STEALTH® CLAMP INSERTS A0A13Recalls & reported problems

Applied Medical Resources Corporation · CLAMP INSERTS · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00607915113001), received between Feb 13, 2026 and Feb 25, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device Slipped” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Slipped3

Patient problems

  • Hemorrhage/Blood Loss/Bleeding3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 25, 2026Device Slipped

    PROCEDURE PERFORMED: DESCENDING THORACIC AORTA TO BILATERAL FEMORAL BYPASS. EVENT DESCRIPTION: I AM FROM "[HOSPITAL]" IN (B)(6). ONE OF OUR SURGEONS REALLY LIKES THE AORTIC STEALTH CLAMPS, HOWEVER THE CLAMP INSERTS ARE QUITE SLIPPERY AND HAVE ACTUALLY COME OFF THE AORTA A FEW DIFFERENT TIMES. DO YOU HAVE A MORE 'STICKY' INSERT THAT WILL NOT SLIDE OFF OF THE AORTA? HE IS USING THEM CORRECTLY. ARE OTHER HOSPITALS REPORTING THIS PROBLEM? HE HAS UNFORTUNATELY, STOPPED USING THEM AS HE CANNOT RISK THEM SLIDING DURING THE PROCEDURE. THREE DEVICE MALFUNCTIONS OCCURRED DURING THREE DIFFERENT PROCEDURES. THE SAME DOCTOR EXPERIENCED THE ISSUE THREE TIMES. IT HAPPENED A COUPLE OF TIMES, THE RUBBER DOESN'T GRIP THE TISSUE THE WHOLE CLAMP FALLS OFF THE AORTA. THE TYPE OF INSERT TYPICALLY USED IS STEALTH INSERT. THERE WAS NO SEPARATION OF THE INSERT PAD FROM THE BASE. ADDITIONAL INFORMATION PROVIDED BY "[NAME]" ON (B)(6) 2026 BY EMAIL: I HAVE THE REFERENCE NUMBER FOR THE STEALTH INSERTS, AND IT IS A0A13. ARE OTHER FACILITIES HAVING THESE SAME ISSUES? IT IS ALMOST LIKE THE RUBBER IS TOO SLIPPERY TO GRIP THE AORTA ADEQUATELY. WE USED A DIFFERENT CLAMP TO RESOLVE THE CASE. EACH TIME, IT HAS BEEN SL

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 24447506 ↗
  • MalfunctionFeb 25, 2026Device Slipped

    PROCEDURE PERFORMED: DESCENDING THORACIC AORTA TO BILATERAL FEMORAL BYPASS. EVENT DESCRIPTION: I AM FROM "[HOSPITAL]" IN (B)(6). ONE OF OUR SURGEONS REALLY LIKES THE AORTIC STEALTH CLAMPS, HOWEVER THE CLAMP INSERTS ARE QUITE SLIPPERY AND HAVE ACTUALLY COME OFF THE AORTA A FEW DIFFERENT TIMES. DO YOU HAVE A MORE 'STICKY' INSERT THAT WILL NOT SLIDE OFF OF THE AORTA? HE IS USING THEM CORRECTLY. ARE OTHER HOSPITALS REPORTING THIS PROBLEM? HE HAS UNFORTUNATELY, STOPPED USING THEM AS HE CANNOT RISK THEM SLIDING DURING THE PROCEDURE. THREE DEVICE MALFUNCTIONS OCCURRED DURING THREE DIFFERENT PROCEDURES. THE SAME DOCTOR EXPERIENCED THE ISSUE THREE TIMES. IT HAPPENED A COUPLE OF TIMES, THE RUBBER DOESN'T GRIP THE TISSUE THE WHOLE CLAMP FALLS OFF THE AORTA. THE TYPE OF INSERT TYPICALLY USED IS STEALTH INSERT. THERE WAS NO SEPARATION OF THE INSERT PAD FROM THE BASE. ADDITIONAL INFORMATION PROVIDED BY "[NAME]" ON 28JAN2026 BY EMAIL: I HAVE THE REFERENCE NUMBER FOR THE STEALTH INSERTS, AND IT IS A0A13. ARE OTHER FACILITIES HAVING THESE SAME ISSUES? IT IS ALMOST LIKE THE RUBBER IS TOO SLIPPERY TO GRIP THE AORTA ADEQUATELY. WE USED A DIFFERENT CLAMP TO RESOLVE THE CASE. EACH TIME, IT HAS BEEN SLIP

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 24447484 ↗
  • MalfunctionFeb 13, 2026Device Slipped

    PROCEDURE PERFORMED: DESCENDING THORACIC AORTA TO BILATERAL FEMORAL BYPASS EVENT DESCRIPTION: I AM FROM "[HOSPITAL]" IN (B)(6). ONE OF OUR SURGEONS REALLY LIKES THE AORTIC STEALTH CLAMPS, HOWEVER THE CLAMP INSERTS ARE QUITE SLIPPERY AND HAVE ACTUALLY COME OFF THE AORTA A FEW DIFFERENT TIMES. DO YOU HAVE A MORE 'STICKY' INSERT THAT WILL NOT SLIDE OFF OF THE AORTA? HE IS USING THEM CORRECTLY. ARE OTHER HOSPITALS REPORTING THIS PROBLEM? HE HAS UNFORTUNATELY, STOPPED USING THEM AS HE CANNOT RISK THEM SLIDING DURING THE PROCEDURE. THREE DEVICE MALFUNCTIONS OCCURRED DURING THREE DIFFERENT PROCEDURES. THE SAME DOCTOR EXPERIENCED THE ISSUE THREE TIMES. IT HAPPENED A COUPLE OF TIMES, THE RUBBER DOESN'T GRIP THE TISSUE THE WHOLE CLAMP FALLS OFF THE AORTA. THE TYPE OF INSERT TYPICALLY USED IS STEALTH INSERT. THERE WAS NO SEPARATION OF THE INSERT PAD FROM THE BASE. ADDITIONAL INFORMATION PROVIDED BY "[NAME]" ON 28JAN2026 BY EMAIL: I HAVE THE REFERENCE NUMBER FOR THE STEALTH INSERTS, AND IT IS A0A13. ARE OTHER FACILITIES HAVING THESE SAME ISSUES? IT IS ALMOST LIKE THE RUBBER IS TOO SLIPPERY TO GRIP THE AORTA ADEQUATELY. WE USED A DIFFERENT CLAMP TO RESOLVE THE CASE. EACH TIME, IT HAS BEEN SLIPP

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 24338686 ↗

FDA clearance and classification

The FDA regulates product code DXC (clamp, vascular) under 21 CFR 870.4450 as a Class II device. 21 CFR 870.4450 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the STEALTH® CLAMP INSERTS A0A13

Has the STEALTH® CLAMP INSERTS A0A13 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00607915113001), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the STEALTH® CLAMP INSERTS A0A13?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Feb 13, 2026 and Feb 25, 2026. In those reports, the most common specific device problem is “Device Slipped” (3); the most common patient problem is “Hemorrhage/Blood Loss/Bleeding” (3). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the STEALTH® CLAMP INSERTS A0A13?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the STEALTH® CLAMP INSERTS A0A13, and is it still sold?

The FDA's device database lists Applied Medical Resources Corporation as the company responsible for it (brand STEALTH® CLAMP INSERTS, model A0A13). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Clamp, Vascular devices

All Clamp, Vascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page