Acrobat-I Stabilizer OM-10000Recalls & reported problems
Maquet Cardiovascular LLC · Acrobat-i Vacuum Stabilizer System · Class 1
All models of Acrobat-I Stabilizer (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Maquet Acrobat-i vacuum stabilizer holds the heart steady during open-heart surgery. The 48 reports mostly describe the arm not locking or tightening so it would not stay in position, parts breaking or springing loose, and suction failing. One report ends in a death; no injuries are described. There is 1 open recall, started in June 2023, because the standard blade may not latch onto the activator drive, which can release a component into the patient or delay the procedure.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 51
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 51 adverse event reports that list this exact device (DI 00607567700581), received between Jul 28, 2020 and Sep 30, 2026 — classified as 1 death, 50 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (20 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jun 28, 2023Class IIOpen, Classified
The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.
Z-2274-2023 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Mechanical Problem20
- Positioning Failure12
- Break8
- Positioning Problem5
- Suction Failure3
- Illegible Information2
- Packaging Problem2
- Suction Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions38
- No Consequences Or Impact To Patient9
- Insufficient Information4
- No Patient Involvement4
- Death1
Recent adverse event reports
The 25 most recent of 51 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Suction Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50952874 ↗MalfunctionSep 23, 2026Mechanical Problem
The hospital reported that during MIDCAB Acrobat-i Stabilizer locking knob was not turning on first use so it was immediately replaced. No patient harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50865373 ↗MalfunctionAug 13, 2026Break; Mechanical Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50232782 ↗MalfunctionAug 5, 2026Mechanical Problem
MalfunctionMar 16, 2026Positioning Failure
THE HOSPITAL REPORTED THAT DURING THE PROCEDURE, THE ACROBAT-I STABILIZER SYSTEM LOCKING LEVER WAS NOT WORKING. IT KEPT ROTATING AND COULD NOT BE TIGHTENED. THE PROCEDURE WAS COMPLETED USING A NEW DEVICE. THERE WAS NO HARM OR DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24604462 ↗MalfunctionDec 23, 2025Mechanical Problem; Mechanical Problem
THE HOSPITAL REPORTED THAT DURING THE OPCAB PROCEDURE, THE ACROBAT-I STABILIZER WAS UNABLE TO LOCK INTO POSITION ON THE UA-5001 BLADE. WHEN ATTEMPTING TO TIGHTEN THE STABILIZER ARM, THE TIGHTENING KNOB CONTINUED TO SPIN FREELY. UPON REMOVING THE STABILIZER, IT WAS DISCOVERED THAT A PIECE OF PLASTIC HAD DETACHED. THE PLASTIC PIECE AND STABILIZER ARM WERE REMOVED, AND A NEW STABILIZER ARM WAS OPENED AND THE PROCEDURE WAS COMPLETED. THE PIECE WAS LOCATED OUTSIDE THE BODY, ON TOP OF DRAPED TOWELS ALONG SIDE THE PATIENT AND CHEST AND RETRACTOR BLADES, ADJACENT TO THE SURGEON. THE PIECE WAS REMOVED BY THE SURGEONS HAND, TO THE SCRUB TECHS HAND AND THEN ONTO THE BACK TABLE. NO INJURY. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23875688 ↗MalfunctionDec 22, 2025Mechanical Problem; Mechanical Problem
MalfunctionDec 3, 2025Suction Problem; Insufficient Device Problem Information; Suction Problem; Insufficient Device Problem Information
THE HOSPITAL REPORTED THAT THE SUCTION FUNCTION OF THE ACROBAT-I STABILIZER WAS NOT WORKING. THE AFFECTED DEVICE WAS REINSTALLED AND TESTED MULTIPLE TIMES; HOWEVER, THE ISSUE COULD NOT BE RESOLVED. THE MALFUNCTION OCCURRED DURING THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING A NEW DEVICE, RESULTING IN A SMALL DELAY. NO HARM TO THE PATIENT WAS REPORTED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23703961 ↗MalfunctionDec 1, 2025Mechanical Problem; Mechanical Problem
THE HOSPITAL REPORTED THAT BEFORE THE PROCEDURE, ACROBAT-I STABILIZER WAS BROKEN, IT WOULD NOT LATCH, WHEN IT WAS OPENED. THE DEVICE WOULD NOT LOCK DOWN ON THE RETRACTOR. ANOTHER STABILIZER WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23677779 ↗MalfunctionNov 25, 2025Mechanical Problem; Mechanical Problem
THE HOSPITAL REPORTED THAT THE PATIENT WAS IN THE OR WHEN THE DEFECT WAS NOTICED. UPON OPENING THE ACROBAT-I STABILIZER, THE LOCKING MECHANISM WAS SPRUNG. ANOTHER DEVICE WAS PULLED AND THE PROCEDURE WAS COMPLETED WITHOUT INJURY. THERE WAS NO PROCEDURAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23645639 ↗MalfunctionNov 21, 2025Mechanical Problem; Mechanical Problem
THE HOSPITAL REPORTED THAT DURING AN OPCAB PROCEDURE, THE ACROBAT-I STABILIZER WAS ATTACHED TO THE RETRACTOR. WHEN THEY TRIED TO TIGHTEN THE DEVICE, THE TIGHTENING KNOB JUST KEPT SPINNING AND WOULD NOT TIGHTEN.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23622506 ↗MalfunctionJul 29, 2025Mechanical Problem; Mechanical Problem
THE HOSPITAL REPORTED WHEN THE ADJUSTMENT KNOB WAS TURNED TO ITS FULLY TIGHTENED POSITION, THE ARM WAS EXPECTED TO BE SECURELY FIXED; HOWEVER, IT REMAINED LOOSE AND FAILED TO STABILIZE AS INTENDED. THE MEDICAL STAFF REPORTED THIS MALFUNCTION DURING SETUP. A NEW SET WAS IMMEDIATELY USED TO REPLACE THE DEFECTIVE ONE, AND THE SURGERY PROCEEDED WITHOUT ANY DELAY. THE ISSUE DID NOT AFFECT THE OPERATION, AND THE PATIENT WAS UNHARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22645618 ↗MalfunctionJul 2, 2025Break; Break
THE HOSPITAL REPORTED THAT BEFORE THE PROCEDURE IN THE OR, THE ACROBAT-I STABILIZER ARM BROKE. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A SLIGHT DELAY IN THE PROCEDURE. THERE WERE NO PATIENT EFFECTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22387490 ↗MalfunctionMay 7, 2025Break; Break
THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE USING ACROBAT-I STABILIZER, AS THE DEVICE WAS BEING TIGHTENED USING THE KNOB, A METAL PIECE "POPPED" OFF. THE PIECE WAS RETRIEVED AND SAVED ALONG WITH THE DEVICE FOR RETURN. THE DEVICE WAS STILL USED TO COMPLETE PROCEDURE. THERE WAS NO PROCEDURAL DELAY. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21981849 ↗MalfunctionFeb 19, 2025Mechanical Problem; Mechanical Problem
THE HOSPITAL REPORTED THAT DURING THE PROCEDURE, THE ACROBAT-I STABILIZER LOCKING MECHANISM WOULD NOT WORK TO STABILIZE THE DEVICE ONTO THE ACCESS RAIL (RETRACTOR). THERE WAS NO PATIENT HARM. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. MINIMAL DELAY WHILE OPENING A NEW DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21416158 ↗MalfunctionFeb 12, 2025Break; Break
THE HOSPITAL REPORTED THAT DURING THE USE OF THE ACROBAT-I STABILIZER (OM-10000), THEY DISCOVERED THAT INTERNAL SCREWS HAD LOOSENED, CAUSING COMPONENTS TO FALL OFF AND MAKING IT IMPOSSIBLE TO SECURE THEM. THE ISSUE WAS RESOLVED BY REPLACING THE (OM-10000) WITH A NEW UNIT. NO HARM WAS CAUSED TO THE PATIENT, AND THE SURGERY PROCEEDED WITHOUT DELAY. PER THE PHOTO, IT SHOWS THAT THE INTERNAL PARTS HAD BECOME LOOSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21370603 ↗MalfunctionJan 8, 2025Mechanical Problem; Suction Failure; Mechanical Problem; Suction Failure
MalfunctionOct 23, 2024Break; Break
THE HOSPITAL REPORTED ACROBAT-I STABILIZER METAL PIECE FROM THE SPRING CONNECTOR THAT ATTACHES THE STABILIZER TO THE RETRACTOR FELL OUT ON THE DRAPES. THEY USED THE DEVICE AS IS. A SLIGHT FEW MINUTES' DELAY AS THEY LOOKED FOR THE LOOSE PIECE. NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20514562 ↗MalfunctionAug 30, 2024Mechanical Problem; Mechanical Problem
THE HOSPITAL REPORTED DURING A CORONARY ARTERY BYPASS PROCEDURE, THE ACROBAT-I STABILIZER WOULD NOT TIGHTEN AND WAS STIFF. THERE WAS A 5 MINUTE PROCEDURAL DELAY WHEN WE HAD TO RUN AND GET A NEW STABILIZER. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT HARM OR INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20107593 ↗MalfunctionApr 17, 2024Mechanical Problem; Positioning Problem; Mechanical Problem; Positioning Problem
ON 28-MAR 2024, THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, ACROBAT-I STABILIZER CANTILEVER COULD NOT BE LOCKED. REPEATED ATTEMPTS WERE MADE, BUT THE CANTILEVER COULD NOT BE LOCKED AND FIXED. THEREFORE, A NEW CARDIAC STABILIZATION SYSTEM (ACROBAT-I STABILIZER) COULD BE LOCKED NORMALLY AFTER REPLACEMENT, AND THE SURGERY WAS SUCCESSFULLY COMPLETED WITHOUT ANY HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19125999 ↗MalfunctionMar 11, 2024Positioning Problem; Positioning Problem
THE HOSPITAL REPORTED THAT DURING PREPARATION FOR A CORONARY ARTERY BYPASS PROCEDURE, ACROBAT-I STABILIZER COULD NOT BE USED DUE TO FAILURE TO STABILIZE IT. THEY DETECTED THIS FAILURE IN THE PRODUCT IN THE PREVIOUS STEP, BEFORE HAVING TO USE IT AND HAD TO RESORT TO ANOTHER UNIT THAT HE HAD AVAILABLE AT THAT TIME. THIS SITUATION DID NOT CAUSE ANY TYPE OF PROBLEM FOR THE PATIENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 18875031 ↗MalfunctionJan 30, 2024Positioning Problem; Positioning Problem
THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, ACROBAT-I STABILIZER SURGEON WAS NOT ABLE TO TIGHTEN THE STABILIZER KNOB. IT KEPT ROTATING AND DIDN'T GET TIGHT. PROCEDURE COMPLETED WITH A NEW DEVICE. THERE WERE NO PROCEDURAL DELAYS AND PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18608747 ↗MalfunctionJan 26, 2024Mechanical Problem; Mechanical Problem
THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, ACROBAT-I STABILIZER WOULDN'T STABILIZE AND WOULDN¿T HOLD ON RAIL TO STABILIZE. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO PROCEDURAL DELAY. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18592211 ↗MalfunctionDec 8, 2023Positioning Problem; Positioning Problem
MalfunctionApr 21, 2023Positioning Failure; Positioning Failure
THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE USING ACROBAT-I STABILIZER, THEY HEARD A STRANGE SOUND WHILE ROTATING THE KNOB AND INSTALLING THE EQUIPMENT. FAILURE TO LOCK ONTO RETRACTOR ARM WAS NOT ABLE TO TIGHTEN IN FIXED POSITION EVEN ROTATED THE KNOB TO HEAR THE CLICK SOUND. AFTER THAT THE STABILIZER IS WORKING NOT FUNCTIONAL, THEY DECIDED TO REPLACE IT WITH A NEW ONE IMMEDIATELY. NO PROCEDURAL DELAY. NO HARM TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16788885 ↗
Other Stabilizer,Heart devices
- Acrobat-I Stabilizer OM-10000ZMaquet Cardiovascular LLC · Stabilizer,Heart2 recalls120 reports
- Acrobat-i Positioner XP-5000ZMaquet Cardiovascular LLC · Stabilizer,Heart2 recalls34 reports
- Acrobat SUV Stabilizer OM-9000ZMaquet Cardiovascular LLC · Stabilizer,Heart1 recall19 reports
- OCTOPUS® TS2000Medtronic, Inc. · stabilizer,heart1 recall15 reports
- Herc Flex Universal Stabilizer Arm T401163Terumo Cardiovascular Systems Corporation · Stabilizer, Heart1 recall10 reports
- Acrobat V Stabilizer OM-9100SMaquet Cardiovascular LLC · Stabilizer,Heart1 recall7 reports
- OCTOPUS® TSMICS1Medtronic, Inc. · Stabilizer, heart1 recall5 reports
- Acrobat SUV Stabilizer OM-9000SMaquet Cardiovascular LLC · Stabilizer,Heart1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page