Vasoview EVH Accsessory Pack VH-2004Recalls & reported problems
Maquet Cardiovascular LLC · Vasoview EVH Accsessory Pack · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00607567700376), received between Feb 25, 2021 and Nov 13, 2025 — classified as 3 injuries, 12 malfunctions.
The most frequently reported specific device problem is “Improper Flow or Infusion” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Improper Flow or Infusion7
- Material Rupture3
- Environmental Particulates2
- Contamination /Decontamination Problem1
- Defective Component1
- No Flow1
- Peeled/Delaminated1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- Insufficient Information3
- Foreign Body In Patient1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 13, 2025Defective Component; Defective Component
PRIOR TO START OF PROCEDURE, THE PURPLE DISSECTION TIP WAS REMOVED FROM KIT AND IT WAS NOTICED THAT THE TIP WAS NOT SMOOTH AND APPEARED TO HAVE RIDGES. PICTURE WAS PROVIDED BY HOSPITAL. THE TIP WAS NOT USED FOR PROCEDURE, A NEW ACCESSORY KIT WAS OPENED. TIP WAS DISCARDED BY ACCOUNT. PATIENT INFORMATION N/A AS IT WAS NOT USED ON A PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23543677 ↗InjuryJan 31, 2025Peeled/Delaminated; Environmental Particulates; Peeled/Delaminated; Environmental Particulates
MalfunctionJan 13, 2025Improper Flow or Infusion; Improper Flow or Infusion
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, CO2 WOULD NOT FLOW THROUGH THE BTT OF THE VV 6 ACCESSORY PACK (VH-2004). PATIENT WAS NOT INJURED. NO TIME WAS ADDED TO THE CASE. A NEW (BTT) WAS USED TO COMPLETE THE PROCEDURE AND THE SAME VV6 WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21143117 ↗MalfunctionJan 9, 2025Improper Flow or Infusion; Improper Flow or Infusion
MalfunctionOct 28, 2024Improper Flow or Infusion; Improper Flow or Infusion
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, VV 6 ACCESSORY PACK INSUFFLATOR WOULD NOT HOLD PRESSURE INSIDE THE PATIENT AND THE VEINS COULD NOT BE VISUALIZED. OPEN VEIN HARVESTING WAS DONE.
Patient problems: Insufficient Information
MAUDE report 20543474 ↗MalfunctionOct 7, 2024
THE VASOVIEW ENDOSCOPIC VESSEL HARVESTING SYSTEM'S INSUFFLATOR WOULD NOT HOLD PRESSURE INSIDE THE PATIENT AND THE VEINS COULD NOT BE VISUALIZED. OPEN VEIN HARVESTING WAS DONE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20383054 ↗MalfunctionSep 30, 2024Contamination /Decontamination Problem; Environmental Particulates; Contamination /Decontamination Problem; Environmental Particulates
MalfunctionApr 8, 2024Improper Flow or Infusion; Improper Flow or Infusion
MalfunctionApr 8, 2024Improper Flow or Infusion; Improper Flow or Infusion
RELATED TO (B)(6). THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, VV 6 ACCESSORY PACK BTT EXCHANGED FOR A NEW ONE VIA ACCESSORY KIT, CO2 STILL NOT FLOWING. CO2 TUBING EXCHANGED FOR NEW. STILL NO FLOW. ANOTHER BTT OPENED VIA ACCESSORY KIT. AFTER PROCEDURE THEY TRIED TO TROUBLE SHOOT AND WAS NOT ABLE TO DUPLICATE THE ISSUE. PROCEDURE COMPLETED WITH ORIGINAL HP2 DEVICE AND THE LAST OPENED BTT. MINIMAL PROCEDURE DELAY DUE TO TROUBLESHOOTING ISSUE. THERE WERE NO PATIENT INJURIES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19059675 ↗InjuryApr 3, 2024Improper Flow or Infusion; Improper Flow or Infusion
MalfunctionMar 6, 2024Improper Flow or Infusion; Improper Flow or Infusion
RELATED TO: (B)(4). THE HOSPITAL REPORTED THAT DURING ENDOSCOPIC VEIN HARVESTING PROCEDURE, VASOVIEW HEMOPRO 2 READ "OCCLUSION" WHEN IT WAS PLUGGED INTO CO2 INSUFFLATION. HARVESTER TRIED NEW INSUFFLATION TUBING WITH NO SUCCESS. WITH NO INSUFFLATION, SURGICAL ASSISTANT COULD NOT CONTINUE WITH LAPAROSCOPIC APPROACH AFTER MULTIPLE ATTEMPTS WITH TWO DIFFERENT HARVESTERS. NEW TOWER WITH INSUFFLATION WAS ATTEMPTED WITH NO SUCCESS EITHER.
Patient problems: Insufficient Information
MAUDE report 18844816 ↗MalfunctionNov 21, 2023Material Rupture; Material Rupture
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, VV 6 ACCESSORY PACK BALLOON ON THE BTT POPPED. A NEW KIT WAS OPENED TO COMPLETE THE PROCEDURE. ONLY DELAY WAS TO OPEN A NEW KIT. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18183177 ↗MalfunctionNov 16, 2021Material Rupture; Material Rupture
RELATED TO (B)(4). THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE USING VV 6 ACCESSORY PACK, THE BALLOON ON THE BTT ON THE VASOVIEW 3500 KIT SPONTANEOUSLY BURST. NO INJURY OR HARM TO THE PATIENT. THE CASE WAS COMPLETED WITH AN ADDITIONAL BTT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12820762 ↗MalfunctionNov 16, 2021Material Rupture; Material Rupture
RELATED TO TW (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12819447 ↗InjuryFeb 25, 2021No Flow; No Flow
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE USING PORT FROM VV 6 ACCESSORY PACK. INSUFFLATION WAS NOT WORKING. THEY SWAPPED THE PORT OUT AND STILL COULD NOT GET IT TO WORK. THIS IS THE SECOND PORT THAT THEY COULD NOT GET CO2 INSUFFLATION TO FLOW THROUGH THE PORT. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11378340 ↗
Other Laparoscope, General & Plastic Surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page