Syntel Thrombectomy Catheter A4545Recalls & reported problems
Applied Medical Resources Corporation · Embolectomy/thrombectomy balloon catheter · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00607915110925), received between May 6, 2016 and Nov 29, 2017 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Detachment Of Device Component” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment Of Device Component3
- Break1
- Material Rupture1
- Material Separation1
Patient problems
- No Consequences Or Impact To Patient5
- Foreign Body In Patient3
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 29, 2017Material Separation; Material Separation
INCIDENT DESCRIPTION: "THE TIP BROKE OFF. " ADDITIONAL INFORMATION WAS RECEIVED VIA EMAIL FROM RN, (B)(6) ON FRIDAY (B)(6) 2017 AT 6:21 AM: "THROMBECTOMY CATHETER INSERTED THROUGH MERIT 8F SHEATH (PSS-8F-4MT) INTO THE BRACHIAL ARTERY. INFLATED AND PULLED BACK TO TIP OF SHEATH, DEFLATED AND REMOVED. THIS WAS DOWN SUCCESSFULLY TWICE. ON THE THIRD TIME, THE TIP OF THE CATHETER CAME OFF INSIDE THE PATIENT'S GRAFT. NEW CATHETER (A4F04) WAS USED TO TRY TO RETRIEVE THE TIP, BUT WAS UNSUCCESSFUL. AT THIS TIME, THERE IS NO ILL EFFECT TO THE PATIENT. THE FAULTY PRODUCT WAS NOT SAVED TO BE RETURNED. TYPE OF INTERVENTION: NONE. PATIENT STATUS: "THERE IS NO ILL EFFECT TO THE PATIENT."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7068026 ↗MalfunctionOct 24, 2017Break; Break
PROCEDURE PERFORMED: "THROMBECTOMY OF DIALYSIS GRAFT" INCIDENT DESCRIPTION: THIS CER IS IN CORRELATION WITH (B)(4). "JUST HAD ANOTHER TIP BREAK OFF, BUT NOT ABLE TO RETRIEVE, STILL IN PATIENT." ADDITIONAL INFORMATION WAS RECEIVED VIA EMAIL FROM REGISTERED NURSE, (B)(6) ON MONDAY, (B)(6) 2017 AT 9:40 AM. "AFTER INSERTING THE FOGARTY THROUGH THE SHEATH, IT WAS INFLATED AND PULLED BACK PER NORMAL USE. ON EACH OF THE THREE OCCASIONS, THE TIP OF THE FOGARTY CAME OFF INSIDE THE GRAFT. IN THE FIRST PROCEDURE ((B)(6) 2017), THE TIPS WERE ABLE TO BE RECOVERED USING ANOTHER FOGARTY WHILE APPLYING SUCTION ON THE PIGTAIL OF THE SHEATH. THE SECOND PROCEDURE ((B)(6) 2017) THE TIP WAS UNABLE TO BE RECOVERED. FIRST PATIENT NO ILL EFFECTS D/T THE FACT THAT THE TIPS WERE RECOVERED BEFORE FLOW WAS RE-ESTABLISHED. THE SECOND PATIENT HAD NO ILL EFFECT EVEN THOUGH THE TIP COULD NOT BE RECOVERED." ADDITIONAL INFORMATION WAS RECEIVED VIA TELEPHONE FROM REGISTERED NURSE, (B)(6) ON MONDAY, 16-OCT-2017 AT 10:38 AM "I WILL HAVE TO CHECK ON THE INCIDENT DATE THAT OCCURRED ON THE FIRST TWO INCIDENTS. I PUT THE WRONG PRODUCT CODE IN MY EMAIL, IT SHOULD BE THE A4545 CATHETER. THERE IS NO POSSIBLE WAY FOR US TO RE
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6973866 ↗MalfunctionOct 9, 2017Detachment Of Device Component; Detachment Of Device Component
PROCEDURE PERFORMED: THROMBECTOMY OF FOREARM AV GRAFT DIALYSIS INCIDENT DESCRIPTION: CER 1 OF 2: (B)(4), CER 2 OF 2: (B)(4). INSERTED A FRENCH MERIT SHEATH INTO THE GRAFT. INFLATED THE CATHETER AND PULLED BACK TO REMOVE THE CLOT. DR NOTICED THE BALLOON OF THE CATHETER WAS DEFLATED AND THAT THE TIP OF THE CATHETER HAD BROKEN INSIDE THE GRAFT. TRIED TO COMPLETE WITH 2ND A4545 OF THE SAME LOT NUMBER AND THE SAME THING HAPPENED. USED ANOTHER A4545 AND THE THIRD CATHETER WORKED. THE SAME LOT NUMBER FOR ALL THREE CATHETERS. RETRIEVED BOTH TIPS, BUT ONLY ONE IS AVAILABLE TO RETURN AS THE DOCTOR DISCARDED 1 OF THE TIPS. TYPE OF INTERVENTION: USED THIRD A4545 TO COMPLETE THE PROCEDURE PATIENT STATUS: "NO ILL EFFECT."
Patient problems: Foreign Body In Patient; No Consequences Or Impact To Patient
MAUDE report 6930148 ↗MalfunctionOct 9, 2017Detachment Of Device Component; Detachment Of Device Component
PROCEDURE PERFORMED: THROMBECTOMY OF FOREARM AV GRAFT DIALYSIS. INCIDENT DESCRIPTION: CER 1 OF 2: (B)(4), CER 2 OF 2: (B)(4). INSERTED A FRENCH MERIT SHEATH INTO THE GRAFT. INFLATED THE CATHETER AND PULLED BACK TO REMOVE THE CLOT. DR NOTICED THE BALLOON OF THE CATHETER WAS DEFLATED AND THAT THE TIP OF THE CATHETER HAD BROKEN INSIDE THE GRAFT. TRIED TO COMPLETE WITH 2ND A4545 OF THE SAME LOT NUMBER AND THE SAME THING HAPPENED. USED ANOTHER A4545 AND THE THIRD CATHETER WORKED. THE SAME LOT NUMBER FOR ALL THREE CATHETERS. RETRIEVED BOTH TIPS, BUT ONLY ONE IS AVAILABLE TO RETURN AS THE DOCTOR DISCARDED 1 OF THE TIPS. TYPE OF INTERVENTION: USED THIRD A4545 TO COMPLETE THE PROCEDURE. PATIENT STATUS: "NO ILL EFFECT."
Patient problems: Foreign Body In Patient; No Consequences Or Impact To Patient
MAUDE report 6930134 ↗MalfunctionJun 27, 2017Material Rupture; Material Rupture
PROCEDURE PERFORMED: THROMBECTOMY. INCIDENT DESCRIPTION: AFTER DR. GRANT WENT TO INSUFFLATE BALLOON, HE PASSED IT THROUGH THE VESSEL AND AFTER PULLING BALLOON OUT, HE NOTICED THE BALLOON WAS RUPTURED. IT WAS NOTED THERE WAS SOME KIND OF SMALL FIBER ATTACHED TO THE BALLOON. THIS HAPPENED IN THE BEGINNING OF THE CASE. THE CATHETER DID NOT COME INTO CONTACT WITH ANY SHARP OBJECTS THROUGHOUT THE CASE. PATIENT STATUS: THERE WAS NO PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6672064 ↗MalfunctionMay 6, 2016Detachment Of Device Component; Detachment Of Device Component
THROMBECTOMY - "WHILE USING THE CATHETER FOR A THROMBECTOMY, PHYSICIAN INJECTED BALLOON 0.7 ML OF CONTRAST. AFTER DEFLATING THE BALLOON AND REMOVING THE CATHETER THE DOCTOR NOTICED THAT THE TIP OF THE CATHETER WAS MISSING. THE CLEAR PLASTIC TIP OVER THE COIL SPRING WAS GONE. STAFF NOTICED AN OBJECT IN THE PATIENT'S WRIST FOR A SHORT TIME BUT DISAPPEARED AFTER THE CONTRAST DISPERSED. THE STAFF WAS THEN UNABLE TO FIND THE CATHETER TIP UNDER FLURO DUE TO THE OBJECT NOT BEING RADIO OPAQUE." PATIENT STATUS - THE PATIENT REMAINED ASYMPTOMATIC THROUGHOUT THE CASE AS WELL AS AFTERWARDS.
Patient problems: Foreign Body In Patient; No Consequences Or Impact To Patient
MAUDE report 5639151 ↗
Other Catheter, Embolectomy devices
- Indigo System CAT6Penumbra, Inc. · CATHETER, EMBOLECTOMY2 deaths408 reports
- Indigo System CATRXKITPenumbra, Inc. · Catheter, Embolectomy1 death339 reports
- Fogarty 120804FPEdwards Lifesciences LLC · Catheter, Embolectomy1 recall1 death200 reports
- LeMaitre Embolectomy Catheter 1651-88Lemaitre Vascular, Inc. · Catheter, Embolectomy1 recall161 reports
- FOGARTY 120404FPEdwards Lifesciences LLC · Catheter, Embolectomy1 recall61 reports
- Indigo System CAT6KIT-APenumbra, Inc. · Catheter, Embolectomy109 reports
- LeMaitre Embolectomy Catheter e1651-88Lemaitre Vascular, Inc. · Catheter, Embolectomy2 recalls4 reports
- Indigo System CAT8XTORQ115Penumbra, Inc. · CATHETER, EMBOLECTOMY2 deaths98 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page