STEALTH® FIBRA® CLAMP INSERT A0G01Recalls & reported problems
Applied Medical Resources Corporation · CLAMP INSERT · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00607915114725), received between Mar 6, 2017 and Feb 6, 2025 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Loss of or Failure to Bond” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loss of or Failure to Bond8
- Detachment of Device or Device Component1
- Fracture1
Patient problems
- No Known Impact Or Consequence To Patient8
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
PROCEDURE PERFORMED: [CARDIAC SURGERY] EVENT DESCRIPTION: [HOSPITAL: [HOSPITAL NAME] DISTRIBUTOR: [DISTRIBUTOR NAME] MODEL #: A0G01. LOT #: 1510129. THE PROCEDURE WAS CARDIAC SURGERY. WHILE USING THE INSERTS TO BLOCK BLOOD FLOW, THE RUBBER PART OF THE INSERTS DETACHED FROM THE PRODUCT. THE SURGEON DID NOT NOTICE ANYTHING WRONG DURING THE SETUP OR WHEN IT STARTED TO BE USED. HOWEVER, THE DETACHMENT WAS DISCOVERED LATER DURING ITS USE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING THE INSERT CONTINUOUSLY. THE PRODUCT IN QUESTION WAS RETURNED FROM THE HOSPITAL. PLEASE INVESTIGATE THE POSSIBLE ROOT CAUSE. THE HOSPITAL HAS REQUESTED THE INVESTIGATION REPORT. ADDITIONAL INFORMATION OBTAINED FROM APPLIED MEDICAL REP VIA EMAIL ON 27JAN2025: NO OTHER DEVICES OR INSTRUMENTS WERE USED. THE SURGICAL TEAM NOTICED DURING AORTIC CLAMPING THAT THE DEVICE MALFUNCTIONED. THE INSERT PAD PARTIALLY DETACHED FROM THE INSERT. THE PAD DID NOT FALL INTO THE PATIENT'S BODY. THERE WAS NO DAMAGE TO THE VESSEL. THE SURGEON NOTICED SOMETHING UNUSUAL WHILE USING THE INSERT, WHICH LED TO THE IDENTIFICATION OF A MALFUNCTION. TYPE OF INTERVENTION: [THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING THE INSERT CO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21324176 ↗MalfunctionMar 7, 2024Fracture; Fracture
PROCEDURE PERFORMED: CARDIAC SURGERY. EVENT DESCRIPTION: ON (B)(6) 2024. [PROCEDURE: CARDIAC SURGERY] WHEN IT WAS OPENED, NURSE NOTICED THAT THE INSERT IN QUESTION WAS BROKEN. THE PROCEDURE WAS COMPLETED WITH NO PROBLEM BY USING ANOTHER A0G01. COMPLAINT SAMPLE WILL BE RETURNED FROM THE HOSPITAL. PRODUCTS TO BE RETURNED: 1 INSERT. ADDITIONAL INFORMATION VIA EMAIL ON 8JAN2024: THIS INSERT WAS NOT ACTUALLY USED IN THE SURGERY BECAUSE THE DAMAGE WAS NOTICED WHEN THE POUCH WAS OPENED. THEREFORE, OTHER A0G01 WAS USED FOR THE SURGERY AND IT WAS COMPLETED WITH NO PROBLEM. IN ADDITION, THE INSERT DID NOT ENCOUNTER ANY SHARP INSTRUMENTS DURING THE PROCEDURE. TYPE OF INTERVENTION: THE PROCEDURE WAS COMPLETED WITH NO PROBLEM BY USING ANOTHER A0G01. PATIENT STATUS: NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18862587 ↗MalfunctionMar 6, 2017Loss of or Failure to Bond; Loss of or Failure to Bond
FEMORAL/TIBIAL FIXATION - "DURING DR. (B)(6) FEM/TIB PROCEDURE THIS MORNING, THEY OPENED THE INSERT AND NOTICED THAT THE RUBBER PORTION OF THE INSERT WAS COMING OFF. THEY OPENED 7-8 OF THE SAME LOT NUMBER AND THEY WERE ALL THE SAME. WE HAVE PULLED THE ENTIRE LOT OFF OF OUR SHELVES." PATIENT STATUS: UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6382263 ↗MalfunctionMar 6, 2017Loss of or Failure to Bond; Loss of or Failure to Bond
FEMORAL/TIBIAL FIXATION- "DURING DR. (B)(6)'S FEM/TIB PROCEDURE THIS MORNING, THEY OPENED THE INSERT AND NOTICED THAT THE RUBBER PORTION OF THE INSERT WAS COMING OFF. THEY OPENED 7-8 OF THE SAME LOT NUMBER AND THEY WERE ALL THE SAME. WE HAVE PULLED THE ENTIRE LOT OFF OF OUR SHELVES." PATIENT STATUS: UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6382262 ↗MalfunctionMar 6, 2017Loss of or Failure to Bond; Loss of or Failure to Bond
FEMORAL/TIBIAL FIXATION- "DURING DR. (B)(6) FEM/TIB PROCEDURE THIS MORNING, THEY OPENED THE INSERT AND NOTICED THAT THE RUBBER PORTION OF THE INSERT WAS COMING OFF. THEY OPENED 7-8 OF THE SAME LOT NUMBER AND THEY WERE ALL THE SAME. WE HAVE PULLED THE ENTIRE LOT OFF OF OUR SHELVES." PATIENT STATUS: UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6382259 ↗MalfunctionMar 6, 2017Loss of or Failure to Bond; Loss of or Failure to Bond
FEMORAL/TIBIAL FIXATION- "DURING DR. (B)(6) FEM/TIB PROCEDURE THIS MORNING, THEY OPENED THE INSERT AND NOTICED THAT THE RUBBER PORTION OF THE INSERT WAS COMING OFF. THEY OPENED 7-8 OF THE SAME LOT NUMBER AND THEY WERE ALL THE SAME. WE HAVE PULLED THE ENTIRE LOT OFF OF OUR SHELVES." PATIENT STATUS: UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6382257 ↗MalfunctionMar 6, 2017Loss of or Failure to Bond; Loss of or Failure to Bond
FEMORAL/TIBIAL FIXATION- "DURING DR. (B)(6)'S FEM/TIB PROCEDURE THIS MORNING, THEY OPENED THE INSERT AND NOTICED THAT THE RUBBER PORTION OF THE INSERT WAS COMING OFF. THEY OPENED 7-8 OF THE SAME LOT NUMBER AND THEY WERE ALL THE SAME. WE HAVE PULLED THE ENTIRE LOT OFF OF OUR SHELVES." PATIENT STATUS: UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6382256 ↗MalfunctionMar 6, 2017Loss of or Failure to Bond; Loss of or Failure to Bond
FEMORAL/TIBIAL FIXATION- "DURING DR. (B)(6)'S FEM/TIB PROCEDURE THIS MORNING, THEY OPENED THE INSERT AND NOTICED THAT THE RUBBER PORTION OF THE INSERT WAS COMING OFF. THEY OPENED 7-8 OF THE SAME LOT NUMBER AND THEY WERE ALL THE SAME. WE HAVE PULLED THE ENTIRE LOT OFF OF OUR SHELVES." PATIENT STATUS: UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6382255 ↗MalfunctionMar 6, 2017Loss of or Failure to Bond; Loss of or Failure to Bond
FEMORAL/TIBIAL FIXATION- "DURING DR. (B)(6)'S FEM/TIB PROCEDURE THIS MORNING, THEY OPENED THE INSERT AND NOTICED THAT THE RUBBER PORTION OF THE INSERT WAS COMING OFF. THEY OPENED 7-8 OF THE SAME LOT NUMBER AND THEY WERE ALL THE SAME. WE HAVE PULLED THE ENTIRE LOT OFF OF OUR SHELVES." PATIENT STATUS: UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6382253 ↗MalfunctionMar 6, 2017Loss of or Failure to Bond; Loss of or Failure to Bond
FEMORAL/TIBIAL FIXATION- "DURING DR. (B)(6)'S FEM/TIB PROCEDURE THIS MORNING, THEY OPENED THE INSERT AND NOTICED THAT THE RUBBER PORTION OF THE INSERT WAS COMING OFF. THEY OPENED 7-8 OF THE SAME LOT NUMBER AND THEY WERE ALL THE SAME. WE HAVE PULLED THE ENTIRE LOT OFF OF OUR SHELVES." PATIENT STATUS: UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6382252 ↗
Other Clamp, Vascular devices
- Heartstring III System 3.8MM HSK-3038Maquet Cardiovascular LLC · Clamp, Vascular1 recall4 deaths809 reports
- Heartstring III System 4.3MM HSK-3043Maquet Cardiovascular LLC · Clamp, Vascular1 recall265 reports
- TR BAND TRB24-REGTerumo Medical Corporation · CLAMP, VASCULAR284 reports
- TR BAND TRB24-REGTerumo Medical Corporation · CLAMP, VASCULAR256 reports
- HeartString III Proximal Seal HS-3045Maquet Cardiovascular LLC · Clamp, Vascular1 recall187 reports
- TR BAND TRB29-LRGTerumo Medical Corporation · CLAMP, VASCULAR67 reports
- TR BAND TRB29-LRGTerumo Medical Corporation · CLAMP, VASCULAR65 reports
- Flo-Thru Intraluminal Shunt® FT-12225Synovis Life Technologies, Inc. · CLAMP, VASCULAR1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page