfunctionMay 5, 2023Battery Problem; Battery Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR WAS ALARMING FOR LOW BATTERY. THE DEVICE WAS NOT IN PATIENT USE. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16877380 ↗MalfunctionAug 26, 2022Battery Problem; Battery Problem
A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S INTERNAL BATTERY NEEDS REPLACED TO ADDRESS THE ISSUE. THE DEVICE WAS RETURNED UNREPAIRED PER THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15300587 ↗MalfunctionMar 1, 2022Degraded; Degraded
A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR ROUTINE PREVENTATIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE THIRD PARTY SERVICE CENTER, THE DEVICE'S INLET AIR PATH ASSEMBLY AND AIR PATH FOAM WAS REPLACED DUE TO CONTAMINATION. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13633850 ↗MalfunctionAug 3, 2021Battery Problem; Battery Problem
A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR ROUTINE PREVENTIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S INTERNAL BATTERY NEEDS REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12268221 ↗MalfunctionApr 14, 2021Mechanical Problem; Mechanical Problem
A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S DETACHABLE BATTERY CABLE WAS REPLACED TO ADDRESS THE ISSUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11665476 ↗MalfunctionFeb 23, 2021Contamination; Mechanical Problem; Contamination; Mechanical Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE CONDITION OCCURRED. THE DEVICE WAS NOT IN PATIENT USE. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S BLOWER MOTOR WAS REPLACED TO ADDRESS THE ISSUE. THE BLOWER HAD EVIDENCE OF CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11364238 ↗MalfunctionFeb 19, 2019Failure to Charge; Battery Problem; Failure to Charge; Battery Problem
A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, AN ERROR CODE RELATED TO THE CHARGING OF THE INTERNAL BATTERY WAS OBSERVED. THE DEVICE'S INTERNAL BATTERY WAS REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8351066 ↗MalfunctionMar 14, 2018Battery Problem; Battery Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A FAILURE OF A VENTILATOR'S INTERNAL BATTERY. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, AN ERROR CODE RELATED TO THE CHARGING OF THE INTERNAL BATTERY WAS OBSERVED. THE DEVICE'S INTERNAL BATTERY NEEDS TO BE REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Patient Involvement
MAUDE report 7337396 ↗MalfunctionOct 25, 2017Device Operational Issue; Device Operational Issue
A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR ROUTINE PREVENTIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, AN ISSUE RELATED TO THE ACTIVE EXHALATION CONTROL MODULE WAS OBSERVED. THE DEVICE'S ACTIVE EXHALATION CONTROL MODULE WAS REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Patient Involvement
MAUDE report 6976073 ↗MalfunctionOct 26, 2016Device Operational Issue; Device Operational Issue
A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE AND "VENTILATOR INOPERATIVE" CODE WERE FOUND IN THE DEVICE'S DOWNLOADED ERROR LOG. THE DEVICE'S BLOWER MOTOR AND SENSOR BOARD WERE REPLACED TO ADDRESS THE ISSUES.
Patient problems: No Patient Involvement
MAUDE report 6057013 ↗MalfunctionSep 7, 2016Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR'S AC INLET CONNECTOR WAS DAMAGED. THERE WAS NO HARM OR INJURY REPORTED. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S AC INLET CONNECTOR WAS REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5932495 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K083526. One submission often covers many models, so check the product and lot details in each record.
- Z-0882-2023 · Dec 7, 2022 · Open, ClassifiedPhilips Trilogy 100 Ventilator, Model Numbers 1054260, 1054096, U1054260, U1054260B, 1054260B, CA1054096B, BT1054260, 1054096B, R1054260, RSL1054260, CA1054096, R1054260B, R1054655
- Z-0884-2023 · Dec 7, 2022 · Open, ClassifiedPhilips Garbin Ventilator, Model Number 1058180B
- Z-1956-2021 · Jun 14, 2021 · Open, ClassifiedTrilogy 100, Trilogy 200, Garbin Plus, Aeris, LifeVent Continuous Ventilators.
- Z-1065-2016 · Jan 18, 2016 · TerminatedTrilogy 100, Trilogy 200, Trilogy O2, Trilogy 202, Trilogy EC, Garbin, Garbin Plus Trilogy 100, Trilogy 200, Garbin, Garbin plus provides continuous or intermittent ventilatory su
- Z-1143-2014 · Feb 11, 2014 · TerminatedTrilogy Ventilators The Respironics Trilogy 100 system provides continuous or intermittent ventilator support for the care of individuals who require mechanical ventilation. Trilog
- Z-2024-2012 · Apr 27, 2012 · TerminatedRespironics Trilogy 100, 200, and 202 Ventilators The Respironics Trilogy systems provides continuous or intermittent ventilatory support for the care of individuals who require m
- Z-0296-2012 · Oct 11, 2011 · TerminatedTrilogy 100, 200 and 202 Ventilators The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used fo
- Z-1896-2010 · Jun 10, 2010 · TerminatedTrilogy 100 Ventilator, Model/Catalog Number 1054260, U1054260 (Domestic). Philips Respironics, Monroeville, PA 15146. Provides continuous or intermittent ventilatory support for
Other Oximeter devices
- Trilogy Trilogy100 Ventilator, InternationalRespironics, Inc. · Oximeter8 deaths2,493 reports
- Trilogy Trilogy 202 Ventilator, United StatesRespironics, Inc. · Oximeter2 deaths1,594 reports
- HAMILTON-G5 159001Hamilton Medical AG · Oximeter18 deaths1,440 reports
- Trilogy Trilogy100 Ventilator, GermanyRespironics, Inc. · Oximeter696 reports
- Model EMV+ Vent, Army, MRI 799-EMVP-04Zoll Medical Corporation · Oximeter1 death605 reports
- Trilogy Evo, O2, Japan JP2100X16BRespironics, Inc. · Oximeter4 recalls5 deaths342 reports
- Model EMV+ Vent, Army, COMMERCIAL 799-EMVP-01-01Zoll Medical Corporation · Oximeter1 death497 reports
- HAMILTON-G5 159002Hamilton Medical AG · Oximeter3 deaths451 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.