Trilogy 200 Trilogy200 Ventilator, Korea (Philips)Recalls & reported problems
Respironics, Inc. · Trilogy 200 Ventilator, Korea · Class 2
All models of Philips Trilogy Ventilator (90) · Trilogy 200 (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00606959047419), received between Jun 13, 2018 and Sep 16, 2021 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Circuit Failure” (4 reports).
5 other recalls cover devices cleared under the same FDA submission (K093416); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Circuit Failure4
- Device Inoperable1
- Electrical /Electronic Property Problem1
- Mechanical Problem1
- Pressure Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Patient Involvement2
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2021Circuit Failure; Circuit Failure
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR HAD A LEAK ISSUE. THE DEVICE WAS NOT IN PATIENT USE. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S SENSOR BOARD WAS REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12480511 ↗MalfunctionJul 8, 2021Pressure Problem; Pressure Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A POSITIVE END-EXPIRATORY PRESSURE MALFUNCTION. THE DEVICE WAS NOT IN PATIENT USE. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12134789 ↗MalfunctionApr 12, 2021Circuit Failure; Circuit Failure
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR HAD A VALVE LEAK. THE DEVICE WAS NOT IN PATIENT USE. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S SENSOR BOARD WAS REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11649695 ↗MalfunctionJan 13, 2021Circuit Failure; Circuit Failure
A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SENSOR BOARD WAS REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11164032 ↗MalfunctionMar 13, 2019Circuit Failure; Circuit Failure
A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, "SERVICE REQUIRED" CODES WERE FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SYSTEM BOARD WAS REPLACED TO ADDRESS THE ISSUE. THE SYSTEM BOARD WAS RETURNED TO THE MANUFACTURER'S QUALITY ASSURANCE LABORATORY FOR FURTHER INVESTIGATION. DURING THE INVESTIGATION, THE CAUSE OF THE SYSTEM BOARD FAILURE WAS DUE TO A FAILURE OF THE CENTRAL PROCESSING UNIT CHIP (U1) ON THE DAUGHTER BOARD OF THE SYSTEM BOARD.
Patient problems: No Patient Involvement
MAUDE report 8416605 ↗MalfunctionDec 13, 2018Mechanical Problem; Mechanical Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE CONDITION OCCURRED. THE DEVICE WAS NOT IN PATIENT USE. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S BLOWER MOTOR WAS REPLACED TO ADDRESS THE ISSUE.
Patient problems: No Patient Involvement
MAUDE report 8161529 ↗MalfunctionJun 13, 2018Electrical /Electronic Property Problem; Device Inoperable; Electrical /Electronic Property Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED. THE DEVICE HAS YET TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7596887 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K093416. One submission often covers many models, so check the product and lot details in each record.
- Z-0883-2023 · Dec 7, 2022 · Open, ClassifiedPhilips Trilogy 200 Ventilator, Model Numbers 1040005, U1040005, 1040005B, U1040005B, CA1032800B, CA1032800, KO1032800, 1032800, R1032804TP, R1032804B, R1032804TPV, 1032804B
- Z-1956-2021 · Jun 14, 2021 · Open, ClassifiedTrilogy 100, Trilogy 200, Garbin Plus, Aeris, LifeVent Continuous Ventilators.
- Z-1065-2016 · Jan 18, 2016 · TerminatedTrilogy 100, Trilogy 200, Trilogy O2, Trilogy 202, Trilogy EC, Garbin, Garbin Plus Trilogy 100, Trilogy 200, Garbin, Garbin plus provides continuous or intermittent ventilatory su
- Z-1143-2014 · Feb 11, 2014 · TerminatedTrilogy Ventilators The Respironics Trilogy 100 system provides continuous or intermittent ventilator support for the care of individuals who require mechanical ventilation. Trilog
- Z-2024-2012 · Apr 27, 2012 · TerminatedRespironics Trilogy 100, 200, and 202 Ventilators The Respironics Trilogy systems provides continuous or intermittent ventilatory support for the care of individuals who require m
Other Oximeter devices
- Trilogy Trilogy100 Ventilator, InternationalRespironics, Inc. · Oximeter8 deaths2,493 reports
- Trilogy Trilogy 202 Ventilator, United StatesRespironics, Inc. · Oximeter2 deaths1,594 reports
- HAMILTON-G5 159001Hamilton Medical AG · Oximeter18 deaths1,440 reports
- Trilogy Trilogy100 Ventilator, GermanyRespironics, Inc. · Oximeter696 reports
- Model EMV+ Vent, Army, MRI 799-EMVP-04Zoll Medical Corporation · Oximeter1 death605 reports
- Trilogy Evo, O2, Japan JP2100X16BRespironics, Inc. · Oximeter4 recalls5 deaths342 reports
- Model EMV+ Vent, Army, COMMERCIAL 799-EMVP-01-01Zoll Medical Corporation · Oximeter1 death497 reports
- HAMILTON-G5 159002Hamilton Medical AG · Oximeter3 deaths451 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page