SOFT / FIBRA® CLAMP INSERT G-3350Recalls & reported problems

Applied Medical Resources Corporation · SURGICAL CLAMP INSERTS · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00607915111243), received between Apr 7, 2017 and Nov 21, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Loss of or Failure to Bond” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loss of or Failure to Bond2
  • Detachment Of Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 21, 2025Loss of or Failure to Bond; Loss of or Failure to Bond

    PROCEDURE PERFORMED: CARDIAC SURGERY. EVENT DESCRIPTION: [HOSPITAL NAME], MODEL: G3350, LOT: 1548224. ON (B)(6) 2025: [PROCEDURE: CARDIAC SURGERY]. IT WAS IDENTIFIED THAT THE RUBBER PART OF THE INSERT DETACHED FROM THE PRODUCT DURING USE. THE SURGEON THEN CHECKED ANOTHER INSERT THAT HAD BEEN USED FOR THE BLOCKAGE AND FOUND THAT THE SAME ISSUE HAD OCCURRED WITH THAT INSERT AS WELL. THIS INDICATES THAT THE TOTAL QUANTITY OF AFFECTED PRODUCTS WAS TWO UNITS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED BY REPLACING THEM WITH NEW ONES. PLEASE NOTE THAT BOTH DETACHMENTS WERE ONLY PARTIAL AND THE RUBBER PARTS DID NOT FALL INTO THE PATIENT¿S BODY. CONSEQUENTLY, THESE ISSUES DID NOT NEGATIVELY AFFECT THE PATIENT. THE PRODUCTS IN QUESTION WERE DISPOSED OF AT THE HOSPITAL. PATIENT STATUS: NO CLINICAL IMPACT. INTERVENTION: THESE DEVICES WERE REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23624848 ↗
  • MalfunctionNov 21, 2025Loss of or Failure to Bond; Loss of or Failure to Bond

    PROCEDURE PERFORMED: CARDIAC SURGERY. EVENT DESCRIPTION: [HOSPITAL NAME] MODEL G3350 LOT 1548224 ON (B)(6) 2025 [PROCEDURE: CARDIAC SURGERY]. IT WAS IDENTIFIED THAT THE RUBBER PART OF THE INSERT DETACHED FROM THE PRODUCT DURING USE. THE SURGEON THEN CHECKED ANOTHER INSERT THAT HAD BEEN USED FOR THE BLOCKAGE AND FOUND THAT THE SAME ISSUE HAD OCCURRED WITH THAT INSERT AS WELL. THIS INDICATES THAT THE TOTAL QUANTITY OF AFFECTED PRODUCTS WAS TWO UNITS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED BY REPLACING THEM WITH NEW ONES. PLEASE NOTE THAT BOTH DETACHMENTS WERE ONLY PARTIAL AND THE RUBBER PARTS DID NOT FALL INTO THE PATIENT¿S BODY. CONSEQUENTLY, THESE ISSUES DID NOT NEGATIVELY AFFECT THE PATIENT. THE PRODUCTS IN QUESTION WERE DISPOSED OF AT THE HOSPITAL. PATIENT STATUS: NO CLINICAL IMPACT. INTERVENTION: THESE DEVICES WERE REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23623643 ↗
  • MalfunctionApr 7, 2017Detachment Of Device Component; Detachment Of Device Component

    EVENT DESCRIPTION: (ON (B)(6) 2017: DURING PROCEDURE) A MEDICAL STAFF ATTACHED THE INSERT TO A CLAMP. THEN THE MEDICAL STAFF FOUND THAT THE INSERT WHITE RUBBER SEPARATED FROM THE PLASTIC BASE. ANOTHER INSERT WAS USED FOR THE PROCEDURE AND THE PROCEDURE WAS COMPLETED WITH NO PROBLEM. THIS EVENT DIDN'T AFFECT THE PATIENT. TYPE OF INTERVENTION: NA. PATIENT STATUS: "NO HEALTH DAMAGE WAS REPORTED FOR THE PATIENT."

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6474312 ↗

Other Clamp, Vascular devices

All Clamp, Vascular devices →