Skip to content

Trilogy Evo and Trilogy Evo O2 recallZ-0638-2024 · Philips Respironics, Inc. · initiated Nov 27, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Respironics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The following was missing from the Contraindications Statement: The AVAPS-AE therapy mode is contraindicated for invasive use and patients less than 10 kg. The AVAPS feature is contraindicated for patients less than 10 kg. Potential harms if devices are used on contraindicated patients include barotrauma, hypoventilation/hypercapnia, and rebreathing of excessive CO2.

FDA root cause category: Device Design

What the company is doing

On November 27, 2023, the firm sent "URGENT Medical Device Correction" letters to dated November 20, 2023, affected customers along with an attached manual addendum for insertion into current manuals. Customers were instructed to keep a copy of the letter and attached addendum with their Trilogy Evo or Trilogy Evo O2 Ventilator IFUs. Distributors were instructed to notify any customers to whom they have distributed affected devices. If you need any further information or support concerning this issue, please contact your local Philips representative: Call 1-(877)-387-3311 or email at [email protected]

Recall details

Recall number
Z-0638-2024
Status
Open, Classified
Initiated by the company
Nov 27, 2023
Posted by the FDA
Jan 4, 2024
Recalling firm
Philips Respironics, Inc.
Quantity
100,126 (54,691 US; 45,435 OUS) in total
Product code
CBK
FDA submissions
K181166
Distribution
Worldwide distribution: US (nationwide)-AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Including PR; and OUS (International) including countries of: AE, AR,AT, AU, BD, BE, BG, BO, BR, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, ;EC EE, EG, ES, FI, FR, GB, GE, GF, GR, GT, HK, HR, HU, IE, IL, IN, IT, JO, JP, KE, KR, KW, LB, LT, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, OM, PA, PE, PF, PH, PK, PL, PR, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, YT, ZA and ZW.

Product description

Trilogy Evo and Trilogy Evo O2 Material Numbers BL2110X15B BR2110X18B CA2110X12B DE2100X13B DE2110X13B DS2100X11B DS2110X11B EE2100X15B EE2110X15B ES2100X15B ES2110X15B EU2100X15B EU2100X19 EU2110X15B FR2100X14B FR2110X14B FX2100X15B GB2110X15B IA2100X15B IA2110X15B IN2100X15B IN2100X19 IN2110X15B IT2100X21B IT2110X21B JP2100X16B JP2110X16B KR2110X15B LA2100X15B LA2110X15B ND2110X15B RDE2100X13B RDE2110X13B RDS2110X11B RIN2100X15B RIN2110X15B RJP2100X16B RJP2110X16B UDS2110X11B

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Material UDI-DI/GTIN BL2110X15B 606959055520 BR2110X18B 606959055476 CA2110X12B 606959052307 DE2100X13B 606959054103 DE2110X13B 606959054110 DS2100X11B 606959051997 DS2110X11B 606959051942 EE2100X15B 606959055698 EE2110X15B 606959055544 ES2100X15B 606959054097 ES2110X15B 606959054080 EU2100X15B 606959055674 EU2100X19 606959055681 EU2110X15B 606959054981 FR2100X14B 606959055643 FR2110X14B 606959055506 FX2100X15B 606959061019 GB2110X15B 606959054127 IA2100X15B 606959055636 IA2110X15B 606959055490 IN2100X15B 606959054059 IN2100X19 606959055575 IN2110X15B 606959051959 IT2100X21B 606959055438 IT2110X21B 606959055421 JP2100X16B 606959055582 JP2110X16B 606959051973 KR2110X15B 606959055483 LA2100X15B 606959055599 LA2110X15B 606959055452 ND2110X15B 606959055513 RDE2100X13B RDE2110X13B RDS2110X11B 606959060517 RIN2100X15B RIN2110X15B RJP2100X16B RJP2110X16B UDS2110X11B 606959060500 All units of listed models are affected by this correction.

Devices named in this recall

24 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page