Skip to content

Sensor External Flow Sensor, Adult/PedRecalls & reported problems

Respironics, Inc. · External Flow Sensor, Adult / Pediatric · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00606959050839), received between Jul 9, 2021 and Feb 22, 2023 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Contamination” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Contamination2
  • Device Sensing Problem1
  • Mechanical Problem1
  • Tidal Volume Fluctuations1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 22, 2023Contamination; Mechanical Problem; Contamination; Mechanical Problem

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR'S EXTERNAL FLOW SENSOR FAILED. THERE WAS NO HARM OR INJURY REPORTED. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER¿S COMPLAINT WAS CONFIRMED. THE DEVICE'S FLOW SENSOR WAS REPLACED TO ADDRESS THE ISSUE. THERE WAS EVIDENCE OF CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16416068 ↗
  • MalfunctionDec 21, 2021Device Sensing Problem; Device Sensing Problem

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR WAS ALARMING FOR A FLOW SENSOR ERROR. THERE WAS NO HARM OR INJURY REPORTED. THE EXTERNAL FLOW SENSOR WAS REPLACED TO ADDRESS THE ISSUE. THE EXTERNAL FLOW SENSOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. THE EXTERNAL FLOW SENSOR WAS CONTAMINATED WITH DUST AND DIRT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13054094 ↗
  • MalfunctionDec 21, 2021Contamination; Contamination

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR WAS ALARMING FOR A FAULTY FLOW SENSOR. THERE WAS NO HARM OR INJURY REPORTED. THE DEVICE'S EXTERNAL FLOW SENSOR WAS REPLACED TO ADDRESS THE ISSUE. THE FLOW SENSOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. THE EXTERNAL FLOW SENSOR WAS FOUND TO HAVE RUST ON THE INTERNAL MESH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13053349 ↗
  • MalfunctionJul 9, 2021Tidal Volume Fluctuations; Tidal Volume Fluctuations

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A MALFUNCTION OF A VENTILATOR'S EXTERNAL FLOW SENSOR CAUSING DECREASED TIDAL VOLUMES. THERE WAS NO HARM OR INJURY REPORTED. THE COMPONENT HAS NOT YET BEEN RETURNED TO THE MANUFACTURER. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12141864 ↗

FDA clearance and classification

The FDA regulates product code NOU (continuous, ventilator, home use) under 21 CFR 868.5895 as a Class II device. 21 CFR 868.5895 ↗

Devices of this type normally need 510(k) clearance before sale.

The FDA lists this type of device as life-sustaining or life-supporting.

FDA definition of this device type: This product code was needed for the home use indication. CBK is not a tracked device and NOU is a tracked device.

Common questions about the Sensor External Flow Sensor, Adult/Ped

Has the Sensor External Flow Sensor, Adult/Ped been recalled?

No FDA recall record in our data names this exact device identifier (DI 00606959050839), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Sensor External Flow Sensor, Adult/Ped?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jul 9, 2021 and Feb 22, 2023. In those reports, the most common specific device problems are “Contamination” (2), “Device Sensing Problem” (1) and “Mechanical Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Sensor External Flow Sensor, Adult/Ped?

None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Sensor External Flow Sensor, Adult/Ped, and is it still sold?

The FDA's device database lists Respironics, Inc. as the company responsible for it (brand Sensor, model External Flow Sensor, Adult/Ped). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CONTINUOUS, VENTILATOR, HOME USE devices

All CONTINUOUS, VENTILATOR, HOME USE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page