Skip to content

OCTOPUS® TS2500Recalls & reported problems

Medtronic, Inc. · STABILIZER TS2500 OCTO EVOLUTION AS 26L · Class 1

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169668904), received between Apr 20, 2022 and Jan 28, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Delivered as Unsterile Product” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 12, 2026Open, Classified

    During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.

    Z-2576-2026 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Delivered as Unsterile Product2
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 28, 2026Delivered as Unsterile Product

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF AN OCTOPUS TISSUE STABILIZER, IT WAS REPORTED THAT THE STERILE BOX OF THE PRODUCT WAS ALREADY OPEN, SO THE CUSTOMER QUESTIONED THE STERILITY OF THE PRODUCT. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE NO MISSING OR WRONG DEVICES OR COMPONENTS IN THE PACKAGE. THE STERILE PACKAGE WAS NOT PROPERLY CLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24198624 ↗
  • MalfunctionJan 19, 2026Delivered as Unsterile Product

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF AN OCTOPUS TISSUE STABILIZER, IT WAS REPORTED THAT WHEN RECEIVING GOODS FROM THE CARRIER'S FORWARDER FROM MEDTRONIC TO AN PHA. THE CUSTOMER OBSERVED THAT 01 BOX HAD DAMAGED PACKAGING, THE PRODUCT INSIDE HAD BLACK STAINS DUE TO DAMAGED OUTER SHELL. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE PACKAGE IS INTACT, CUSTOMER KEPT THE DEFECTIVE PACKAGE AS IT WAS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24108300 ↗
  • MalfunctionApr 20, 2022Break; Break

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THIS OCTOPUS TISSUE STABILIZER, THE CUSTOMER REPORTED THAT WHEN THEY TOOK THE DEVICE OUT OF IT'S PACKAGING, THEY NOTED THAT HEAD WAS SEPARATED FROM THE BODY. THE DEVICE WAS REPLACED FOR THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO DAMAGE TO THE PACKAGING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14159179 ↗

Other stabilizer,heart devices

All stabilizer,heart devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.