Stealth® Clamps A3203Recalls & reported problems
Applied Medical Resources Corporation · Vascular Clamp, Reusable · Class 2
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 00607915112721), received between Mar 30, 2017 and Mar 2, 2026 — classified as 16 malfunctions.
The most frequently reported specific device problem is “Difficult to Open or Close” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Open or Close10
- Migration or Expulsion of Device4
- Patient Device Interaction Problem2
- Failure to Align1
Patient problems
- No Known Impact Or Consequence To Patient14
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 2, 2026Patient Device Interaction Problem
NAME OF PROCEDURE BEING PERFORMED: UROLOGIC PROCEDURE ON THE SPERMATIC CORD. EVENT DESCRIPTION: PARALLEL CLAMPING NO LONGER POSSIBLE DUE TO FREQUENT USE. IT'S A SERVICE REQUEST: ORIGINAL CONDITIONS PROBABLY NOT REPAIRABLE. PLEASE PROVIDE A FREE REPLACEMENT. PRODUCT IS ALREADY IN AME. OR NURSE SENDS THE A3203 CLAMP TO STERILIZATION. ROUTINE FUNCTIONALITY CHECK ACCORDING TO THEIR STANDARDS: STERILIZATION SENDS IT TO US WITH A SERVICE REQUEST. NO PATIENTS AFFECTED. INTERVENTION: THEY USE THEIR SECOND CLAMP. PATIENT STATUS: NA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24488181 ↗MalfunctionAug 19, 2025Failure to Align; Patient Device Interaction Problem; Failure to Align; Patient Device Interaction Problem
PROCEDURE PERFORMED: UROLOGICAL PROCEDURE ON THE SPERMATIC CORD. EVENT DESCRIPTION: THE STEALTH CLAMP WAS USED REGULARLY FOR A UROLOGICAL PROCEDURE ON THE SPERMATIC CORD. REASON: TEAR DUE TO FREQUENT USE. PARALLEL CLAMPING NO LONGER POSSIBLE DUE TO FREQUENT USE. ADDITIONAL INFORMATION RECEIVED BY APPLIED MEDICAL REP VIA EMAIL ON 11AUG25: THE SURGEON DESCRIBES IT COLLOQUIALLY AS LOOKING LIKE ¿FINGER CROSSING¿ OF THE JAWS. IT WAS NOTICED BEFORE USE. PATIENT STATUS: NO CLINICAL IMPACT TO THE PATIENT. INTERVENTION: USER USED HIS SECOND A3203 STEALTH CLAMP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22841566 ↗MalfunctionApr 22, 2017Difficult to Open or Close; Difficult to Open or Close
THE DOCTOR MENTIONED TO THE INSTRUMENT SPECIALIST THAT WHEN THE INSERT IS ATTACHED TO THE JAW, IT IS DIFFICULT FOR THE CLAMP TO MAINTAIN CLAPPED IN THE LOCKED POSITION. THE DOCTOR USED THE A0C02 INSERTS. APPLIED MEDICAL RECEIVED ADDITIONAL INFORMATION ON MARCH 14,2017 FROM THE CUSTOMER VIA PHONE: THE CUSTOMER HAS CONFIRMED THAT WHEN THE JAW IS CLAMPED CLOSED IT WOULD EASILY UNCLASP. IF THE HANDLE WERE TO BE TAPPED THE CLAMP WOULD RELEASE. PREVAC STEAM WAS USED TO STERILIZATION . PROCEDURE PERFORMED: UNK TYPE OF INTERVENTION: UNK PATIENT STATUS: THE CALLER WAS NOT INFORMED OF ANY PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6513676 ↗MalfunctionApr 22, 2017Difficult to Open or Close; Difficult to Open or Close
THE DOCTOR MENTIONED TO THE INSTRUMENT SPECIALIST THAT WHEN THE INSERT IS ATTACHED TO THE JAW IT IS DIFFICULT FOR THE CLAMP TO MAINTAIN CLAPPED IN THE LOCKED POSITION. THE DOCTOR USED THE A0C02 INSERTS. APPLIED MEDICAL RECEIVED ADDITIONAL INFORMATION ON MARCH 14,2 017 FROM THE CUSTOMER VIA PHONE: THE CUSTOMER HAS CONFIRMED THAT WHEN THE JAW IS CLAMPED CLOSED IT WOULD EASILY UNCLASP. IF THE HANDLE WERE TO BE TAPPED THE CLAMP WOULD RELEASE. PREVAC STEAM WAS USED TO STERILIZATION. PROCEDURE PERFORMED: UNK. TYPE OF INTERVENTION: UNK. PATIENT STATUS: THE CALLER WAS NOT INFORMED OF ANY PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6513673 ↗MalfunctionApr 22, 2017Migration or Expulsion of Device; Migration or Expulsion of Device
PROCEDURE PERFORMED: NA. THE CUSTOMER WAS TESTING THEIR CLAMPS AND IDENTIFIED THAT AT WHEN THE JAW IS CLAMPED CLOSED IT WOULD EASILY UNLOCK. TYPE OF INTERVENTION: NA. PATIENT STATUS: NA.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6513667 ↗MalfunctionApr 22, 2017Migration or Expulsion of Device; Migration or Expulsion of Device
PROCEDURE PERFORMED: NA. THE CUSTOMER WAS TESTING THEIR CLAMPS AND IDENTIFIED THAT AT WHEN THE JAW IS CLAMPED CLOSED IT WOULD EASILY UNLOCK. TYPE OF INTERVENTION: NA. PATIENT STATUS: NA.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6513665 ↗MalfunctionApr 22, 2017Migration or Expulsion of Device; Migration or Expulsion of Device
PROCEDURE PERFORMED: NA. THE CUSTOMER WAS TESTING THEIR CLAMPS AND IDENTIFIED THAT AT WHEN THE JAW IS CLAMPED CLOSED IT WOULD EASILY UNLOCK. TYPE OF INTERVENTION: NA. PATIENT STATUS: NA.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6513664 ↗MalfunctionApr 22, 2017Migration or Expulsion of Device; Migration or Expulsion of Device
PROCEDURE PERFORMED: NA. THE CUSTOMER WAS TESTING THEIR CLAMPS AND IDENTIFIED THAT AT WHEN THE JAW IS CLAMPED CLOSED IT WOULD EASILY UNLOCK. TYPE OF INTERVENTION: NA. PATIENT STATUS: NA.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6513663 ↗MalfunctionApr 7, 2017Difficult to Open or Close; Difficult to Open or Close
PROCEDURE PERFORMED: UNKNOWN - THE DOCTOR MENTIONED TO THE INSTRUMENT SPECIALIST THAT WHEN THE INSERT IS ATTACHED TO THE JAW IT IS DIFFICULT FOR THE CLAMP TO MAINTAIN CLAPPED IN THE LOCKED POSITION. THE DOCTOR USED THE A0C02 INSERTS. APPLIED MEDICAL RECEIVED ADDITIONAL INFORMATION ON MARCH 14,2017 FROM THE CUSTOMER VIA PHONE: THE CUSTOMER HAS CONFIRMED THAT WHEN THE JAW IS CLAMPED CLOSED IT WOULD EASILY UNCLASP. IF THE HANDLE WERE TO BE TAPPED THE CLAMP WOULD RELEASE. PREVAC STEAM WAS USED TO STERILIZATION. TYPE OF INTERVENTION: UNK. PATIENT STATUS: THE CALLER WAS NOT INFORMED OF ANY PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6474311 ↗MalfunctionApr 7, 2017Difficult to Open or Close; Difficult to Open or Close
PROCEDURE PERFORMED: UNKNOWN - THE DOCTOR MENTIONED TO THE INSTRUMENT SPECIALIST THAT WHEN THE INSERT IS ATTACHED TO THE JAW IT IS DIFFICULT FOR THE CLAMP TO MAINTAIN CLAPPED IN THE LOCKED POSITION. THE DOCTOR USED THE A0C02 INSERTS. APPLIED MEDICAL RECEIVED ADDITIONAL INFORMATION ON MARCH 14,2017 FROM THE CUSTOMER VIA PHONE: THE CUSTOMER HAS CONFIRMED THAT WHEN THE JAW IS CLAMPED CLOSED IT WOULD EASILY UNCLASP. IF THE HANDLE WERE TO BE TAPPED THE CLAMP WOULD RELEASE. PREVAC STEAM WAS USED TO STERILIZATION TYPE OF INTERVENTION: UNK. PATIENT STATUS: THE CALLER WAS NOT INFORMED OF ANY PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6474310 ↗MalfunctionApr 5, 2017Difficult to Open or Close; Difficult to Open or Close
PROCEDURE PERFORMED: UNKNOWN - THE DOCTOR MENTIONED TO THE INSTRUMENT SPECIALIST THAT WHEN THE INSERT IS ATTACHED TO THE JAW IT IS DIFFICULT FOR THE CLAMP TO MAINTAIN CLAPPED IN THE LOCKED POSITION. THE DOCTOR USED THE A0C02 INSERTS. APPLIED MEDICAL RECEIVED ADDITIONAL INFORMATION ON MARCH 14,2017 FROM THE CUSTOMER VIA PHONE: THE CUSTOMER HAS CONFIRMED THAT WHEN THE JAW IS CLAMPED CLOSED IT WOULD EASILY UNCLASP. IF THE HANDLE WERE TO BE TAPPED THE CLAMP WOULD RELEASE. PREVAC STEAM WAS USED TO STERILIZATION. ADDITIONAL CERS CREATED 2017-0423, 2017-0463, 2017-0464, 2017-0465, 2017-0466, 2017-0467, 2017-0468, 2017-0469, 2017-0470, 2017-0471, 2017-0472, 2017-0473, 2017-0474. TYPE OF INTERVENTION: UNK. PATIENT STATUS: THE CALLER WAS NOT INFORMED OF ANY PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6462520 ↗MalfunctionApr 5, 2017Difficult to Open or Close; Difficult to Open or Close
PROCEDURE PERFORMED: UNKNOWN - THE DOCTOR MENTIONED TO THE INSTRUMENT SPECIALIST THAT WHEN THE INSERT IS ATTACHED TO THE JAW IT IS DIFFICULT FOR THE CLAMP TO MAINTAIN CLAPPED IN THE LOCKED POSITION. THE DOCTOR USED THE A0C02 INSERTS. APPLIED MEDICAL RECEIVED ADDITIONAL INFORMATION ON MARCH 14,2017 FROM THE CUSTOMER VIA PHONE: THE CUSTOMER HAS CONFIRMED THAT WHEN THE JAW IS CLAMPED CLOSED IT WOULD EASILY UNCLASP. IF THE HANDLE WERE TO BE TAPPED THE CLAMP WOULD RELEASE. PREVAC STEAM WAS USED TO STERILIZATION (B)(4). TYPE OF INTERVENTION: UNK. PATIENT STATUS: THE CALLER WAS NOT INFORMED OF ANY PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6462501 ↗MalfunctionApr 5, 2017Difficult to Open or Close; Difficult to Open or Close
PROCEDURE PERFORMED: UNKNOWN - THE DOCTOR MENTIONED TO THE INSTRUMENT SPECIALIST THAT WHEN THE INSERT IS ATTACHED TO THE JAW IT IS DIFFICULT FOR THE CLAMP TO MAINTAIN CLAPPED IN THE LOCKED POSITION. THE DOCTOR USED THE A0C02 INSERTS. APPLIED MEDICAL RECEIVED ADDITIONAL INFORMATION ON MARCH 14,2017 FROM THE CUSTOMER VIA PHONE: THE CUSTOMER HAS CONFIRMED THAT WHEN THE JAW IS CLAMPED CLOSED IT WOULD EASILY UNCLASP. IF THE HANDLE WERE TO BE TAPPED THE CLAMP WOULD RELEASE. PREVAC STEAM WAS USED TO STERILIZATION ADDITIONAL CERS CREATED (B)(4). TYPE OF INTERVENTION: UNK. PATIENT STATUS: THE CALLER WAS NOT INFORMED OF ANY PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6462428 ↗MalfunctionApr 5, 2017Difficult to Open or Close; Difficult to Open or Close
PROCEDURE PERFORMED: UNKNOWN - THE DOCTOR MENTIONED TO THE INSTRUMENT SPECIALIST THAT WHEN THE INSERT IS ATTACHED TO THE JAW IT IS DIFFICULT FOR THE CLAMP TO MAINTAIN CLAPPED IN THE LOCKED POSITION. THE DOCTOR USED THE A0C02 INSERTS. APPLIED MEDICAL RECEIVED ADDITIONAL INFORMATION ON MARCH 14,2017 FROM THE CUSTOMER VIA PHONE: THE CUSTOMER HAS CONFIRMED THAT WHEN THE JAW IS CLAMPED CLOSED IT WOULD EASILY UNCLASP. IF THE HANDLE WERE TO BE TAPPED THE CLAMP WOULD RELEASE. PREVAC STEAM WAS USED TO STERILIZATION. (B)(4). TYPE OF INTERVENTION: UNK. PATIENT STATUS: THE CALLER WAS NOT INFORMED OF ANY PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6462411 ↗MalfunctionApr 5, 2017Difficult to Open or Close; Difficult to Open or Close
PROCEDURE PERFORMED: UNKNOWN - THE DOCTOR MENTIONED TO THE INSTRUMENT SPECIALIST THAT WHEN THE INSERT IS ATTACHED TO THE JAW IT IS DIFFICULT FOR THE CLAMP TO MAINTAIN CLAPPED IN THE LOCKED POSITION. THE DOCTOR USED THE A0C02 INSERTS. APPLIED MEDICAL RECEIVED ADDITIONAL INFORMATION ON MARCH 14,2017 FROM THE CUSTOMER VIA PHONE: THE CUSTOMER HAS CONFIRMED THAT WHEN THE JAW IS CLAMPED CLOSED IT WOULD EASILY UNCLASP. IF THE HANDLE WERE TO BE TAPPED THE CLAMP WOULD RELEASE. PREVAC STEAM WAS USED TO STERILIZATION. ADDITIONAL CERS CREATED (B)(4). TYPE OF INTERVENTION: UNK. PATIENT STATUS: THE CALLER WAS NOT INFORMED OF ANY PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6462355 ↗MalfunctionMar 30, 2017Difficult to Open or Close; Difficult to Open or Close
PROCEDURE PERFORMED UNKNOWN. THE DOCTOR MENTIONED TO THE INSTRUMENT SPECIALIST THAT WHEN THE INSERT IS ATTACHED TO THE JAW IT IS DIFFICULT FOR THE CLAMP TO MAINTAIN CLAPPED IN THE LOCKED POSITION. THE DOCTOR USED THE A0C02 INSERTS. APPLIED MEDICAL RECEIVED ADDITIONAL INFORMATION ON MARCH 14,2017 FROM THE CUSTOMER VIA PHONE: THE CUSTOMER HAS CONFIRMED THAT WHEN THE JAW IS CLAMPED CLOSED IT WOULD EASILY UNCLASP. IF THE HANDLE WERE TO BE TAPPED THE CLAMP WOULD RELEASE. PREVAC STEAM WAS USED TO STERILIZATION TYPE OF INTERVENTION: UNK. PATIENT STATUS: THE CALLER WAS NOT INFORMED OF ANY PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6446726 ↗
Other Clamp, Vascular devices
- Heartstring III System 3.8MM HSK-3038Maquet Cardiovascular LLC · Clamp, Vascular1 recall4 deaths809 reports
- Heartstring III System 4.3MM HSK-3043Maquet Cardiovascular LLC · Clamp, Vascular1 recall265 reports
- TR BAND TRB24-REGTerumo Medical Corporation · CLAMP, VASCULAR284 reports
- TR BAND TRB24-REGTerumo Medical Corporation · CLAMP, VASCULAR256 reports
- HeartString III Proximal Seal HS-3045Maquet Cardiovascular LLC · Clamp, Vascular1 recall187 reports
- TR BAND TRB29-LRGTerumo Medical Corporation · CLAMP, VASCULAR67 reports
- TR BAND TRB29-LRGTerumo Medical Corporation · CLAMP, VASCULAR65 reports
- Flo-Thru Intraluminal Shunt® FT-12225Synovis Life Technologies, Inc. · CLAMP, VASCULAR1 recall1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.