Kii® Sheilded Bladed Access Sysem C0Q61Recalls & reported problems
Applied Medical Resources Corporation · Trocar · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00607915116149), received between Oct 21, 2016 and Aug 13, 2019 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Physical Resistance1
Patient problems
- No Consequences Or Impact To Patient2
- No Patient Involvement1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 13, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
PROCEDURE PERFORMED: HYSTERECTOMY. EVENT DESCRIPTION: LIMITED INFORMATION AVAILABLE AT THE TIME OF REPORT. TWO OF THE TROCARS BROKE DURING SURGERY. SURGEON USED A 12MM TROCAR TO RETRIEVE THE PIECES FROM THE PATIENT. THERE WAS NO PATIENT INJURY. THE DEVICES ARE AVAILABLE FOR RETURN. ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 31JUL2019 FROM MATERIAL COORD THERE WERE NO CONCOMITANT DEVICES USED WITH THE TROCARS. PATIENT LEFT THE OR STABLE - UNSURE OF CURRENT STATUS. PIECES THAT BROKE OFF OF TROCARS WERE RETRIEVED AND SURGERY WAS COMPLETED USING 12 MM TROCAR INSTEAD OF 8 MM. UNSURE WHERE THE FRACTURE OCCURRED. THE FRACTURE OCCURRED AFTER THE TROCARS WERE INSERTED IN USUAL FASHION, A SMALL ROUND PIECE OF PLASTIC FELL OUT FROM SOMEWHERE INSIDE THE TROCAR INTO THE PATIENT ABDOMEN. THE FRACTURE DID NOT CAUSE PARTICULATION. ALL PIECES WERE RETRIEVED FROM THE PATIENT. THE PORT WAS AN ANCILLARY PORT FOR LAPAROSCOPY. PORT WAS INSERTED THROUGH THE ABDOMEN FOLLOWING INSUFFLATION TO ALLOW FOR INSERTION OF INSTRUMENTS DURING LAPAROSCOPY. THERE ARE PHOTOS AVAILABLE THAT SHOW THE BREAK. INTERVENTION: USED A 12MM TROCAR TO RETRIEVE PIECES FROM PATIENT. PATIENT STATUS: PATIENT LEFT THE OR STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8889324 ↗MalfunctionAug 13, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
PROCEDURE PERFORMED: HYSTERECTOMY. EVENT DESCRIPTION: LIMITED INFORMATION AVAILABLE AT THE TIME OF REPORT. TWO OF THE TROCARS BROKE DURING SURGERY. SURGEON USED A 12MM TROCAR TO RETRIEVE THE PIECES FROM THE PATIENT. THERE WAS NO PATIENT INJURY. THE DEVICES ARE AVAILABLE FOR RETURN. ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 31JUL2019 FROM MATERIAL COORD THERE WERE NO CONCOMITANT DEVICES USED WITH THE TROCARS. PATIENT LEFT THE OR STABLE - UNSURE OF CURRENT STATUS. PIECES THAT BROKE OFF OF TROCARS WERE RETRIEVED AND SURGERY WAS COMPLETED USING 12 MM TROCAR INSTEAD OF 8 MM. UNSURE WHERE THE FRACTURE OCCURRED. THE FRACTURE OCCURRED AFTER THE TROCARS WERE INSERTED IN USUAL FASHION, A SMALL ROUND PIECE OF PLASTIC FELL OUT FROM SOMEWHERE INSIDE THE TROCAR INTO THE PATIENT ABDOMEN. THE FRACTURE DID NOT CAUSE PARTICULATION. ALL PIECES WERE RETRIEVED FROM THE PATIENT. THE PORT WAS AN ANCILLARY PORT FOR LAPAROSCOPY. PORT WAS INSERTED THROUGH THE ABDOMEN FOLLOWING INSUFFLATION TO ALLOW FOR INSERTION OF INSTRUMENTS DURING LAPAROSCOPY. THERE ARE PHOTOS AVAILABLE THAT SHOW THE BREAK. INTERVENTION: USED A 12MM TROCAR TO RETRIEVE PIECES FROM PATIENT. PATIENT STATUS: PATIENT LEFT THE OR STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8889319 ↗MalfunctionOct 21, 2016Physical Resistance; Physical Resistance
PROCEDURE PERFORMED: UNKNOWN - "WHEN THE SCRUB ATTEMPTED TO INSERT OBTURATOR INTO CANNULA, IT WOULD NOT FIT. THAT IS WHEN THEY DISCOVERED THE PLASTIC PIECE FOR THE BLADE SHIELD HAD POPPED OUT. DR. (B)(6) WAS THE SURGEON." ADDITIONAL INFORMATION RECEIVED VIA EMAIL FROM (B)(6) ON SEPTEMBER 22, 2016 - "THE TROCAR HAD NO CONTACT WITH THE PATIENT, BECAUSE THEY WERE NEVER ABLE TO PUT THE TROCAR TOGETHER". SEE ATTACHING PICTURES FOR EXPLANATION OF MALFUNCTION. "I WAS TOLD BY THE MATERIALS MANAGER THAT THE SCRUB TECH WAS GETTING THE PRODUCTS READY FOR THE CASE AND WHEN SHE WENT TO INSERT THE OBTURATOR INTO THE CANNULA IT WOULD NOT FIT. THAT IS WHEN THEY NOTICED THAT PLASTIC PIECE HAD PARTIALLY POPPED OUT. THAT IS ALL SHE COULD TELL ME." PATIENT STATUS - NO CONTACT WITH PATIENT. TYPE OF INTERVENTION - NA.
Patient problems: No Patient Involvement
MAUDE report 6049249 ↗
Other Laparoscope, General & Plastic Surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page