Convertible Trocar System C0E05Recalls & reported problems
Applied Medical Resources Corporation · Laparoscopic Trocar Sleeve/Reusable · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00607915115593), received between Nov 26, 2018 and Nov 26, 2018 — classified as 1 malfunction.
The most frequently reported specific device problem is “Sharp Edges” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Sharp Edges1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 26, 2018Sharp Edges; Sharp Edges
PROCEDURE PERFORMED: UNKNOWN ABDOMINAL PROCEDURE. EVENT DESCRIPTION: DISTAL EDGE OF PRODUCT SEVERELY DAMAGED LAPAROSCOPIC INSULATION ON 2 L-HOOKS. INSULATION WAS STRIPPED FROM INSTRUMENT PASSING THROUGH CANNULA AND FELL INTO ABDOMINAL CAVITY. THE OUTER DISTAL OPENING EDGE APPEARED TO BE SHARP CAUSING THE DAMAGE. HOSPITAL WILL BE REPORTING THIS TO THE HEALTH AUTHORITY. ADDITIONAL INFORMATION RECEIVED FROM DISTRIBUTOR VIA EMAIL ON 01NOV2018: RECEIVED CLARIFICATION ON PRODUCT RETURN STATUS. IT IS NOT IMMEDIATELY AVAILABLE AS IT HAS BEEN SENT TO THEIR REGIONAL CLINICAL ENGINEERS FOR INVESTIGATION. ONCE THEIR INVESTIGATION IS COMPLETE THEY WILL BE SENDING THE INSTRUMENT BACK TO THE HOSPITAL. THE HOSPITAL WILL THEN SEND IT TO [DISTRIBUTOR ]. WE WILL THEN SEND IT TO APPLIED. THE HOSPITAL HAS CONFIRMED THAT THE PATIENT DID NOT RECEIVE ANY HARM. PATIENT STATUS: NO ADVERSE PATIENT OUTCOMES REPORTED. THE HOSPITAL HAS CONFIRMED THAT THE PATIENT DID NOT RECEIVE ANY HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8102872 ↗
Other Laparoscope, General & Plastic Surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page